ELTRA CS-800 Service Manual (322 pages)
Double Dual Range Carbon/Sulfur Determinator
Brand:
ELTRA
| Category:
Laboratory Equipment
| Size: 11 MB
Table of Contents
Section 1
3
Service Manual
2
Table of Contents
3
0.1 Overview of the CS-800 Modifications
6
Faults
23
Gas Flow too High or too Low
24
The Gas Flow Rate Fluctuates
26
No Combustion
27
Bad Combustion
34
Electronic Drivers Malfunctioning
35
The Furnace Does Not Close
36
The Analysis Takes too Long
37
Development of Dust When Closing the Furnace
41
Erratic Results
42
The Results Are Always 0.0000% or 00.0 Ppm
43
Beeper Responding
44
The Life Time of the Combustion Tube Is too Short
45
The Fuse Burns out at the Beginning of Combustion
48
The Fuse Blows at the End of the Analysis
51
Large Collection of Dust When Opening the Furnace
52
No or Low Oxygen Pressure
53
Saved Data Are Lost
55
Noise or Drift of the Infrared Baseline
56
Dust in the Oxygen Carrying Tube When the Furnace Opens
57
The Word Eltra Appears on the Display
58
The Oxygen Pressure Falls When the Furnace Opens
59
Combustion Without Pressing " START
60
The Crucible Cracks or Melts
62
The Analysis Doesn't Start
63
Vibrations on Setting
64
1.28 the Analysis Cycle Stops Only 25 Seconds after Start
65
1.29 the Chopper Motor Doesn`t Rotate
66
High Carbon Results Are Erretic
67
Pressure Switch
72
Infrared Zero-Baseline, Control and Adjustment
73
IR-Source Voltage Setting
76
Temperature Regulation of the IR-Cell
77
Pneumatics
82
Check the Following
84
Pressure Regulation
85
Closing Cone, Adjustment
88
Linearisation
89
Programming the Balance
91
Induction Generator Control
93
Current Control
93
Diagnostic Printous
95
Setting Check
96
Service
98
IR-Paths, Cleaning and Replacing
98
Infrared Electronics
102
Infrared Power Supply
103
Infrared Temperature Regulation
106
Leak Checking
108
Replacing Eproms
111
Clearing Memory
112
Electronic Reset
113
Coding IR-Ranges
114
Cleaning the Solenoid Valves
117
Replacing the Flow Sensor
118
Installing an AC-Power Stabiliser
119
Additional Safety Features for Furnace Pneumatics (Optional)
120
Installing a Halogen Trap
121
Rubber Ring
121
Gas Saving Mode
122
3.19 3.19 Upgrade for Copper Analysis Without Accelerators
123
Upgrading of Analysers with Electronic Rack for Using the New Windows Software
125
3.21 Preventive Maintenance Checklist
126
Miscellaneous
127
Ordering Numbers
127
Valve Wiring
139
Cable Connection
149
Starting through External Key
156
Spare Parts Kits
158
Section 2
164
0.1 Overview of the CS-800 Modifications
167
Faults
184
Gas Flow too High or too Low
185
The Gas Flow Rate Fluctuates
187
No Combustion
188
Bad Combustion
195
Electronic Drivers Malfunctioning
196
The Furnace Does Not Close
197
The Analysis Takes too Long
198
Development of Dust When Closing the Furnace
202
Erratic Results
203
The Results Are Always 0.0000% or 00.0 Ppm
204
Beeper Responding
205
The Fuse Burns out at the Beginning of Combustion
209
The Fuse Blows at the End of the Analysis
212
Large Collection of Dust When Opening the Furnace
213
No or Low Oxygen Pressure
214
Saved Data Are Lost
216
Noise or Drift of the Infrared Baseline
217
Dust in the Oxygen Carrying Tube When the Furnace Opens
218
The Word Eltra Appears on the Display
219
The Oxygen Pressure Falls When the Furnace Opens
220
Combustion Without Pressing " START
221
The Crucible Cracks or Melts
223
The Analysis Doesn't Start
224
Vibrations on Setting
225
1.28 the Analysis Cycle Stops Only 25 Seconds after Start
226
