One Point with an Offset Correction

Use the following procedure to calibrate consistency with one sample point (you can also use the water point
stored in the transmitter as the first sample point):
,
,
,
,
,
,
,
,
,
,

Two-point Sampling Method for Calculating Offset and Gain

Use the following procedure to calibrate the consistency measurement with two sample points (you can also use
the water point stored in the transmitter as the first sample point):
,
,
,
,
,
13
An OK message will appear in and the display will automatically return to
the Sample sub-menu. Press the ESC key until the PV reading in the main
menu is displayed.
Starting at the default PV display value; Press the ESC button.
Use the UP/DOWN arrows to move to the Calibration sub-menu and press
ENTER.
Use the UP/DOWN arrows to move to the Calibrate sub-menu and press
ENTER.
Transmitter asks you to enter first calibration data point. Use the
UP/DOWN arrows to move to the corresponding sample and press ENTER.
An OK message shortly appears in the display followed by the request to
enter next data point and next sample item will appear in the display.
Press the ESC button to skip and perform one-point calibration.
New gain and offset values will scroll through the display. When SAVE?
appears in the display, press ENTER to accept or ESC to exit without
saving.
An OK message will appear in and the display will automatically return to
the Calibrate sub-menu. Press the ESC key until the PV reading in the
main menu is displayed
Starting at the default PV display value; Press the ESC button.
Use the UP/DOWN arrows to move to the Calibration sub-menu and press
ENTER.
Use the UP/DOWN arrows to move to the Calibrate sub-menu and press
ENTER.