2. Installation

(1) Scope of delivery.
Main body(1set), Operation manual (1EA), Fuse (2EA), Shelf(2EA), Shelf guide(4EA) ,
Communication CD(1EA), RS-232 Communication cable(1EA)
(2) This unit will work correctly on proper power supply. Please check power supply and ID
Plate information are the same. User must use power supply connected earth and power
cord must be connected to wall outlet supplying ground point.
We don't have any responsibility for accident and lose of personal and asset if the ground earth is not
connected.
(3) Please install the unit in the flat place where prevent vibration and shock.
(4) Please let the unit avoid heat source and direct sun light and let the unit located in ambient
temperature range in 5°C ~ 40°C relative humidity lower than 80%.
(5) Please don't install the unit where water and organic solvent is easily penetrates in the unit.
They cause short circuit.
(6) Please don't install the unit in dangerous place. (Where there are flammable gas and
explosive material)
(7) Please secure enough space for installation because the door of it opened 180°.
(8) Do not install near to instruments of strong high-frequency ( high-frequency welding
machine, sewing machines, SCR Controller for big capacity)
(9) This unit is quite heavy. Please move the unit with proper moving tool or 2 people together.
(10) Power cord should includes Ground terminals when installing the instrument.
(11) Install the instrument near to consents for easy usage.
Please think about the safety of user and instrument when you find out a place to install.