3. Make sure the cylinder is connected to the compressor.
Refer to Connecting the Cylinder to the Compressor in
this section of the manual.
4. Push the power switch A on the control panel to the
On ( I ) position.
5. Examine the control panel B. The following sequence
of events should occur:
a. 0-3 minutes - The O
C will be lit while the compressor warms up.
b. After three minutes - The FILLING (GREEN) light D
will be on while the cylinder is filling.
c. The FULL (GREEN) light E will be on when the
cylinder is finished filling. Proceed to STEP 6.
The O
BELOW NORMAL (YELLOW) light
2
may come on if the oxygen level from the
concentrator has not yet reached or drops below
90%. The compressor will begin or resume
filling when the oxygen level reaches 90%. If
light does not go Off within 10 minutes, refer to
Troubleshooting in this manual.
If the ATTENTION (RED) light and audible alarm
F are on refer to Troubleshooting in this manual.
6. Remove the full cylinder. Refer to Disconnecting the
cylinder from the Compressor in this section of the
manual.
7. Push the power switch to the Off (O) position.
For cylinder operation and use, refer to the specific
user manual for the cylinder/regulator assembly.
1195663-C