CAUTION: Portable and mobile RF communications equipment can affect the GastroCH
ECK®
4
Gastrolyzer®.
CAUTION: Sensor use beyond the specified component lifetime without servicing may compromise
result accuracy.
CAUTION: Use only accessories designed for use with the GastroCH
ECK® Gastrolyzer®.
4
CAUTION: Please do not attempt to modify the equipment in any way or use accessories not
specified by the manufacturer. Any attempt to do so, will invalidate the warranty and may cause
product failure or damage to your GastroCH₄ECK® Gastrolyzer® monitor. The product warranty does
not cover product failure or damage resulting from use with non-approved accessories.
NOTE: The instrument should be calibrated prior to first use, after transportation, and every 4
weeks.
NOTE: Please allow the GastroCH
ECK® to be switched on for at least 5 minutes prior to calibration.
4
NOTE: The GastroCH
ECK® device temperature needs to be at 21°C ± 4°C (69.8°F ± 39.2°F) to be able
4
to perform a calibration.
NOTE: Bedfont® recommends an annual service in order to check sensor and component parts
performance.
NOTE: Servicing should only be carried out by a Bedfont® trained representative
NOTE: Bedfont® will make available on request service training to appropriately qualified personnel.
NOTE: Please refer to Bedfont's infection control and maintenance guidelines for further
information on infection control.