Hettich Roto Silenta 630RS
General
Maintenance and servicing
•
Pull the mains plug before cleaning. For permanently connected
devices, switch off the mains switch to disconnect the mains supply
to the device.
•
Centrifuges, rotors and accessories must not be cleaned in rinsing
machines.
•
They may only be cleaned by hand and disinfected with liquids.
•
The water temperature must be between 20 – 25°C.
•
Only detergents/disinfectants may be used which:
have a pH between 5 – 8
o
do not contain caustic alkalis, peroxides, chlorine compounds,
o
acids and alkaline solutions
•
In order to prevent appearances of corrosion through cleaning
agents or disinfectants, the application guide from the manufacturer
of the cleaning agent or disinfectant are absolutely to be heeded.
Surface disinfection
•
If infectious materials penetrates into the centrifugal chamber this is
to be disinfected immediately.
•
Ingredients of suitable disinfectants: ethanol, n-propanol, isopropyl alcohol, glutardialdehyde, quaternary ammonium
compounds.
•
After using disinfectants, remove the disinfectant residue by wiping with a damp cloth.
•
The surfaces must be dried immediately after disinfecting
Autoclaving
•
Swing-out rotors, angle rotors made of aluminium, suspension made of metal, lids with bio-containment as well as
stands and reductions can be autoclaved at 121° C / 250°F (20 mins).
•
Otherwise you must ask the manufacturer.
•
No statement can be made about the degree of sterility.
•
The lids of the rotors and containers must be removed prior to autoclaving.
•
Autoclaving accelerates the ageing process of plastics. In addition, autoclaving may discolour plastics.
•
We recommend that the packing rings of the bio-safety system be replaced after autoclaving.