Zeroing of the
measure. It must
be executed in
accordance with
all those specified
Starting
the Process
Alignment of the
measure with the
real value of
consistency.
Take a pulp sample to provide the
laboratory and enter the menu "init
setup -> Freeze" confirming the
acquisition of the sample, appear "***"
flashing on the display
Put in the real value
of consistency on the
%CONST menu
VALCOM® - www.valcom.it - valcom@terranova-instruments.com
a TERRANOVA® Srl brand - www.terranova-instruments.com
Factory & Sales: Via Gramsci, 1 – 26827 Terranova Passerini (LO) - Italy
Legal Office: Via Rosso Medardo, 16 – 20159 Milano (MI) - Italy
TRASMETTITORE ELETTRONICO DI CONSISTENZA SMART
ELECTRONIC CONSISTENCY TRANSMITTER SMART
Sistema Gestione Qualità, Ambiente e Sicurezza
Quality Management System, Environment and Safety
INIT SETUP
MOUNT
in point 2.
%CONSIST
dough
CALIBRAT
(facoltativo)
Linearization of the
measure curve through
the insertion of
calibration point, which
are obtained by real
consistency value of the
sample taken on-line
INSERTION OF
POINT
Take a pulp sample and at the
same time write the value read
from display
Consistency value
read from display:
%CS_READ
Point of calibration
Pi (%CS_READ, %CS_CALIBR)
MOD. MN
Ph: +39 0377 911066 – Fax: +39 0377 919156
VAT IT07848810151
MNT7C01
Rev. 19.01
Issued
Consistency value
of the sample:
%CS_CALIBR
Rev. 0
DC
by:
Pag. 19