Setting The Range For I02 (Vc Multichannel) - Satron Instruments VC Quick Start Manual

Multichannel transmitter
Table of Contents

Advertisement

,
,
,
,
,
,
,

Setting the Range for I02 (VC Multichannel)

,
,
,
,
,
,
,
,
,
,
,
,
,
,
,
,
,
Use the UP/DOWN arrows to place the decimal separator in the proper
location and press ENTER. A separator will appear to indicate the digit that
will be modified. Use the UP/DOWN arrows to change the number and
press ENTER. The separator will move to the right position. Press ENTER
until the right-hand digit is indicated and press ENTER one more time.
When SAVE? appears in the display, press ENTER to accept or ESC to exit
without saving.
Press the Esc button until the PV reading in the main menu is displayed.
Starting at the default PV display value; Press the ESC button.
Press the ENTER button to enter the Configuration sub-menu.
Use the UP/DOWN arrows to move within the Configuration sub-menus.
When the MA Output sub-menu appears, press the ENTER key and MA 1 ?
menu item will appear in the display.
Use the UP/DOWN arrows to move within the MAOUTPUT sub-menu.
When MA 2 ? menu item appears, press ENTER key and lower range value
(LRV) menu item will appear.
Press the ENTER button to display the current LRV.
Use the UP/DOWN arrows to place the decimal separator in the proper
location and press ENTER. A separator will appear to Indicate the digit that
will be modified. Use the UP/DOWN arrows to change the number and
press ENTER. The separator will move to the right position. Press ENTER
until the right-hand digit is indicated and press ENTER one more time.
When SAVE? appears in the display, press ENTER to accept or ESC to exit
without saving.
Use the DOWN arrow to move to the upper range value (URV).
Press the ENTER button to display the current URV.
Use the UP/DOWN arrows to place the decimal separator in the proper
location and press ENTER. A separator will appear to indicate the digit that
will be modified. Use the UP/DOWN arrows to change the number and
press ENTER. The separator will move to the right position. Press ENTER
until the right-hand digit is indicated and press ENTER one more time.
When SAVE? appears in the display, press ENTER to accept or ESC to exit
without saving.