Notes on the Service Manual ........................ 8 Modification overview ..........................9 Up to S/N 0766xxxxxx From S/N 0767xxxxxx to 0950xxxxxx From S/N 0951xxxxxx to 1079xxxxxx From S/N 1080 xxxx to 1556xxxxx From S/N 1557xxxxxx new pneumatics (from end 2007) Fault messages ............................. 27 Check base lines Fluctuating gas flow 3.2.1...
Page 5
Analysis takes too long 3.8.1 Additional reasons for excessive analysis duration could be: 3.8.1.1 One of the IR-sources is unstable: Development of dust when closing the furnace 3.10 Irregular results 3.11 The results are always 0.0000% or 00.0ppm: 3.12 Combustion tube - life time is too short 3.13 IR-cell temperature control is out of range 3.14...
Page 6
4.2.2.1 Supply voltages and thermostatic control circuit 4.2.2.2 Supply voltages for IR signal processing circuit 4.2.3 Test points IR-ranges 4.2.3.1 Supply voltages for IR signal processing circuit 4.2.3.2 Outputs (base lines about 8VDC. For golden paths adjust 6VDC) 4.2.4 Base line adjustments 4.2.5 Infrared cell board IRC1.x sections assignment 4.2.5.1...
Page 7
Gas flow system 6.2.3 Valve controlling 6.2.3.1 Opening the furnace: 6.2.3.2 Closing the furnace: Spare parts kits 6.3.1 Common spare parts kit for all analysers 6.3.2 Spare parts kit for CS-800 Maintenance ............................151 Index ..............................152 Appendix ..........................following page...
Pos : 1.1 /0020 Ü bersc hriften/1. Ü bers chriften/1 Hi nweis e z ur Ser vic e Anl eitung @ 7\mod_1379492578032_9.doc x @ 48845 @ 11 @ 1 Notes on the Service Manual Pos : 1.2 /0025 Warnhinweise/G0002 GEF AHR Lebensg efahr durch Str omsc hlag - H oc hs pannung @ 6\mod_1373271950818_9.doc x @ 43119 @ 2 @ 1 DANGER G0001 Mortal danger from electric shock...
Pos : 3.2 /0010 ELTR A/0015 s er vice_i nstr uctions /CS- 800_Ser vice/0010 Änderungen Ü bersic ht/1005 Modul Änderungen C S-800 @ 7\mod_1379493542394_9.doc x @ 48878 @ 22222222 @ 1 Up to S/N 0766xxxxxx The CS-800 analyzers are equipped with a flow sensor with optical elements and with the corresponding board HF-3 and HF-4.
From S/N 0767xxxxxx to 0950xxxxxx The CS-800 is equipped with a flow sensor employing a pressure transducer, and with the corresponding boards HF-41 and later the HF-42. These boards also offer: • Current limit for the induction furnace (on board HF 41/42 adjustment) •...
Fig. 7: Draft 11665 (Gas flow diagram) From S/N 0951xxxxxx to 1079xxxxxx The CS-800 has an additional purging inlet on top of the combustion chamber.
Page 16
• The position of the solenoid valves is rearranged and their controlling software is modified accordingly. Therefore, a very good care has to be taken when changing the EPROMS. • The wiring has also changed. It has been modified to be used with the IR-4, and the four IR-cell rack.
Fig. 11: Draft connector pin assignmen From S/N 1080 xxxx to 1556xxxxx The CS-800 is equipped with the four IR-cell racks. All electronic boards as well as the chopper motor are different to those of the previous 3-range IR-cell. NOTICE The software version changed to the four IR-ranges! The gas saving mode is available and selectable in the PC software PLOTCS.
From S/N 1557xxxxxx The electronic unit is replaced by the board UNI 1.3 and the electronic rack is removed. The corresponding wiring diagram is 12061. The modification took place in the second half of 2004. Fig. 17: Draft 12061 (signal and control board)
new pneumatics (from end 2007) Fig. 18: Draft new pneumatics (from end 2007) Pos : 4 /0010 ELTR A/------- Seitenumbruc h ----------- @ 6\mod_1368796429449_0.doc x @ 41468 @ @ 1...
Pos : 5.1 /0020 Ü bersc hriften/1. Ü bers chriften/1 F ehl er mel dungen @ 2\mod_1259056589728_9.doc x @ 17580 @ 222 @ 1 Fault messages Pos : 5.2 /0020 Ü bersc hriften/ELTR A 1.1 Ü bersc hriften/11 C hec k base lines ELTRA @ 6\mod_1368452266990_9.doc x @ 40029 @ 3333333333333333333 @ 1 Check base lines Pos : 5.3 /0010 ELTR A/0015 s er vice_i nstr uctions /CS- 800_Ser vice/0015 Fehler/0105 M odul C hec k base lines ELTR A @ 6\mod_1377688597729_9.doc x @ 46483 @ 23333333444444 @ 1...
