Hettich Roto Silenta 630RS
General
Maintenance and servicing
Pull the mains plug before cleaning. For permanently connected
devices, switch off the mains switch to disconnect the mains supply
to the device.
Centrifuges, rotors and accessories must not be cleaned in rinsing
machines.
They may only be cleaned by hand and disinfected with liquids.
The water temperature must be between 20 – 25°C.
Only detergents/disinfectants may be used which:
have a pH between 5 – 8
o
do not contain caustic alkalis, peroxides, chlorine compounds,
o
acids and alkaline solutions
In order to prevent appearances of corrosion through cleaning
agents or disinfectants, the application guide from the manufacturer
of the cleaning agent or disinfectant are absolutely to be heeded.
Surface disinfection
If infectious materials penetrates into the centrifugal chamber this is
to be disinfected immediately.
Ingredients of suitable disinfectants: ethanol, n-propanol, isopropyl alcohol, glutardialdehyde, quaternary ammonium
compounds.
After using disinfectants, remove the disinfectant residue by wiping with a damp cloth.
The surfaces must be dried immediately after disinfecting
Autoclaving
Swing-out rotors, angle rotors made of aluminium, suspension made of metal, lids with bio-containment as well as
stands and reductions can be autoclaved at 121° C / 250°F (20 mins).
Otherwise you must ask the manufacturer.
No statement can be made about the degree of sterility.
The lids of the rotors and containers must be removed prior to autoclaving.
Autoclaving accelerates the ageing process of plastics. In addition, autoclaving may discolour plastics.
We recommend that the packing rings of the bio-safety system be replaced after autoclaving.
Tel: 1800
LABGEAR
LabGear Australia 30 Finchley Street, Milton, Brisbane, QLD 4064
or 07 3368 1112 | Email:
sales@labgearaustralia.com.au
| Web:
www.labgearaustralia.com.au

Advertisement

loading

Summary of Contents for Labgear 630RS

  • Page 1 Autoclaving accelerates the ageing process of plastics. In addition, autoclaving may discolour plastics. • We recommend that the packing rings of the bio-safety system be replaced after autoclaving. LabGear Australia 30 Finchley Street, Milton, Brisbane, QLD 4064 Tel: 1800 LABGEAR or 07 3368 1112 | Email: sales@labgearaustralia.com.au...
  • Page 2 Trunnions With swing-out rotors the trunnions must be regularly lubricated (Hettich Lubricating Grease No. 4051) in order to ensure consistent swinging out of the hangers. LabGear Australia 30 Finchley Street, Milton, Brisbane, QLD 4064 Tel: 1800 LABGEAR or 07 3368 1112 | Email: sales@labgearaustralia.com.au...