ELGA PURELAB flex 1 Service Manual page 51

Table of Contents

Advertisement

Service Bulletin – LWGSBN0086
The following screen will list the types of
system available. Choose the correct model and
press accept:
TOC System – PURELAB flex 2
Basic System – PURELAB flex 1
RO System – PURELAB flex 3
Direct Feed System – PURELAB flex 4
Select the below options to exit the test procedure:
Skip Service Calibration? Select No and press accept, calibrate the system as per the
instructions in the service manual.
Test Procedure Diagnostics. Press accept.
Purge System? Select Skip and press accept.
Skip Outlet Valve Adjustment? Select No and press accept, follow the instructions on
page 3 of this bulletin to ensure that the outlet valve is adjusted correctly.
Skip Tubing Test? Select Yes and press accept.
Skip Conditioning? Select Yes.
Please be aware that a purge will now be performed on the system; please
ensure that there is a container available to collect any water that may be
dispensed from the handset.
Press accept.
Press accept again to stop the purge and complete the process.
Power down the system.
When turned back on the system will be in commissioning mode.
Page 6 of 10