1.29 the Chopper Motor Doesn`t Rotate
227
High Carbon Results Are Erretic
228
Adjustment, Test and Function Explanations
229
Gas Flow Controller Adjustment and Jumper Setting
229
Pressure Switch
233
Infrared Zero-Baseline, Control and Adjustment
234
IR-Source Voltage Setting
237
Temperature Regulation of the IR-Cell
238
Pneumatics
243
Pressure Regulation
246
Closing Cone, Adjustment
249
Linearisation
250
Programming the Balance
252
Induction Generator Control
254
Setting Check
257
IR-Paths, Cleaning and Replacing
259
Service
259
IR-Source
262
Infrared Electronics
263
Infrared Power Supply
264
Infrared Temperature Regulation
267
Leak Checking
269
Replacing Eproms
272
Clearing Memory
273
Electronic Reset
274
Coding IR-Ranges
275
Cleaning the Solenoid Valves
278
Replacing the Flow Sensor
279
Installing an AC-Power Stabiliser
280
Additional Safety Features for Furnace Pneumatics (Optional)
281
Installing a Halogen Trap
282
Gas Saving Mode
283
3.19 3.19 Upgrade for Copper Analysis Without Accelerators
284
Upgrading of Analysers with Electronic Rack for Using the New Windows Software
286
3.21 Preventive Maintenance Checklist
287
Miscellaneous
288
Ordering Numbers
288
Starting through External Key
317
Spare Parts Kits
319
Advertisement
ELTRA CS-800 Manual (158 pages)
Carbon/Sulfur Determinator
Brand:
ELTRA
| Category:
Industrial Equipment
| Size: 22 MB
Table of Contents
Table of Contents
4
Notes on the Service Manual
8
Modification Overview
9
Up to S/N 0766Xxxxxx
9
From S/N 0767Xxxxxx to 0950Xxxxxx
13
From S/N 0951Xxxxxx to 1079Xxxxxx
15
From S/N 1080 XXXX to 1556Xxxxx
20
From S/N 1557Xxxxxx
25
New Pneumatics (from End 2007)
26
Fault Messages
27
Check Base Lines
27
Fluctuating Gas Flow
28
Blocked Metal Dust Filter
29
Blocked Chemicals
29
Blocked Paper Filters
30
Blocked Lance Hole over the Crucible
30
Blocked Furnace Outlet
30
Too Low Oxygen Pressure
30
Outlet Tap of the Oxygen Regulator Is Not Properly Open
30
Contaminated or Corroded Solenoid Valve
30
Faulty Oxygen Regulator
30
Bent Plastic Tube of Oxygen Supply
30
Heavy Item Is Placed on the Plastic Tube of the Oxygen Supply
31
Twisted Plastic Tube Inside the Analyzer
31
Oxygen Supply Is Interchanged with the Compressed Air Supply
31
Blocked Dust Filter Cartridge
31
Blocked Fittings
31
Blocked Flow Sensor Assembly
31
Electronic Problem
31
High Carbon Results Are Erratic
32
No Combustion
33
No Oxygen Pressure
33
No Flow
34
Pre-Purging
34
Waiting for Stability
34
Integration Delay
35
Analyzing
35
The Command from the Microcontroller to Start the Generator Is Not Received
35
Measurements on TP9
35
Checking the Generator Tube
36
If the Filament of the Generator Tube Doesn't Glow
37
Checking the Primary Voltage on the High Voltage Transformer
38
The Rectifier Could also be Faulty
39
Loud Buzzing Sound When the Generator Turns on
42
Bad Combustion
42
Electronic- Drivers Malfunctioning
43
Furnace Does Not Close
45
Analysis Takes too Long
46
Additional Reasons for Excessive Analysis Duration Could be
47
One of the IR-Sources Is Unstable
47
Development of Dust When Closing the Furnace
50
Irregular Results
50
The Results Are Always 0.0000% or 00.0Ppm
51
Combustion Tube - Life Time Is too Short
51
IR-Cell Temperature Control Is out of Range
53
Breaker Turns off at the Beginning of Combustion
53
Further Causes Could be Found in the Oscillator Circuit
55
Breaker Turns off at the End of the Analysis
57
Furnace - Large Collection of Dust When Opening
58
No or Low Oxygen Pressure
59
