Moisture analyser for single paper sheets

Advertisement

Device maintenance instructions
• Please note that the moisture sensor is a sensitive measuring
instrument and handle it with care.
• The white sensor plate (measuring plate) most not be exposed to
mechanical stress. Heavy mechanical load may cause the break-
ing of the sensor plate. In this case, return the sensor to Schaller
GmbH. The breaking of the sensor plate is not subject to warranty.
• If the device is transported, secure the lid to prevent unintentional
opening.
• We recommend to regularly save all measuring data in the data-
base.
• Furthermore we recommend to regularly save the database with
eventual added or modified calibration curves.
Technical data
Grammage range
Paper size
Paper thickness
Resolution
Measuring range
Operating temperature
Storage temperature
Temperature compensation automatic
Menu languages
Power supply
Dimensions
Weight
Protection class
Scope of supply
page 12
30 to 800 g/m²
minimum 180x180mm
up to 0.8mm (higher range on request)
0.01% water content; 0.1°C temperature
1 to 20 % water content (depending of
characheristics curve)
+10°C to +40°C / 50°F to 104°F
-20°C to +60°C / -4°F to 140°F
English, German
Power supply via PC 5VDC (USB
connector)
280 x 200 x 135 mm
4.8 kg
IP 30
moisture sensor, PC analysing software,
USB connection cable for PC