Page 1
User Manual PEAK INSTRUMENTS INC P-510 pH Meter User Manual PEAK INSTRUMENTS INC Version 1801 Address: 16223 Park Row, Houston, TX-77084, USA. Website: www.peakii.com. Tel: +1 2819353455...
User Manual PEAK INSTRUMENTS INC CONTENTS I. General Information ....................1 II. Specifications ......................2 III. Device Instructions ....................3 1. Display indicators ....................3 2. Operation Keys ....................3 3. Save, view and delete data................. 4 IV. pH Measurement ...................... 5 1.
PEAK INSTRUMENTS INC I. General Information Thanks for using P-510 pH meter. In order to help you operate and maintain the instrument properly, please read the user manual before using it. We reserve the rights to update the manual and its parts subject to the purpose of improving the instrument’s performance.
User Manual PEAK INSTRUMENTS INC III. Device Instructions 1. Display indicators ① Parameter mode tag ② Measurement value ③ Data storage and replay number and symbol. Indicator of special status. M+ is symbol for data storage. RM is data replay icon.
User Manual PEAK INSTRUMENTS INC key to change measuring unit between pH and mV. 2.4. Backlight and delete key. 2.4.1. In measuring mode, short press(less than 1.5 seconds) this key to turn on or off backlit. 2.4.2. When viewing saved data, long press(more than 5 seconds) this key to delete saved data.
User Manual PEAK INSTRUMENTS INC measuring mode. IV. pH Measurement 1. Preparations 1.1. Press to switch on device, then press to choose pH measuring mode. 1.2. Check if the glass bulb of pH combined electrode is moist and complete. If the bulb is broken, then the electrode will not work, if the bulb is too dry, it should be soaked in saturated KCL solution for 24 hours.
User Manual PEAK INSTRUMENTS INC the reading is stable(the symbol will be shown on the screen), then press and display shows 9.18, next press to confirm the third point calibration and quit calibration mode and enter measuring mode, “ ” will be shown on the right bottom and means the instrument is calibrated at three points.
User Manual PEAK INSTRUMENTS INC 4. Parameter setup 4.1. Table of pH meter parameter setup Indicator Description Remarks Manual temperature compensation (0-99.9)℃ setup pH resolution setup 0.1pH/0.01pH pH buffer setup CH, USA, NIST ℃ ℉ Temperature unit setup Ammonia pure water compensation...
User Manual PEAK INSTRUMENTS INC 4.3.3. Press to enter next parameter setting or press to back to measuring mode. 4.4. pH standard buffer setting(P3) 4.4.1. Press to choose P3 in mode P2, Then press to enter P3. 4.4.2. Press to select buffer standard. There are three standards: USA, NIST and China.
User Manual PEAK INSTRUMENTS INC water compensation. Press to confirm the change and back. 4.6.3. Press to enter next parameter setting or press to back to measuring mode. 4.7. Auto shutdown timer of backlight(P6) 4.7.1. Press to choose P6 in mode P5, Then press to enter P6.
User Manual PEAK INSTRUMENTS INC blinks on the screen, which means it is restored to default settings. Press to back to measuring mode. 4.9.3. Be careful to use this function, because all saved data will be removed once restored to default settings.
User Manual PEAK INSTRUMENTS INC response slow, reading inaccurate. The following methods could be used in different situations. 5.8.1. The glass bulb is passivated: Soak the bulb in 0.1mol/L dilute hydrochloric acid(add pure water into 9ml hydrochloric acid to 1000ml) for 24 hours, wash it with pure water, then soak it in electrode immersion solution for 24 hours.
User Manual PEAK INSTRUMENTS INC V. mV measurement 1. Press to switch on the instrument, press to choose mV measuring mode 2. Connected to ORP or ion composite electrode(optional). Wash it and wave it dry, put them into tested solution, stir the electrode and let it be static, then wait until the...