To Set The Rf Output Power Level; To Set An Amplitude Reference And An Amplitude Offset - HP ESG-D Series User Manual

Signal generator
Hide thumbs Also See for ESG-D Series:
Table of Contents

Advertisement

Using Functions
Setting Frequency and Power Levels

To Set the RF Output Power Level

1. Press
Preset
to return the signal generator to the factory-defined instrument state.
2. Observe the amplitude area of the display (in the upper middle of the display). It should
display −135 dBm. This is the normal preset RF output power level.
3. The signal generator is set to output an RF signal, however the
set to
On
before the RF signal is available at the RF OUTPUT connector. Press
. Notice that the display annunciator changes from RF OFF to RF ON. The
RF On/Off
RF signal is now being output at a −135 dBm level at the RF OUTPUT connector.
4. Change the power level to −20 dBm. Press
displayed in the active entry area of the display. Using the numeric keypad and the
key, enter −
20
power is now displayed in the amplitude area of the display and also in the active entry
area.
5. Amplitude is still the active function until you press another front panel function key.
You can also change the amplitude using the up and down arrow keys and the front
panel knob. Practice changing the amplitude using these methods also.

To Set an Amplitude Reference and an Amplitude Offset

The following procedure sets the RF output power as an amplitude reference to
which all other amplitude parameters are relative. The amplitude initially shown
on the display will be 0 dB (the power output by the hardware minus the reference
power). Although the display changes, the output power does not change. Any
subsequent power changes are shown as incremental or decremental to 0 dB.
1. Press
to return the signal generator to the factory-defined instrument state.
Preset
2. Set the RF output power to −20 dBm.
3. Turn on amplitude reference mode and set the current output power (−20 dBm) as the
reference value. Press
0 dB (the power output by the hardware, −20 dBm, minus the reference value,
−20 dBm.) Notice that the REF indicator is turned on and the
toggled to
.
On
4. Set the
key to
RF On/Off
RF OFF to RF ON. The power at the RF OUTPUT connector is −20 dBm.
5. Use the up arrow key to increase the output power by 10 dB. The amplitude display
changes to show 10 dB (the power output by the hardware, −20 dBm + 10 dBm, minus
the reference power, −20 dBm) and the output power changes to −10 dBm.
2-4
Amplitude
softkey. The new −20 dBm RF output
and then press the
dBm
Ampl
,
More (1 of 2)
,
Ampl Ref Set
. Notice that the display annunciator has changed from
On
RF On/Off
key must be
. The current power level is now
. The amplitude displayed is
softkey has
Ampl Ref