Teledyne Isco; Retriever 500; Remote Chart Drive Control By Non-Isco Equipment - Teledyne UA-6 Installation And Operation Manual

Absorbance detector
Table of Contents

Advertisement

UA-6 Detector
Section 3 Operation

3.13.2 Teledyne Isco

Retriever 500

Figure 3-12 Cable connection to the Retriever 500
3.14 Remote Chart Drive
Control by Non-Isco
Equipment
3-18
Use cable 60-1020-217 to connect the UA-6 with a Teledyne Isco
Retriever 500 fraction collector. The information transmitted
between the two units consists of tube change and rack change
event marks.
RETRIEVER 500
CABLE #60-1020-217
The REMOTE CHART DRIVE jack is a polarized, four-pin jack
that mates with a Cinch-Jones S-304-CCT plug or equivalent.
The pins of the plug are:
1. 50 or 60 Hz time base (at 19 VAC)
2. common
3. pulse input
4. –15 VDC
This plug can be used to stop the chart drive motor, or to drive
the motor at a variable rate.
Connect a manual switch across pins 1 and 2. The chart drive
stops when the switch is closed.
Short pins 1 and 2. The recorder chart motor may be driven with
line frequency (60 Hz at 12 to 120 VAC) across pins 3 and 4. TTL
pulses may be used by placing an optoisolator between the TTL
signal and pins 3 and 4 using the suggested circuit in Figure
3-13.