Fig. 4: Fluctuating gas flow Pos : 5.6 /0010 ELTR A/0015 s er vice_i nstr uctions /CS- 800_Ser vice/0015 Fehler/0115 M odul Fl uktuier ender Gasfl uss ELTRA @ 6\mod_1377688598185_9.doc x @ 46507 @ 2223333333333333333 @ 1 The flow rate is electronically controlled; therefore there should be no significant...
In case of a fault however, a control of the flow rate is no longer possible and the gas flow fluctuates. This can be caused, for example, by an excessive obstruction of the gas flow system or by a faulty component that needs to be replaced. See the next pages of this chapter and chapter Gas flow controller adjustment and jumper settings...
3.2.3 Blocked paper filters The paper filters are small so that they are frequently forgotten or neglected, although they can block the flow when they are not replaced for long. 3.2.4 Blocked lance hole over the crucible The hole in the ceramic heat shield over the crucible can be restricted or blocked. Clean by using the brass brush as described in the operation manual.
be too low oxygen pressure in the analyzer, especially at moments of high oxygen consumption, like when burning samples or purging the furnace. 3.2.11 Heavy item is placed on the plastic tube of the oxygen supply Years of experience have shown that not only bent plastic tubes can be a reason for insufficient oxygen supply, but also that a foot of the analyzer’s cabinet was placed on the tube, squeezing it and restricting the oxygen flow.
High carbon results are erratic Pos : 5.8 /0010 ELTR A/0015 s er vice_i nstr uctions /CS- 800_Ser vice/0015 Fehler/0120 M odul U nreg elmäßige Kohl enstoff Ergebnis se ELTRA @ 6\mod_1377688598908_9.doc x @ 46519 @ 32333333444444 @ 1 Even if the filter may be brushed and cleaned in ultrasonic cleaner, someday will not be able to be cleaned any more.
Fig. 9: oxygen gauge If the pressure is remarkably lower than 1.5 bar (22 psi) then – Check the pressure of the oxygen bottle. – Check the pressure regulator of the oxygen bottle – Do leakage check. See chapter Leak Checking 3.4.2 No flow If this message permanently remains in the “status window”, check the flow on the...
Fig. 12: Icon Base line 3.4.5 Integration delay The message “Integration delay” appears for as many seconds as entered in the “configuration window”. Reasonable range is something between zero and 5 seconds. When the integration delay starts, the induction generator starts at the same time.
Fig. 14: TP13 test point When the command for the generator is received ( 0.5 V DC ), the triac triggering pulses must appear on the TP13 test point. If there is no oscilloscope available you can test this with a digital voltmeter. The displayed value depends on what type of digital voltmeter you have.
Fig. 15: Air vent holes of the furnace DANGER Mortal danger from electric shock Exposed power contacts - High Voltage If the light from the filament is not visible, then remove the cover from the generator section, in order to see the tube. Fig.
10 VAC 220 VAC (between contact 2 and 3) Fig. 17: Measurements Check whether the filament is faulty, at the filament connections, only 1 Ohm must be measured. A very high or unmeasured resistance means that the filament is burnt through and the tube needs to be replaced. 3.4.7.4 Checking the primary voltage on the high voltage transformer: DANGER...
Fig. 18: Measurements - high voltage transformer Once there is voltage on the transformer the triac must be OK. A relatively low voltage on the transformer could be a result of too high current so that the HF-42 board reduces the phase angle. IF so there is most probably a faulty component in the oscillator circuit preventing it from oscillating so that the oscillator tube takes a high DC current.
Page 40
Due to high voltage in the range of 7000 Volts (amplitude), there should be no measurement on the rectifier under power. If no spare rectifier available to exchange, then test by using a scope according to the following drawing. 220 VAC Insulating transformer 220 VAC...
Page 41
Fig. 20 Draft high frequency generator (1-4-4 + 1-4-5S) As there is high voltage (5000V) and high frequency (19.2MHz) present, a measurement would be very difficult and, above all, very dangerous. Before the generator cover plate is unscrewed, the device should be turned off or, better, the power plug pulled out.
Bad combustion Pos : 5.14 /0010 ELTR A/0015 s ervic e_ins truc tions/C S-800_Ser vic e/0015 F ehl er/0135 Modul Schl ec hte Verbr ennung - Bad combusti on ELTRA @ 6\mod_1377688600953_9.doc x @ 46555 @ 2 @ 1 In most cases the reasons for bad combustion are: –...
Electronic- drivers malfunctioning Pos : 5.16 /0010 ELTR A/0015 s ervic e_ins truc tions/C S-800_Ser vic e/0015 F ehl er/0140 Modul El ektr oni k - Trei ber Fehlfunktion ELTRA @ 6\mod_1377688601518_9.doc x @ 46567 @ 324 @ 1 IC 15 and 16 are electronic driver integrated circuits, found on the mother Microcontroller board.
