Current inputs

The input can be used as the data source for limit switches and logbooks, for example.
Furthermore, external values can be made available as set points for controllers.
Menu/Setup/Inputs/Current input x:y
Function
Mode
Input variable
Meas. value format
Parameter name
Input variable = Parameter
Unit of measure
Input variable = Parameter
Lower range value
Input variable = Parameter
Upper range value
Input variable = Parameter
Damping
1)
x:y = slot no. : input number
1)
Options
Info
Selection
• Off
• 0 - 20mA
• 4 - 20mA
Factory setting
4 - 20mA
Selection
• Parameter
• Current
Factory setting
Current
Selection
• #
• #.#
• #.##
• #.###
Factory setting
#.#
Customized text, 16
characters
Customized text, 16
You cannot choose the unit from a list. If you
characters
want to use a unit, you must enter it here as
customized text.
-20.0 ... Upper range
value <unit of measure>
Factory setting
0.0 <engineering unit>
Lower range value to
10000.0 <engineering
unit>
Factory setting
10.0 <engineering unit>
0 to 60 s
The damping causes a floating average curve of
the measured values over the time specified.
Factory setting
0 s
Select the same current range as in the data
source (connected device).
Select the input variable.
Specify the number of decimal places.
Assign a useful name, e.g. the parameter
name that the data source also uses.
Enter the measuring range. The lower and
upper range values are assigned to the 0 or
4 mA value and the 20 mA value
respectively. The system uses the
engineering unit which you entered
beforehand.