MAGNETIC RESONANCE IMAGING (MRI) SAFETY INFORMATION (SCS)

All of the components of the Nalu Neurostimulation System are MR Unsafe except for the
following:
Nalu Implantable Pulse Generators (Dual Integrated Model 11002-040, Single Integrated
Model 11001-040)
Nalu Implantable Pulse Generators (Dual Ported Model 11004-002 connected to two 40
cm Nalu Lead Model 12001-040, Single Ported Model 11003-002 connected to a 40 cm
Nalu Lead Model 12001-040)
Nalu Anchor (Model 13001).
WARNING
Do not bring MR Unsafe components of the Nalu Neurostimulation System into the
MRI system room.
It is important to read this entire section prior to conducting or recommending an MRI examination on a user
implanted with the Nalu Neurostimulation System. These instructions only apply to the Nalu
Neurostimulation System and do not apply to other products. If you have any questions, please contact Nalu
Medical or visit www.nalumed.com.
SCS Head and Extremities scan using a transmit/receive head and extremities coil – see page 64
SCS Scan using a whole body RF transmit coil – See page 70
PNS Head and Extremities scan using a transmit/receive head and extremities coil – See page 75
Nalu User's Kit Instructions for Use PN: MA-000007 Rev J