Page 44
Fig. 22: Partial image of the mother microcontroller board Fig. 19: Circuit diagram These IC’s act as a power switch in order to energize solenoid valves inside the analyzer or to send control signals. These drivers are designed for continuous current of 500mA each output, so that the 200mA taken by the valves are far below their specified current.
Pos : 5.17 /0020 Übersc hriften/ELTRA 1.1 Übersc hriften/11 T he furnac e does not cl os e ELTR A @ 6\mod_1368610904372_9.doc x @ 40292 @ 22 @ 1 Furnace does not close Pos : 5.18 /0010 ELTR A/0015 s ervic e_ins truc tions/C S-800_Ser vic e/0015 F ehl er/0145 Modul Ofen s chli eßt nic ht ELTRA @ 6\mod_1377688602089_9.doc x @ 46579 @ 2 @ 1 11190...
Pos : 5.19 /0020 Übersc hriften/ELTRA 1.1 Übersc hriften/11 Anal ys edauer zu lang - The anal ysis takes too long ELTR A @ 6\mod_1368613082100_9.doc x @ 40314 @ 22 @ 1 Analysis takes too long Pos : 5.20 /0010 ELTR A/0015 s ervic e_ins truc tions/C S-800_Ser vic e/0015 F ehl er/0150 Modul Anal ys edauer z u l ang ELTRA @ 6\mod_1377688604344_9.doc x @ 46591 @ 2224 @ 1...
Fig. 25: Analysis time The analysis respectively the combustion, takes normally a maximum of 100 seconds. At the beginning of combustion, the IR-cells signals start going up on the computer screen, reaching a peak maximum depending on the C resp. S concentrations in the sample.
Page 48
The base line should remain within 10mV, 50 seconds after START. If the drift is larger, then check the IR-source voltage. RF-interference on the IR-cell, or on the electronic units: If the generator's high frequency disrupts the electronic units or the infrared cell, then the base line on the PC monitor appears jerky.
Page 49
DANGER Mortal danger from electric shock Exposed power contacts - High Voltage 1… 1… 1… Fig. 20: Draft high frequency generator (1-4-4 + 1-4-5S) The causes: – The screws (1) are not tightened properly, or not all 22 of them are screwed in.
Development of dust when closing the furnace Pos : 5.22 /0010 ELTR A/0015 s ervic e_ins truc tions/C S-800_Ser vic e/0015 F ehl er/0155 Modul Entwic kl ung von Staub bei m Schli eßen des Ofens ELTRA @ 6\mod_1377688605074_9.doc x @ 46603 @ 2 @ 1 Fig.
The results are always 0.0000% or 00.0ppm: Pos : 5.26 /0010 ELTR A/0015 s ervic e_ins truc tions/C S-800_Ser vic e/0015 F ehl er/0165 Modul Ergebniss e si nd 0,0% oder 00,0ppm ELTRA @ 6\mod_1377688605901_9.doc x @ 46627 @ 2 @ 1 Pos : 5.27 /0020 Übersc hriften/ELTRA 1.1 Übersc hriften/11 C ombusti on tube - life ti me is too s hort ELTR A @ 6\mod_1368628553320_9.doc x @ 40424 @ 222 @ 1...
Page 52
– Don't use copper as accelerator. – The combustion tube is too long, i.e. longer than 145 mm. In this case, after fixing the four wing nuts (3) and (9), the combustion tube is pressed between the upper furnace closing part (17) and the lower part (10) In this case, when the combustion tube heats up during combustion, it may break.
IR-cell temperature control is out of range Pos : 5.30 /0010 ELTR A/0015 s ervic e_ins truc tions/C S-800_Ser vic e/0015 F ehl er/0175 Modul IR-C ell Temperatur Kontr olle ELTRA @ 6\mod_1377688607482_9.doc x @ 46663 @ 32 @ 1 If in the device state the graphs of IR-temperature °C or %IR-power are not in the...
Page 54
Transformer Fig. 29: • Disconnect the cables (1) from the primary side of the transformer (2) and connect the cables (1) to a 220 V light bulb (3) which is inserted to a socket. An ordinary desk lamp can serve for this purpose. In case only 110 V light bulbs are available, connect two of them in series.
If yes (O.K.), unplug the mains power plug and reconnect the power cables (1) to the transformer (2). Unplug the power plug and reconnect the cables (1). When reconnecting the power cables (1), do not let these wires approach the connectors for the cables (4) and (5).
Page 56
Fig. 21: Draft 11001-9002 (1-15-3-1S) • Check that all the screws and soldering connections have a good contact. – If the resistor 200 K (11) is a short circuit, then the fuse will blow when starting the combustion. • Check that all power conductors and copper bands are placed far enough away, from any nearby objects! The wire (1) and the copper band (2), for example, should be placed far away from the plate (3).