Possible Faults on the Front Panel of the Analyzer
59
On the Rear-Side of the Analyzer
60
Inside the Analyzer
60
3.17.3.1 Oxygen Solenoid Valve (V5) May be Defective
60
3.17.3.2 Outflow Valve Does Not Close
61
3.17.3.3 if the Voltage on the Outflow Valve Is Constantly at 24 V, then
61
Other Faults
61
No Oxygen Flow
61
Status Window Message: "Waiting for Stability
61
Reasons for Instability
62
3.19.1.1 too Low Deviation Value
62
3.19.1.2 too High Noise
62
3.19.1.3 too High Drift
62
Furnace Opens with Delay after the End of Analysis
62
Oxygen Pressure Falls When the Furnace Opens
64
Combustion Without Starting Analysis
65
Crucible Cracks or Melts
68
Analysis Doesn't Start / no Peak on Screen
68
Vibrations on Setting 2
68
Analysis Cycle Stops before the Peak Begins
69
Chopper Motor Doesn't Rotate
70
Sample Burns but no Peak on the Screen
70
IR Signal Does Not Come down to the Base Line at the End of Analysis
70
Too Long Combustion of a very Low Carbon Sample
71
Copper Oxide Has a Bad Quality
72
Path Window Is Not Sealed or It Is Broken
72
Flow Stops During Analysis
73
Adjustments, Tests and Working Instructions
74
Gas Flow Controller Adjustment and Jumper Settings
74
Jumper Settings
75
Gas Flow Regulation
76
Infrared Base Line Adjustments
78
Infrared Cell Modification - General Information
78
General Test Points
81
Supply Voltages and Thermostatic Control Circuit
81
Supply Voltages for IR Signal Processing Circuit
81
Test Points IR-Ranges
83
Supply Voltages for IR Signal Processing Circuit
83
Outputs (Base Lines about 8VDC. for Golden Paths Adjust 6VDC)
83
Base Line Adjustments
84
Infrared Cell Board Irc1.X Sections Assignment
84
Explanations
84
Example
84
Voltage of a Base Line Is too High (Higher than 9.5V)
85
Source Voltage - Reducing
85
Drift too Large
86
IR Source Voltage Setting
87
IR Cell Temperature Control
88
Examples
88
Modifications for Higher Sensitivity (Gain) in the Low Sulphur Range
89
IRC Board and Cross-Talk
89
Pneumatics
90
Pneumatics Adjusting for Correct Furnace Movement
92
Pressure Regulation
93
Purge Pressure
94
Closing Cone Adjustment
96
Linearisation
97
Balance Programming
97
Induction Generator Control
97
Service
100
IR-Paths, Cleaning and Replacing
100
Available Infrared Paths
101
Infrared Electronics
102
General Test Points
103
Supply Voltages and Thermostatic Control Circuit
103
Supply Voltages for IR Signal Processing Circuit
103
IR-Channels Test Points
104
Supply Voltages for IR Signal Processing Circuit
104
Outputs (Base Lines about 8VDC. Less no Problem)
104
Infrared Power Supply
105
IR-Source
106
Leak Checking
108
Furnace Inlet System - Leaks
111
Leaks in the Furnace
111
Leaks in the Furnace Outlet System
111
Leaks Inside the Analyzer
111
Solenoid Valves Cleaning
112
Flow Sensor - Replacing / Cleaning
113
Flow Sensor Replacing
113
Furnace Pneumatics - Additional Safety Features (Optional)
114
Halogen Trap - Installing
115
Gas Saving Mode
115
Miscellaneous
117
Ordering Numbers
117
Infrared Cells
118
Front Side
120
Right Side
122
Pneumatics
124
Oscillating Circuit
126
Furnace
128
Furnace Cleaning
130
Wiring (I)
133
Uni 1
143
Gas Flow System
144
Valve Controlling
146
Opening the Furnace
147
Closing the Furnace
147
Spare Parts Kits
147
Common Spare Parts Kit for All Analysers
147
Maintenance
151
Index
152
ELTRA CS-800 Manual (84 pages)
Carbon and Sulfur Analyzer
Brand:
ELTRA
| Category:
Measuring Instruments
| Size: 4 MB
Table of Contents
Table of Contents
3
Contact Information
6
