To use roll mode

To use roll mode
Roll mode continuously moves data across the display from right to left. It
allows you to see dynamic changes (like adjusting a potentiometer) on low
frequency signals. Two frequently used applications are transducer
monitoring and power supply testing.
Press
. Then press the
1
Mode
Press
2
Main/Delayed
Press the
Roll
softkey.
3
The oscilloscope is now untriggered and runs continuously. Also notice that
the time reference softkey selection changes to center and right.
Press
. Then press the
4
Mode
The oscilloscope fills either
reference selection), then it searches for a trigger. After a trigger is found,
the remainder of the display is filled. Then, the oscilloscope stops acquiring
data.
You can also make automatic measurements in the roll mode. Notice that the
oscilloscope briefly interrupts the moving data while it makes the
measurement. The acquisition system does not miss any data during the
measurement. The slight shift in the display after the measurement is
complete is that of the display catching up to the acquisition system.
Roll mode operating hints
• Roll mode operates on channels 1 and 2 only.
• Math functions, averaging, and peak detect are not available.
• Holdoff and horizontal delay do not affect the signal.
• Both a free running (nontriggered) display and a triggered display (available
in the single mode only) are available.
• Roll mode is available at sweep speeds up to 200 ms.
1–16
,