Breaker turns off at the end of the analysis Pos : 5.34 /0010 ELTR A/0015 s ervic e_ins truc tions/C S-800_Ser vic e/0015 F ehl er/0185 Modul Br aker turns off at the End ELTRA @ 6\mod_1377688608489_9.doc x @ 46687 @ 222 @ 1 Causes: The triac is malfunctioning.
Pos : 5.35 /0020 Übersc hriften/ELTRA 1.1 Übersc hriften/11 F urnac e - l arge c ollec tion of dust when openi ng ELTR A @ 6\mod_1368692602462_9.doc x @ 40514 @ 2222 @ 1 3.16 Furnace - large collection of dust when opening Pos : 5.36 /0010 ELTR A/0015 s ervic e_ins truc tions/C S-800_Ser vic e/0015 F ehl er/0190 Modul Large coll ecti on of dus t when opening the furnac e ELTR A @ 6\mod_1377688608847_9.doc x @ 46699 @ 3334443 @ 1...
No or low oxygen pressure Pos : 5.38 /0010 ELTR A/0015 s ervic e_ins truc tions/C S-800_Ser vic e/0015 F ehl er/0195 Modul N o or low oxyg en pr ess ure ELTRA @ 6\mod_1377688609900_9.doc x @ 46711 @ 23334443 @ 1 After pressing the START button, the following message appears: "no oxygen pressure"...
– Check if the pressure gauge for compressed air (5) shows 4 to 6 bar (60 to 90 psi). – Check if the cock (3) is properly shut, or else the O-ring (4) will not seal. Check if the O-ring (4) is in a good condition. –...
Status window message: "Waiting for stability" Pos : 5.42 /0010 ELTR A/0015 s ervic e_ins truc tions/C S-800_Ser vic e/0015 F ehl er/019505 M odul Waiting for s tability ELTRA @ 6\mod_1377688610363_9.doc x @ 46723 @ 23444 @ 1 When this message appears in the status window after starting analysis, it means that the base lines are not within the "Deviation"...
Furnace opens with delay after the end of analysis Pos : 5.44 /0010 ELTR A/0015 s ervic e_ins truc tions/C S-800_Ser vic e/0015 F ehl er/019510 M odul F urnac e opens with del ay ELTRA @ 6\mod_1377688610559_9.doc x @ 46735 @ 22444 @ 1 There can be two different reasons for this.
Page 63
Pos : 5.45 /0020 Übersc hriften/ELTRA 1.1 Übersc hriften/11 F urnac e does not open ELTR A @ 6\mod_1368703750758_9.doc x @ 41165 @ 22 @ 1 Furnace does not open Pos : 5.46 /0010 ELTR A/0015 s ervic e_ins truc tions/C S-800_Ser vic e/0015 F ehl er/019515 M odul F urnac e does not open ELTR A @ 6\mod_1377688610853_9.doc x @ 46747 @ 2 @ 1...
Pos : 5.48 /0010 ELTR A/0015 s ervic e_ins truc tions/C S-800_Ser vic e/0015 F ehl er/019520 M odul Oxygen pres sur e falls when the fur nac e opens ELTRA @ 6\mod_1377688611553_9.doc x @ 46759 @ 2 @ 1...
If the voltage on the valve (V2) is correct, then the valve itself is defective. Pos : 5.49 /0020 Übersc hriften/ELTRA 1.1 Übersc hriften/11 C ombusti on without starting anal ysis ELTR A @ 6\mod_1368706920531_9.doc x @ 41209 @ 222 @ 1 3.22...
Page 66
The power switch should be set immediately to pos. 1, in order to avoid any overheating of the generator. NOTE Always keep the furnace closed while testing availability of induction power. Fig. 39: microcontroller board The drivers of the microcontroller board are out of order, when the combustion starts without pressing F5 or clicking START and without "Analyzing"...
Page 67
Fig. 40: microcontroller board When the pulses at TP13 are present without pressing "START" and with 15 V DC at TP9, then the circuit board HF 42 is out of order. If no scope is available, the pulses can be detected by using a voltmeter. Normal values are about 0 V AC without pulses, and about 1 V AC with pulses.
Crucible cracks or melts Pos : 5.52 /0010 ELTR A/0015 s ervic e_ins truc tions/C S-800_Ser vic e/0015 F ehl er/019530 M odul Cr ucibl e cr ac ks or melts ELTRA @ 6\mod_1377688613120_9.doc x @ 46783 @ @ 1 One reason could be that: The quality of the crucible is bad.
Analysis cycle stops before the peak begins Pos : 5.58 /0010 ELTR A/0015 s ervic e_ins truc tions/C S-800_Ser vic e/0015 F ehl er/019545 M odul anal ysis c ycl e stops befor e the peak begins ELTRA @ 6\mod_1377688613716_9.doc x @ 46819 @ @ 1...
If so, check the bypass valve V4. Pos : 5.63 /0020 Übersc hriften/ELTRA 1.1 Übersc hriften/11 IR signal does not c ome down to the base line at the end of anal ysis @ 7\mod_1381404509510_9.doc x @ 49850 @ 31222 @ 1 3.29...
