Liquiline System CA80PH
Endress+Hauser
Menu/Setup/Sample preparation/Installation type = Inlet pipe sampling/Sample preparation 1 ... 2
Function
Options
Filter backflush
Selection
• On
• Off
Factory setting
On
Cleaning duration
Selection
10 to 60 s
Factory setting
00:30
Cleaning discard time Selection
0 to 1800 s
Factory setting
180 s
Lead time
0:00 to 30:00 min:ss
Factory setting
25:00 min
Duration
0:00 to 60:00 min:ss
Factory setting
25:00 min
Timeout no sample
3:00 to 30:00 min:ss
Reset settings
10.5
Current inputs
The input can be used as the data source for limit switches and logbooks, for example.
Furthermore, external values can be made available as set points for controllers.
Menu/Setup/Inputs/Current input x:y
Function
Mode
Input variable
Meas. value format
Parameter name
Input variable = Parameter
Unit of measure
Input variable = Parameter
Info
Cleaning duration of the automatic water or compressed
air cleaning system
Discard time once cleaning is complete. If water is used as
the flushing medium, this must be replaced by a fresh
sample before starting the next measurement.
Resets all settings specific to sample preparation. All other
settings are retained.
1)
Options
Info
‣
Selection
• Off
• 0 - 20mA
• 4 - 20mA
Factory setting
4 - 20mA
‣
Selection
• Parameter
• Current
Factory setting
Current
‣
Selection
• #
• #.#
• #.##
• #.###
Factory setting
#.#
‣
Customized text, 16
characters
Customized text, 16
You cannot choose the unit from a list. If you
characters
want to use a unit, you must enter it here as
customized text.
Select the same current range as in the data
source (connected device).
Select the input variable.
Specify the number of decimal places.
Assign a useful name, e.g. the parameter
name that the data source also uses.