Gill LevelPro 7010 Manual

Robust level sensors for varied applications
Hide thumbs Also See for LevelPro 7010:

Advertisement

Quick Links

Robust level sensors for varied applications
Manual
gillsc.com
7010-30-086 Iss. 5
© 2021 Gill Sensors & Controls Limited

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the LevelPro 7010 and is the answer not in the manual?

Questions and answers

Summary of Contents for Gill LevelPro 7010

  • Page 1 Robust level sensors for varied applications Manual gillsc.com 7010-30-086 Iss. 5 © 2021 Gill Sensors & Controls Limited...
  • Page 2: Table Of Contents

    4.5 SAE 5-Bolt installation 4.6 2” NPT Thread installation 5. Specifications 5.1 General 5.2 Mechanical 5.3 Electrical 5.4 Analogue outputs 5.5 Connections 5.6 Part number configuration 6. Maintenance 7. Appendix gillsc.com 7010-30-086 Iss. 5 © 2021 Gill Sensors & Controls Limited...
  • Page 3: Product Description

    If requested, further accuracy can be gained by compensating for the effect of temperature changes on the output in a given liquid with offsets across the temperature range. Your chosen calibration has passed through Gill Sensors & Controls Limited’s quality control to ensure the sensor provides market leading measurement accuracy in your specified liquid.
  • Page 4: Product Overview

    Robust level sensors for varied applications A breakdown of the sensor is provided below with key measurement considerations. Please contact Gill for any additional information required. The sensor electronics are housed in the flange, outside the measurement tank. The outer tube is placed into the measurement tank, such that will be submerged in liquid.
  • Page 5: Npt Models

    Robust level sensors for varied applications 1.3 NPT models Cable Flange NPT Thread Outer Tube NOTES Diagram NOT to scale All dimensions in mm See technical drawings for full detail gillsc.com 7010-30-086 Iss. 5 © 2021 Gill Sensors & Controls Limited...
  • Page 6: Getting Started

    Over Voltage Protection >32 VDC >32 VDC Supply Current @ 12 VDC <15 mA <15 mA Reverse Polarity Protection Analogue Resolution 10 bit 10 bit Report Rate 10 Hz 10 Hz gillsc.com 7010-30-086 Iss. 5 © 2021 Gill Sensors & Controls Limited...
  • Page 7: Configuration

    One free USB port to connect a sensor A connector is required to configure the sensor with a computer. Gill Sensors & Controls Limited provide an RS232 to USB adaptor (P/N 1484-00-086). Please contact us if your require this device.
  • Page 8: Tank Profiling

    In this situation, 50% of the tank volume is not half way up the probe. To correct for this tank profiling can be used. This can be done in the GSlevel software and is covered in detail in the GSlevel User Manual. gillsc.com 7010-30-086 Iss. 5 © 2021 Gill Sensors & Controls Limited...
  • Page 9: Installation

    4. Installation Do not remove the packaging until you are ready to install the sensor. Prior to installation, please check the sensor for any signs of transit damage. Contact Gill Sensors & Controls Limited immediately if this is the case. IMPORTANT DO NOT attempt to cut the sensor to length or bend the sensor.
  • Page 10: Parts Supplied

    Please contact Gill Sensor & Controls Limited to discuss specific sensor requirements when mounted at an angle.
  • Page 11: 5" Unf Thread Installation

    Fit the O-ring above the male thread on the sensor. Ensure the O-ring is centred and sits within the chamfer; tighten the sensor using a adjustable spanner until the sensor face is flush with the top of the tank and a torque of 50 Nm is achieved. gillsc.com 7010-30-086 Iss. 5 © 2021 Gill Sensors & Controls Limited...
  • Page 12: Using The Gasket

    Fit the gasket over the thread mount, black face up. Ensure the gasket is centred under the sensor and tighten the sensor using a adjustable spanner to a torque of 50 Nm. gillsc.com 7010-30-086 Iss. 5 © 2021 Gill Sensors & Controls Limited...
  • Page 13: Bsp Thread Installation

    Fit the O-ring above the male thread on the sensor. Ensure the O-ring is centred and sits within the chamfer; tighten the sensor using a adjustable spanner until the sensor face is flush with the top of the tank and a torque of 50 Nm is achieved. gillsc.com 7010-30-086 Iss. 5 © 2021 Gill Sensors & Controls Limited...
  • Page 14: Using The Gasket

    Fit the gasket over the thread mount, black face up. Ensure the gasket is centred under the sensor and tighten the sensor using a adjustable spanner to a torque of 50 Nm. gillsc.com 7010-30-086 Iss. 5 © 2021 Gill Sensors & Controls Limited...
  • Page 15: Sae 5-Bolt Installation

    The sensor should be fitted with a gasket if the SAE 5-bolt installation method is used. Fit the gasket over the bolts, black face up, and tighten the bolts to a torque of 10 Nm each. gillsc.com 7010-30-086 Iss. 5 © 2021 Gill Sensors & Controls Limited...
  • Page 16: 2" Npt Thread Installation

    2” NPT fitting. Use of PTFE tape or pipe sealant is recommended for this installation type. PTFE or Tapered threads Pipe sealant gillsc.com 7010-30-086 Iss. 5 © 2021 Gill Sensors & Controls Limited...
  • Page 17: Specifications

    0.25 V – 4.75 V Maximum Output Range: 0.25 V – 10.0 V Current Output: 4 – 20 mA Switch Output: Open collector output 50 V / 1A max Switch to ground gillsc.com 7010-30-086 Iss. 5 © 2021 Gill Sensors & Controls Limited...
  • Page 18: Connections

    5.6 Part number configuration The configuration of the sensor can be determined from its part number designation, engraved on the flange. The interpretation of the part number designation is outlined below. gillsc.com 7010-30-086 Iss. 5 © 2021 Gill Sensors & Controls Limited...
  • Page 19: Maintenance

    Please contact Gill Sensors or Controls Ltd using the details at the end of this document in the first instance with any further issues to be resolved.
  • Page 20: Appendix

    Methods of testing and required test results Electrical equipment for measurement, control and laboratory use. EMC requirements. EN 61326-1:2013 Particular requirements. Signed for and on behalf of: Gill Sensors & Controls Limited Place of Issue Lymington, United Kingdom Date of Issue...
  • Page 21 Robust level sensors for varied applications Important Notices: Gill Sensors & Controls Limited can take no responsibility for installation and/or use of its equipment if „ this is not done in accordance with the appropriate issue and/or amendment of the manual.