3.29.1 Too long combustion of a very low carbon sample Fig. 24: Programm screenshot (graph example 1) It appears like a long tailing at the end of the peak. It appears in the low carbon range only and the distance to the base line is in the range of 10 to 50mV. Depending on the blanks of the crucibles and on the oxygen purity, the level of the tailing can be as high as 100mV.
3.29.2 Copper oxide has a bad quality Fig. 25: Programm screenshot (graph example 2) Memory effects of the copper oxide can cause tailing of the low carbon range in the level of 10 to 50 mV. However the tailing level does not grow due to faulty copper oxide.
3.29.4 Flow stops during analysis 101B Fig. 26: Programm screenshot (graph example 3) In such case all base lines remain high, keeping their level which they had at the moment the flow stops. Therefore the voltages of the flat continuation of the base lines can be high, depending on the voltage level of each peak at the moment the flow stopped.
Pos : 7.1 /0020 Ü bersc hriften/ELTR A 1. Ü berschriften/1 Adj ustments, tests and wor ki ng instruc tions ELTRA @ 6\mod_1368775845833_9.doc x @ 41360 @ 1 @ 1 Adjustments, tests and working instructions Pos : 7.2 /0020 Ü bersc hriften/ELTR A 1.1 Ü bersc hriften/11 Gas fl ow c ontroll er adj ustment and j umper s etti ngs ELTR A @ 6\mod_1368776355883_9.doc x @ 41371 @ 233332 @ 1 Gas flow controller adjustment and jumper settings Pos : 7.3 /0010 ELTR A/0015 s er vice_i nstr uctions /CS- 800_Ser vice/0020 Einstellungen, T ests und F unkti ons besc hrei bung/0205 Modul Gas flow contr oller adj ustment and j umper s ettings @ 6\mod_1377688625026_9.doc x @ 46855 @ 31 @ 1...
Pos : 7.4 /0020 Ü bersc hriften/ELTR A 1.1.1. Ü berschriften/111 Jumper settings ELTRA @ 6\mod_1368777217783_9.doc x @ 41393 @ 323323 @ 1 4.1.1 Jumper settings 102B Pos : 7.5 /0010 ELTR A/0015 s er vice_i nstr uctions /CS- 800_Ser vice/0020 Einstellungen, T ests und F unkti ons besc hrei bung/0210 Modul J umper s ettings - Gas fl ow C ontroller @ 6\mod_1377688625403_9.doc x @ 46867 @ 3 @ 1 Abb.
Pos : 7.6 /0020 Ü bersc hriften/ELTR A 1.1.1. Ü berschriften/111 Gas fl ow r egulation ELTR A @ 6\mod_1368781633747_9.doc x @ 41415 @ 322332 @ 1 4.1.2 Gas flow regulation 103B Pos : 7.7 /0010 ELTR A/0015 s er vice_i nstr uctions /CS- 800_Ser vice/0020 Einstellungen, T ests und F unkti ons besc hrei bung/0215 Modul Gas Fl ow - D etails @ 6\mod_1377688625763_9.doc x @ 46879 @ 343 @ 1 UNI Board Fig.
Page 77
Abb. 45: Front panel The flow controller mainly consists of the following components: – The flow sensor (1) – The flow regulating valve (V6) – The flow control board HF 42 The flow sensor measures the gas flow, and converts its flow rate into an electrical signal (DC voltage).
Fig. 28: Flow sensor Pos : 7.8 /0020 Ü bersc hriften/ELTR A 1.1 Ü bersc hriften/11 Infrar ed bas e line adj ustments ELTRA @ 6\mod_1369121936836_9.doc x @ 41542 @ 2232 @ 1 Infrared base line adjustments Pos : 7.9 /0020 Ü...
Page 79
Pos : 7.10 /0010 ELTR A/0015 s ervic e_ins truc tions/C S-800_Ser vic e/0020 Eins tell ungen, Tests und F unktions beschr eibung/0225 M odul Infrared baseli ne adjustment @ 6\mod_1377688626500_9.doc x @ 46903 @ 222344344 @ 1 Eltra has changed the IR cell design. This modification makes the cells less susceptible to contamination problems and makes the calibration more stable.
4.2.2 General test points 105B 4.2.2.1 Supply voltages and thermostatic control circuit TP 1: TP 7: IR source voltage TP 12: + 5V supply TP 13: Heating transistors control voltage TP 26: IR temperature actual value TP 27: +24V supply TP 28: IR temperature set point 4.2.2.2...
Page 82
Fig. 29: Mainboard (Cut-out) IR-Cell – general test points...
TP 24: Output section [3] TP 23: Output section [4] TP 20: Analog GND NOTICE Use a very thin screwdriver. With a big screwdriver, the trimmer can be mechanically destroyed before you feel it. The screwdriver should never get in touch with the hole resp.
