PEAK INSTRUMENTS E-1000 Series User's Manual And Maintenance Manual

PEAK INSTRUMENTS E-1000 Series User's Manual And Maintenance Manual

Spectrophotometer

Advertisement

Quick Links

E-1000 Series User Manual
PEAK INSTRUMENTS INC
E-1000 Series Spectrophotometer
User Manual and Maintenance Guide
PEAK INSTRUMENTS INC.
Version 1701

Advertisement

Table of Contents
loading

Summary of Contents for PEAK INSTRUMENTS E-1000 Series

  • Page 1 E-1000 Series User Manual PEAK INSTRUMENTS INC E-1000 Series Spectrophotometer User Manual and Maintenance Guide PEAK INSTRUMENTS INC. Version 1701...
  • Page 2: Table Of Contents

    E-1000 Series User Manual PEAK INSTRUMENTS INC CONTENTS Section 1 Introduction ............... 1 1.1 Basic working principle ............1 1.1.1 The nature of absorption ..........1 1.1.2 Absorption law ..............1 1.1.3 Influence factors ............. 2 1.2 Applications ................ 4 1.3 Use conditions ..............
  • Page 3: Introduction

    E-1000 Series User Manual PEAK INSTRUMENTS INC Section 1 Introduction 1.1 Basic working principle 1.1.1 The nature of absorption Spectrophotometer is the method set up by the use of substances to choose different wavelengths of light absorption properties. Generally utilize a prism or grating to get monochromatic light, so...
  • Page 4: Influence Factors

    E-1000 Series User Manual PEAK INSTRUMENTS INC The premise of absorption Law mathematical formula: ① the incident light is monochromatic, ② in the absorption process no interaction of each substance, the absorbance of each substance has additive property, ③the role of light and matter is limited to the absorption process, no fluorescent and photochemical scattering phenomenon, ④...
  • Page 5 E-1000 Series User Manual PEAK INSTRUMENTS INC the Beer's law applies only to dilute solution. 1.1.3.6 Stray light, stray light is wavelength component which is not required, but enters the detector and is outside the tested spectral bandwidth range. It comes mainly from the...
  • Page 6: Applications

    E-1000 Series User Manual PEAK INSTRUMENTS INC 1.2 Applications Discipline for physics, chemistry, medicine, biology, pharmacology, geology and other scientific research, is one of the most important quality control instruments which is widely used in chemical, pharmaceutical, biological and chemical, metallurgy, light industry,...
  • Page 7: Technical Data

    E-1000 Series User Manual PEAK INSTRUMENTS INC reasonable mechanical structure, and the use of large-screen LCD, it will provide very effective and intuitive means for the analytical testing of every laboratory . The menu on large-screen LCD selects and recognizes each corresponding steps that the function you need to complete.
  • Page 8: Installation

    E-1000 Series User Manual PEAK INSTRUMENTS INC 2.3 Installation 2.3.1 After unpacking, carefully check the packing list whether the things inside are complete and intact; 2.3.2 Ensure the working environment meets the above requirements, environmental temperature is 10 ~ 35 ℃, relative humidity is less than 85%, operating voltage (220 ±...
  • Page 9: Display Content Description

    E-1000 Series User Manual PEAK INSTRUMENTS INC 3.2 Display content description Transmittance Absorbance Concentration Factor Test data display Storage status display 3.3 Panel diagram Screen Function Buttons 3.4 Button Functions <Esc>: exit <Mode>: the key used to select the Transmittance, Absorbance,...
  • Page 10: Programs

    E-1000 Series User Manual PEAK INSTRUMENTS INC Concentration and Factor mode. <0%T>: the key has two functions: a. To set zero. It is effective only in T mode. Insert black block into cell holder, then close the sample compartment cover. If it does not show 0.0, Press the key, and it will show 0.0.
  • Page 11: Transmittance

    E-1000 Series User Manual PEAK INSTRUMENTS INC the substance to be tested, which dissolved in preparation solution. 4.1 Transmittance 4.1.1. Turn the instrument on. Allow the instrument to warm up for 30 minutes. 4.1.2. Put the reference solution and the solution to be tested into cuvettes separately.
  • Page 12: Quantitation

    E-1000 Series User Manual PEAK INSTRUMENTS INC close the sample compartment cover. 4.2.4 Rotate the wavelength knob to set wavelength. 4.2.5 Press Mode key to T mode, pull the black block into light path. Press 0%T key until display reads 0.0. Please note to set zero one more time if wavelength make a change.
  • Page 13: Measure Concentration With Factor(Coefficient)

    E-1000 Series User Manual PEAK INSTRUMENTS INC 4.3.2 Measure concentration with factor(coefficient). 4.3.2.1 Press Mode key to ‘A’ mode. 4.3.2.2 Rotate the wavelength knob to set wavelength. Please note to set zero one more time if wavelength make a change.
  • Page 14: Delete Data

    E-1000 Series User Manual PEAK INSTRUMENTS INC 4.5 Delete data Hold ESC key 3 seconds or more after turn on the instrument to cancel all saved data. 4.6 Print data Connect to thermo printer with serial port. Press Enter key to print value under T/ A mode.

This manual is also suitable for:

E-1000vE-1000uv