Notes on the Manual
6
Explanations of the Safety Instructions
7
General Safety Instructions
9
Explanation of Signs and Symbols
10
Packaging, Transport and Installation
11
Packaging
11
Transport
11
Intended Use
11
Conditions for the Installation Site
11
Type Plate Description
12
Installation
13
Setting Check up
13
Front Panel Illustration
14
Mains Power Connections
15
Data Interface
16
Gas Connections
17
Gas Purification Furnance Connecting (Optional)
18
Autoloader (Optional)
19
Analysis
20
Working Procedure
20
Analysis Example
22
Combustion Peak
22
Combustion Double Peak
22
Work Breaks
23
Crucibles Preheating, Optional
23
Operating the Pre-Heating Furnace (Optional)
24
Applications
24
Maintenance
29
General Information
29
Maintenance Every 100 Analyses
29
Maintenance Every 500 Analyses
29
Maintenance Every 1000 Analyses
29
Maintenance Every 2000 Analyses
29
Reagent Tubes - Removing and Installing
30
Reagent Tubes Filling
32
Chemicals
32
Reagent Tube Filling
34
Oxygen Purification Furnace Quartz Tube Filling (Optional)
37
Halogen Trap Filling (Optional)
38
Paper Filter Changing
40
O-Rings Replacement
41
Replacing the O-Rings 11 and 12 for Combustion Tube
42
Replacing the O-Ring 13 for Lower Furnace Lock
42
Replacing the O-Rings for Furnace Seal
43
Furnace Cleaning Brush - Replacing
44
Dust Trap Cleaning
45
Fast Filter Cleaning
46
Thorough Cleaning
47
Oxygen Purification Furnace Refilling
48
Combustion Tube Replacement
49
Generator Tube Replacing
51
Combustion Coil Replacing
51
Pedestal Removing
51
Gas Leaks Checking
52
Leaks in the Furnace Inlet System
53
Leaks in the Furnace
53
Leaks in the Furnace Outlet System
54
Function Description
55
System Overview
55
Measuring Principle
55
Gas Flow System
56
Infrared Cell
58
Furnace
59
Miscellaneous
60
Ordering Numbers
60
Chopper Motor
60
Infrared Cells
61
Front Side
62
Right Side
64
Pneumatics
66
Oscillating Circuit
68
Furnace
70
Furnace Cleaning Mechanism
72
Gas Purification Furnace, Optional
74
Packing
75
Pre-Installation Guide
77
Approved Methodologies to Which Eltra Instruments Conform
79
Inorganic Materials (Metals)
79
Organic Materials (Oil, Coal, Foodstuffs)
80
Disposal
81
Index
82
Advertisement
ELTRA CS-800 Operating Manual (28 pages)
Autoloader
Brand:
ELTRA
| Category:
Measuring Instruments
| Size: 2 MB
Table of Contents
Table of Contents
4
Operating the Machine
5
General Information
5
Loader Models
5
Space Requirements
5
Connecting and Disconnecting Plastic Tubes
6
Mechanical Connection to the Analyzer
7
Chain Motor Belt Adjustment
8
Chain Adjustment
8
Finger Adjustment
10
Arm Installation
11
Electrical and Pneumatic Connections
12
Up/Down Adjustment of the Arm
15
Single Step Operation for Mechanical Adjustments
15
Vertical Laser
15
Horizontal Laser
16
Single Step Cycles
18
Finger Position over the Pedestal
18
Finger Up/Down Speed
18
Crucible Position on the Platform
19
Height of the Platform
19
Repeating the Single Step Cycle
20
Non Stop Cycles
20
Error Messages
20
Final Installation after Adjusting
21
Maintenance
21
Analysis Mode
21
Crucibles Disposal
22
Index
23
Advertisement
Related Products
ELTRA CS-2000
ELTRA CS-580A
ELTRA CSi/d
ELTRA CS-i
ELTRA CS-d
ELTRA CHS-580A
ELTRA ELEMENTRAC CS-i
ELTRA ELEMENTRAC ONHp2
ELTRA OH-900
ELTRA ON-900
ELTRA Categories
Measuring Instruments
Industrial Equipment
Laboratory Equipment
Stereo System
Radio
More ELTRA Manuals
Login
Sign In
OR
Sign in with Facebook
Sign in with Google
Upload manual
Upload from disk
Upload from URL