Install the IR module into the analyzer. . Allow sufficient time for the base lines to stabilize. NOTICE It is necessary to turn slowly in order for the IR sources to stabilize their temperature with increasing IR source voltage. • Adjust the other cells of the module (if any) at 8V baselines or 5V to 6V in case of clean golden paths.
4.2.7 IR source voltage setting Fig. 31: IR source voltage setting The jumper JP2 is for setting voltages of the IR sources. All available sources in an IR cell are connected in series. Normally the voltage of each source is 1.75 VDC. The total voltage of all available cells can be measured on test point TP7 against TP1=GND.
4.2.8 IR cell temperature control 111B Fig. 32: IR cell temperature control TP28 is the test point for measuring the set point of the IR temperature against TP1=GND. At 3000 mV the temperature is 27°C growing with 10 mv /°C. 4.2.8.1 Examples 3.10V ...
The temperature is set by turning the pot and measuring the set value on The actual value of the temperature is measured on TP26. Caution: After changing the temperature set point, it takes some time until the actual temperature value is stabilized at the new set level. This is the trimmer in the IR cell module for adjusting the cell temperature.
Combustion coil Muffler Pos : 7.17 /0020 Übersc hriften/ELTRA 1.1.1. Ü bersc hriften/111 Pneumatic s adj usti ng for corr ect fur nace movement @ 6\mod_1369230967371_9.doc x @ 41855 @ 324333 @ 1 4.3.1 Pneumatics adjusting for correct furnace movement 112B Pos : 7.18 /0010 ELTR A/0015 s ervic e_ins truc tions/C S-800_Ser vic e/0020 Eins tell ungen, Tests und F unktions beschr eibung/0245 M odul Pneumatic adj ustment for c orrec t furnac e movement @ 6\mod_1377688629923_9.doc x @ 46951 @ 244244434433444 @ 1...
According to our experience, the combustion coils of our analyzers don’t need replacement, even after years of operation. Pos : 7.19 /0020 Übersc hriften/ELTRA 1.1 Übersc hriften/11 Press ure regul ati on ELTR A @ 6\mod_1369298860002_9.doc x @ 41930 @ 32233 @ 1 Pressure regulation Pos : 7.20 /0010 ELTR A/0015 s ervic e_ins truc tions/C S-800_Ser vic e/0020 Eins tell ungen, Tests und F unktions beschr eibung/0255 M odul Pres sur e R egulation @ 6\mod_1377688630939_9.doc x @ 46975 @ 2424 @ 1...
Fig. 49: Front side Purge pressure Fig. 50: PR2...
Page 95
PR2 is the pressure regulator for the purging system of the IR-cell and it is situated inside the analyzer, just behind the left-side of the panel. It is positioned horizontally. – The oxygen pressure at the outlet of the PR2 should be 0.35 bar (5psi). The pressure can be checked by applying a T-piece on the outlet of the pressure regulator, and connecting a pressure gauge (1), as indicated above.
Pos : 7.24 /0010 ELTR A/0015 s ervic e_ins truc tions/C S-800_Ser vic e/0020 Eins tell ungen, Tests und F unktions beschr eibung/0270 M odul Linearisati on @ 6\mod_1377688631703_9.doc x @ 47011 @ 42244 @ 1 See software helps. Pos : 7.25 /0020 Übersc hriften/ELTRA 1.1 Übersc hriften/11 Bal ance pr ogramming ELTR A @ 6\mod_1369384797393_9.doc x @ 42074 @ 2422242 @ 1 Balance programming Pos : 7.26 /0010 ELTR A/0015 s ervic e_ins truc tions/C S-800_Ser vic e/0020 Eins tell ungen, Tests und F unktions beschr eibung/0275 M odul Balance pr ogramming @ 6\mod_1377688631889_9.doc x @ 47023 @ 4 @ 1...
Page 98
3BAdjustments, tests and working instructions Fig. 53: Jumper setting – Current control: The current control provides a safer and more accurate control of the furnace intensity. • Set the jumper J2 to B1-B2. • Connect a digital multi meter on TP1 (TP15 is GND). Adjust only, while the furnace is active: •...
Page 99
3BAdjustments, tests and working instructions Fig. 55: Diagramm generator power vs. setpoint Pos : 8 /0010 ELTR A/------- Seitenumbruc h ----------- @ 6\mod_1368796429449_0.doc x @ 41468 @ 24222222 @ 1...
Pos : 9.1 /0020 Ü bersc hriften/ELTR A 1. Ü berschriften/1 Ser vic e ELTRA @ 6\mod_1369640568215_9.doc x @ 42168 @ 2222221 @ 1 5 Service Pos : 9.2 /0020 Ü bersc hriften/ELTR A 1.1 Ü bersc hriften/11 IR-paths, cleaning and r epl acing ELTRA @ 6\mod_1369640666897_9.doc x @ 42179 @ 24 @ 1 IR-paths, cleaning and replacing Pos : 9.3 /0010 ELTR A/0015 s er vice_i nstr uctions /CS- 800_Ser vice/0025 Ser vice/0305 M odul IR-paths, cleani ng and r epl acing @ 6\mod_1377688640144_9.doc x @ 47083 @ 32222211 @ 1...
