Battery Charger

Note:
Caution!
ASD Document 600541 Rev. G
Recharge a fully discharged battery as soon as possible (usually within 24
hours). Uncharged batteries will be incapable of accepting a charge if the
period of time between charges is too long. A partially discharged battery
does not need charging immediately. However, prior to storing batteries
ensure that they are fully recharged. When 100% charged, these batteries can
be stored up to six months, at this time perform a topping charge before use.
Always store the battery in a cool, dry location.
Do not use batteries other than those supplied by ASD.
Do not use ASD batteries in a manner unauthorized by
ASD. Using improper batteries or improper use of ASD
batteries could result in bodily injury or damage to the
instrument.
The battery charger specifications:
Type:
AC Input:
DC Input:
Charge Time:
Review the charger instruction manual for details.
LED Status indicator:
The LED will be ORANGE during charge.
The LED will turn GREEN when charge is completed.
The battery charger is designed for indoor use and should not come in contact
with water or dust.
Do not unplug the charger and leave the batteries in the charger. Remove the
batteries first, or they will discharge through the charger.
Constant overcharging or undercharging will shorten battery life.
The TerraSpec spectrometer will stop collecting spectra
when the battery drains to about 9.7 volts and the software
will post a "
Spectrometer Isn't Ready
instrument and attached probe will continue to draw
current if power is left on. Power down the instrument and
remove probe as soon as possible when the above message
occurs.
32
Chapter 5 Batteries
External Desktop
90-240 VAC, 50/60 Hz
12 V 1.5 A
under 8 hours for a fully discharged 9 AH battery
" message. However, the