AriaMx_Poster_18x12_2.6.pdf
2
1/15/15
8:56 AM
step
3
Install the optical modules
5) Open the boxes containing the optical modules. Remove the top piece of foam from each box (Figure 8) then remove the
plastic bag containing the optical module.
Figure 5
Optical
module housing
carrier, centered
in the instrument
door opening
Figure 7
Slots
for the optical
modules
C
M
Y
CM
MY
CY
CMY
6) Install the optical modules into the slots.
K
a) Open the plastic bag and remove the optical module.
b) Peel off the plastic film from the edge of the optical module (see figure 9). Once the film is removed, do not touch
the exposed edge.
c) Put the optical module into an available slot. The correct orientation for the optical module is label side up with the
Agilent spark closer to the front of the instrument (see Figure 10).
NOTE: If you are installing fewer than six optical modules, make sure that the empty slots are on the left-most side
of the housing.
7) Lower the lid on the optical module housing until it clicks shut. When you turn on the instrument for the first time, it will
prompt you to calibrate the background for the optical modules.
Figure 9
Removal
of plastic film from
optical module
(continued )
Figure 6
Opening
of the optical
module housing
Figure 8
Optical
module in its
shipping box
Figure 10
Installed
optical modules
step
4
Clean the thermal block
With the instrument door still open, clean the outside and inside surfaces of the thermal block.
1) Lift the lid of the thermal block by pulling forward on the handle of the lid and then lifting the lid up and away from
the thermal block.
2) Using an aerosol can of compressed air, clean out the wells of the thermal block. Hold the can 3–4 inches away from
the thermal block as you press the trigger.
3) Moisten a lint-free cleansing tissue with dH
0, and gently wipe down the thermal block and the underside of the lid.
2
Then, close the lid of the thermal block and wipe down the top of the lid.
4) Close the instrument door.
step
5
Connect the instrument
to a power supply
You must connect the instrument to a grounded AC outlet.
1) Plug the power cord into the power connector at the rear
of the instrument.
2) Connect the cable plug to the outlet.
step
7
Connect the instrument to a network or directly to a PC
If you do not connect your instrument to a PC, you must transfer post-run experiment data by copying it from the
instrument to a USB drive, and then from the USB drive to your PC.
1) Plug one end of an ethernet cable into the ethernet port on the back of the instrument.
Use a standard Cat 6 straight/crossover ethernet cable.
2) Plug the other end of the cable into either a network port or a PC.
Note: If you are connecting the instrument directly into a PC, rather than
through a network, you need to set a static IP address, subnet mask, and static
default gateway on both the PC and the instrument. See the AriaMx Setup and
User Guide for instructions.
step
6
Connect a keyboard or mouse
to the instrument
(optional)
If desired, you can connect a keyboard or mouse to the
instrument via the USB ports on the front and back of the
instrument. Plug the USB cable of the device into a USB port
on the instrument. Multimedia keyboards are not supported.
Copyright ©2015, 2017-2018 Agilent Technologies, Inc
Printed in Malaysia May 2018, revision D0
*G8830-90011*
G8830-90011
For Research Use Only. Not for use in diagnostic procedures.