Service Fig. 56: Infrared cell By reinstalling the infrared cell unit in the analyzer, you must pay attention to the color markings of the tubes and their connections. black marked tube connection yellow marked tube connection red marked tube connection 5.1.1 Available infrared paths 113B...
Service Pos : 9.4 /0020 Ü bersc hriften/ELTR A 1.1 Ü bersc hriften/11 Infrar ed el ectr onics ELTR A @ 6\mod_1369640821864_9.doc x @ 42223 @ 2432 @ 1 Infrared electronics Pos : 9.5 /0010 ELTR A/0015 s er vice_i nstr uctions /CS- 800_Ser vice/0025 Ser vice/0315 M odul Infrar ed el ectronics @ 6\mod_1377688640668_9.doc x @ 47107 @ 12222 @ 1 See chapter Infrared power supply and chapter...
Service Pos : 9.6 /0020 Ü bersc hriften/ELTR A 1.1 Ü bersc hriften/11 General test poi nts ELTR A @ 6\mod_1373448456679_9.doc x @ 43494 @ 4222 @ 1 General test points Pos : 9.7 /0010 ELTR A/0015 s er vice_i nstr uctions /CS- 800_Ser vice/0025 Ser vice/039515 Modul General Test Poi nts @ 7\mod_1381924872474_9.doc x @ 49938 @ 33221223 @ 1 Fig.
TP 21: -12V Pos : 9.8 /0020 Ü bersc hriften/ELTR A 1.1 Ü bersc hriften/11 IR-channels tes t poi nts ELTRA @ 6\mod_1373448523586_9.doc x @ 43505 @ 2222 @ 1 IR-Channels test points Pos : 9.9 /0010 ELTR A/0015 s er vice_i nstr uctions /CS- 800_Ser vice/0025 Ser vice/039520 Modul IR cannel tes t poi nts @ 7\mod_1381925324500_9.doc x @ 49949 @ 221333 @ 1 Fig.
Infrared cell board IRC1.x sections assignment. Pos : 9.10 /0020 Übersc hriften/ELTRA 1.1 Übersc hriften/11 Infr ared power s uppl y ELTRA @ 6\mod_1369640877414_9.doc x @ 42245 @ 2234222333 @ 1 Infrared power supply Pos : 9.11 /0010 ELTR A/0015 s ervic e_ins truc tions/C S-800_Ser vic e/0025 Ser vic e/0320 M odul Infr ared power s uppl y @ 6\mod_1377688640960_9.doc x @ 47119 @ 21333 @ 1 The IRC module is powered by a 24VDC power supply which is specified with 5A.
Service Pos : 9.12 /0020 Übersc hriften/ELTRA 1.1 Übersc hriften/11 IR-s ource ELTRA @ 6\mod_1369640743211_9.doc x @ 42201 @ 2222232444 @ 1 IR-source Pos : 9.13 /0010 ELTR A/0015 s ervic e_ins truc tions/C S-800_Ser vic e/0025 Ser vic e/039525 M odul IR-source @ 7\mod_1382344471044_9.doc x @ 49968 @ 2313333 @ 1 Fig.
Page 107
Service Correct Wrong 3-2-1 Fig. 38: IR sources in their housing If the position of a source is wrong, the radiation reaching the IR sensor is lower than when the source is in the right position. It may happen that after transportation due to a strong shock that the base lines are not in the right level due to the sources being not in the right position any more.
Service • Release the push button (43). • Close the furnace, squeeze the tube (6) tightly and press and hold the button (43). If the pressure on gauge (12) remains constant, then the furnace has to be checked for leaks. If the pressure shown on gauge (12) drops then the leakage has to be found somewhere along the furnace inlet system.
If the voltages on V2, V3, or V4 are wrong replace the electronic drivers. Pos : 9.16 /0020 Übersc hriften/ELTRA 1.1 Übersc hriften/11 Sol enoi d val ves cleaning ELTR A @ 6\mod_1369643206541_9.doc x @ 42465 @ 222223222 @ 1 Solenoid valves cleaning Pos : 9.17 /0010 ELTR A/0015 s ervic e_ins truc tions/C S-800_Ser vic e/0025 Ser vic e/0365 M odul Sol enoid val ves cleaning @ 6\mod_1377688643571_9.doc x @ 47227 @ 2332333444 @ 1...
