Bedfont GastroCH4ECK Gastrolyzer User Manual page 24

Hide thumbs Also See for GastroCH4ECK Gastrolyzer:
Table of Contents

Advertisement

CH
sensor check sum/end char error: Press the retry
4
button. If the error occurs more than 3 times, please
restart the unit. If failure persists contact Bedfont® or
their distributor for support.
CH
sensor error: Check that the ambient
4
temperature is within range and that the unit is
plugged correctly. Then press the retry button. If the
error occurs more than 3 times, please restart the
unit. If failure persists contact Bedfont® or their
distributor for support.
CH
sensor failure: Press the retry button. If the error
4
occurs more than 3 times, please restart the unit. If
failure persists contact Bedfont® or their distributor
for support.
CH
sensor reading error: Verify that no gas is flowing
4
through the device. If gas was being used, please wait
5 minutes before attempting a new test and then
press the retry button. If the error occurs more than
3 times, please restart the unit. If failure persists
contact Bedfont® or their distributor for support.
23

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the GastroCH4ECK Gastrolyzer and is the answer not in the manual?

Subscribe to Our Youtube Channel

Related Products for Bedfont GastroCH4ECK Gastrolyzer

Table of Contents