Calibrate Setup Menu; Legal Standard - MSI CellScale 3750CS User Manual

Rf digital weight indicator
Table of Contents

Advertisement

3750CS

CALIBRATE SETUP MENU

LEGAL STANDARD

The STANDARD function allows the Model 3750CS to be configured as required by various regulatory
agencies. The "INDUSTRY" selection is the default. Choose "NIST" for US and Canadian Legal-for-Trade
units. Choose "OIML" for European L-F-T applications. "METRIC" is the same as "INDUSTRY" except
only kg and g units are enabled.
Before starting the following procedure, you must be in the Top Menu Level of Calibration. See "Enable Calibration". Contact MSI for details on the
differences between the various configurations.
To Select the Legal Standard
1) Use the UP or DOWN scroll key to scroll to the
"STANDARD" message.
2) Press ENTER. The current Standard is displayed.
3) Use the UP or DOWN key to scroll through the
Standards options. For this example we'll pick
the "METRIC" standard. This will limit the units
choices to kg and g (and Metric Tons if the capacity
is high enough).
4) Push ENTER to set your selection.
5) You are now back in the "CAL SETUP" menu. Use
the UP/DOWN keys to scroll to the next calibration
operation or push EXIT to return to normal scale
operation.
Page 78 MSI-3750CS RF Weighmeter • User Guide
C
S
ELL
CALE RF DIGITAL WEIGHT INDICATOR
®
...
1
STANDARD
1
INDUSTRY
2
NIST
3
OIML
3
METRIC
3
STANDARD
4
or
5

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the CellScale 3750CS and is the answer not in the manual?

Subscribe to Our Youtube Channel

Table of Contents