• Adjust the flow sensor. Pos : 9.20 /0020 Übersc hriften/ELTRA 1.1 Übersc hriften/11 F urnac e pneumatic s - additi onal s afety featur es (opti onal) ELTR A @ 6\mod_1369643455549_9.doc x @ 42531 @ 222222232 @ 1 5.10 Furnace pneumatics - additional safety features (optional) Pos : 9.21 /0010 ELTR A/0015 s ervic e_ins truc tions/C S-800_Ser vic e/0025 Ser vic e/0380 M odul F urnac e pneumatic s - additi onal s afety features (opti onal) @ 6\mod_1377688644179_9.doc x @ 47263 @ 323243344 @ 1...
Service Pos : 9.22 /0020 Übersc hriften/ELTRA 1.1 Übersc hriften/11 H alog en trap - i nstalling ELTR A @ 6\mod_1369643531895_9.doc x @ 42553 @ 2222223223 @ 1 5.11 Halogen trap - installing Pos : 9.23 /0010 ELTR A/0015 s ervic e_ins truc tions/C S-800_Ser vic e/0025 Ser vic e/0385 M odul H alogen trap - i nstalling @ 6\mod_1377688644390_9.doc x @ 47275 @ 323343344 @ 1...
Page 116
Service TP13 TP14 TP10 TP11 HF41: External On board TP3 TP4 flow enable flow enable On board Continuous power adjust. TP12 flow TP15 TP16 TP17 TP18 TP19 TP20 HF41 12008 Fig. 64: HF41 board jumper setting for gas saving mode If the analyzer has the HF42 board: Set the jumper setting to connect A1 with A2 on J3 TP13...
Miscellaneous 75190 Washer 6.1.1 Infrared cells 120B 05060 Reflector 05067 IR-path for high carbon 05150 IR-path connector with window source side 05160 IR-path connector with window detector side 05170 Gas conditioner 05244 IR 4 chopper 05260 IR-path tube ( advice the length ) 05270 Threaded rod ( advice the length ) 05275...
Page 119
Miscellaneous 77510 Heaters for IR-cell...
After a short delay by the software, V2, V3 and V4 will be de-energized. The closed furnace condition is now reached. Pos : 11.6 /0020 Übersc hriften/ELTRA 1.1 Übersc hriften/11 Spare parts kits @ 7\mod_1382427709237_9.doc x @ 50008 @ 2232232 @ 1 Spare parts kits Pos : 11.7 /0010 ELTR A/0015 s ervic e_ins truc tions/C S-800_Ser vic e/0030 Sonstig es/3015 Modul Spare parts ki ts @ 7\mod_1382427765325_9.doc x @ 50019 @ 321222333333333 @ 1...
Page 148
Miscellaneous 05278 Infrared preamplifier for 3 pins 05279 Infrared preamplifier for 4 pins 06032 Infrared detector with CO2 filter 06034 Infrared detector with SO2 filter 06058 IR Temperature control board TH61 06421 Infrared power supply assembly IR13 06630 IR Temperature control board TH62 06733 IRC board for 4 channels 11035...
Page 149
Washer 77140 HF-filter 250 V 78010 Mains power switch 90290 Copper oxide 100g 92610 Tube of grease 6.3.2 Spare parts kit for CS-800 131B description 11105 Metal dust filter 11115 Plastic tube 75mm 11120 Paper filter holder 11180 Dust filter cartridge...
Page 150
Miscellaneous 12005 Gasflow and furnace control board HF3 12016 Gasflow and furnace control board HF42 12080 Rectifier 13067 Combustion coil 13175 Insulator 13260 High voltage filter 13262 Capacitor 700 pF/40KV 13270 Resistor 14045 Cleaning brush for combustion tube 14072 Ceramic heat shield for brush 14080 Cleaning mechanism rod 14130...
Index Analysis takes too long 46 Analysis time 47 %IR-power 53 Analyzing 35 Averages 27 (long C path) 84 (short C path) 84 B1-B2 75 (short S path) 84 B3-B4 75 (single C path) 84 b7-10-1 109 (single S path) 84 Bad combustion 42 Balance programming 97 °...
Page 153
combustion tubes 52 connector pin assignmen 20 IC 15 43 flow sensor and HF-3 9 IC 16 43 high frequency generator (1-4-4 + 1-4-5S) 41, Icon Base line 35 IR-45 power supply 23 configuration window 34 IR-sources 24 Induction generator control 97 new pneumatics (from end 2007) 26 Infrared base line adjustments 78 VG-96 connector 21...
Page 154
Miscellaneous 117 service manual Modification overview 9 notes 8 Signal and control 14 signal and control board 19 12061 25 new pneumatics (from end 2007) 26 single carbon range 84 No combustion 33 Solenoid valves cleaning 112 No flow 34 No or low oxygen pressure 59 Source voltage No oxygen pressure 33...
Page 158
Copyright ® Copyright by Eltra GmbH Haan, Retsch-Allee 1-5 D-42781 Haan Federal Republic of Germany...
Need help?
Do you have a question about the CS-800 and is the answer not in the manual?
Questions and answers