Fig. 1
3.3. Meter Mounting – Panel mount version (designated by "P" suffix)
3.3.1. Panel mount versions are equipped with the probe and secondary
sensor connections mounted on the rear of the enclosure. They are
also equipped with two threaded studs on the rear with a U-bracket,
wing nuts and washers. An additional gasket is placed on the rear
facing surface of the H-frame to seal the panel opening.
3.3.2. For detailed installation instructions see Wahl document WD1100
and reference Wahl drawing 13615, both included in section 11 of
this manual.
3.4. Wire Connections – In the event that your probe was shipped
disconnected from the meter, the following describes the probe wiring and
installation process.
3.4.1. Loosen the four Phillips head screws in the front cover until the
cover is removed. Note: The screws are held captive by retaining
washers and should not be removed completely.
3.4.2. Remove the 6 pin pluggable terminal strip connector (J1) from the
PCB mating connector and connect probe wires per the following
table and as shown in Fig. 4.
Probe Connection Wiring
P1 Pin
1
2
3
4
5
6
3.4.3. Connect the pluggable terminal strip to PCB connector J1.
DST600 Series USER MANUAL
Signal Name
Excite +
RTD +
RTD -
Excite -
Probe ID Ground
Probe ID Signal
5 of 20
Bracket Mounting
Holes
Digi-Stem
Mounting Screw
Wire Color
Fixed Probe
Remote Probe
White
White
Black
Black
Brown
Orange
Brown
White
Blue
Black
Gray
Pink