Huazheng Electric Manufacturing HZ-1521 User Manual

Pec energy meter field calibrator

Advertisement

Quick Links

HZ-1521 PEC Energy Meter Field
Calibrator
USER MANUAL
Huazheng Electric Manufacturing(Baoding) Co.,Ltd

Advertisement

loading
Need help?

Need help?

Do you have a question about the HZ-1521 and is the answer not in the manual?

Questions and answers

Summary of Contents for Huazheng Electric Manufacturing HZ-1521

  • Page 1 HZ-1521 PEC Energy Meter Field Calibrator USER MANUAL Huazheng Electric Manufacturing(Baoding) Co.,Ltd...
  • Page 2 As for updating product design and technical specifications, prior notification shall not be given, and the physical products shall prevail. Huazheng Electric Manufacturing(Baoding) Co.,Ltd reserves the right to final interpretation of this product.
  • Page 3: Table Of Contents

    Contents Ⅰ.spare parts..............................1 Ⅱ. Introduction to single-phase energy meter multi-function calibrator..........1 Ⅲ. How to use the functions........................5 Ⅳ. How to use HZ-1521..........................11 Ⅴ. Frequently Asked Questions.......................16 Ⅵ. Instrument maintenance........................17 Ⅶ. Appendix..............................18...
  • Page 4: Ⅰ.spare Parts

    Ⅰ.spare parts 1.1 List of parts(50Aclamp meter) Name Qty. Remark HZ-1521 single-phase handheld field calibrator 1 set Standard configuration 50A current clamp 1 pcs Standard configuration Voltage test line 1 pcs Standard configuration Current test line 1 pcs Standard configuration...
  • Page 5 Before work, users can make task plans on the HZ-1521 , which is beneficial to orderly working. Calibration plan can be downloaded via PEC management software through serial port into the instrument. ● Input: the instrument supports for Chinese characters, English case and Arabic number input.
  • Page 6 system; The accuracy of electric energy is ensured within the temperature range of - 20℃ ~ + 40℃;  With 0.3- magnitude 50A clamp optional (including opening and closing non-repeatability  error, contact error, interference error of outside magnetic field and angle error, etc.). With 50A, 100A, 500A and 1000A clamps optional, used for direct measurement and low ...
  • Page 7 direction of current. 2.5 Wiring Check the wiring methods of the single - phase energy meter: Load on the field Connect the energy meter "live" wire on the instrument side to the instrument’s U + port, and connect the "zero" line to the instrument’s U - port; connect the clamp meter line to the instrument’s "clamp meter"...
  • Page 8: Ⅲ. How To Use The Functions

    ratio is 1:1 and it is input under other situations according to the transformation ratio of field TA); Error: the error parameters generated by the energy meter during field calibration. The instrument can calculate the percentage of the values of electric quantity measured by the energy meter and the difference in electric quantity actually measured by a standard meter according to the input energy meter constant and collected energy meter power pulses).
  • Page 9 parameter value by pressing [Confirm] key. Electric meter series Electric meter series: refers to the level of the measured energy meter, which will be indicated on the electric meter. Operation method: press "←, →" and the cursor jumps to the electric meter series setting; press [Confirm] key and jump out of a variety of options;...
  • Page 10 Ratio: generally select the default value. Notes to the calibration parameter settings: The difference between "photoelectric" mode and manual mode is large. For example: check the three-phase mechanical active energy meter, and the electric energy constant is 450 r/kW. h and the set number of laps is 2. When calibrating by use of the photoelectric mode, every time when the energy meter’s black mark moves a lap, the photoelectric sampler generates a pulse;...
  • Page 11 Harmonic order - refers to the multiples of fundamental wave frequency, relative to fundamental wave. If the frequency of mains supply is 50Hz, the 2nd harmonic is 100Hz AC component, the 3rd harmonic is 150Hz AC component and so on. Channel indicator -- the harmonic state of the channel currently tested.(six channels are respectively:);...
  • Page 12 5. Leading to the metering and measuring instruments (such as: energy meter) not to give correct instructions or measurement; 6. Interfering with the communication system, lowering down the quality of signal transmission, disturbing the normal transmission signal, and even damaging the communications equipment. Instrument voltage-current harmonic interface Timing Press [7] key to enter the time settings interface and set the required year, month, day and time...
  • Page 13 input "55"; next, press "↓" key, move the cursor to second, and input "22"; press [Confirm] to complete time setting, and return to the main interface system. Waveform Press [8] key to enter into the waveform display interface, and there are two options voltage and current on the right of the interface;...
  • Page 14: Ⅳ. How To Use Hz-1521

    4. Press [Confirm] key to save the user information and return to the main interface, press [Exit] key not to save the edited and modified information and return to the main interface. Ⅳ. How to use HZ-1521 How to calibrate the single-phase energy meter...
  • Page 15 instrument U-, and connect the clamp meter line with the clamp meter interface); 2. Connect the energy meter voltage line, current line, and current clamp meter (when the instrument voltage is input from the live wire inlet, the clamp meter should be connected to the live wire outlet;...
  • Page 16 "LINAN", move the cursor by [↑↓] to the "meter number" column, and press [Switch] key to switch the input method and input 2096157 (if it is mistyped, press [F5] to single delete, and press [F3] to delete the entire line); move the cursor by [↑↓] to the "inspector" column, press [Switch] key to switch the input method and input GDZ (if it is mistyped, press [F5] to single delete, and press [F3] to delete the entire line).
  • Page 17 live wire outlet; when the instrument voltage is input from the live wire outlet, the clamp meter should be connected to the live wire inlet; attention is paid to the current direction of the clamp meter, and the current flows from the clamp meter "+" into "-") 4.
  • Page 18 [Switch] key to switch the input method and input GDZ (if it is mistyped, press [F5] to single delete, and press [F3] to delete the entire line). After the settings are completed, press [Confirm] key for storage. Note: when testing the energy meter with no load (superposition), directly connect the current test line to raise the current automatically.
  • Page 19: Ⅴ. Frequently Asked Questions

    energy meter error value each phase voltage, current, power and other data detailed in the calibration records in this interface). 3. Press [Exit] key to return to the data query interface Note: if the calibration task you want to find can't be found on the current interface, search by turning pages through [←...
  • Page 20: Ⅵ. Instrument Maintenance

    b) When the input is greater than 100A, the set transformation ratio is not correct; c) The meter constant setting is abnormal; d) Active and reactive power settings are abnormal; e) Manual and photoelectric settings are not correct; (2) Check whether the clamp meter is in the reverse direction. 4.
  • Page 21: Ⅶ. Appendix

    Ⅶ. Appendix Appendix 1:How to use the photoelectric sampler The photoelectric sampler shows accurate and fast beam focusing at the time of field calibration, which directly affects the efficiency of field work. Our company has paid close attention to the development of photoelectric sampler and the quality of relevant parts since PEC-1 field calibrator was launched.
  • Page 22 Appendix 2:Technical indicators of HZ-1521 Accuracy when Function Accuracy when using a clamp meter using a terminal Voltage 0.1% 0.1% Main Current 0.3%(0.1A~60A ) 0.5%(60A~1000A) 0.1% (0.1A~10A ) functions 0.3% Guaranteed long-term accuracy (1A~50A) Power 0.1% Active power 0.3% Guaranteed long-term accuracy (1A~50A) 0.1%...