Note: Ø Please read the manual carefully before operating the product. Warning: Ø Tubing may have crack due to wear. It results in the overflow of fluid from tubing. In that time human body and instruments may be harmed. So users must check frequently and change tubing in time. Ø...
1. Product Introduction KA1 adopts ABS engineering plastic housing, 2.4 inch LCD screen, low power, ultra-quiet, compact structure. Easy to operate with digital knob speed control. A variety of external control methods for option. Support RS485 communication, standard MODBUS communication protocol (RTU mode); It is suitable for complex working environment with super anti-interference and wide voltage design.
4. Operation Interface Structure 4.1 Booting Interface When power on, enter to the booting interface, the system will enter to the English main interface after waiting for 2.5s. 4.2 Main Interface Main interface structure as the follow picture: Running Flow status Speed Head...
Page 6
4.3.1 Basic Configuration Interface Head&Tube Head&Tube FD15 FD15 Head: KT15 Head: KT15 Tube: Tube: Suck back: Suck back: Operating steps: Press [Knob] to select tubing in setting interface, when in status (1) , rotate [Knob], the cursor up and down, select the tubing or back suction angle.
Page 7
4.3.3 Setting Communication Interface Communication Communication RS485 RS485 Port: Port: Address: Address: 9600.8.e.1 9600.8.e.1 Baud rate: Baud rate: Operation steps: Serial communication is always on, and the on / off option is not set. In setting communication interface, press [Knob] button to select native address, when in status (1), rotate [Knob] button to select native address or baud rate.
Page 8
When in status (1), press [Knob] button to status (2). When in status (1), rotate knob to ESC position and press the [Knob], then back to [Menu selection interface]. When in status (2), rotate [Knob] button to select turning on/off external control start/stop, turning on/off external control reversing or selecting signal to level/pulse (The corresponding inner control button does not make effect in level mode).
Page 9
Note: Select analog speed regulation signal according to external port inputting signal, there are three options: 0-5V, 0-10V,4-20mA. External control speed regulation can set the maximum speed for the pump, the maximum speed is 350rpm, the analog signal voltage range is liner with the motor speed. If the maximum speed is not 350rpm, the motor speed will be limited by the analog signal, motor speed analog signal When the motor speed and analog signal reach the set maximum speed according to the corresponding proportional relationship, the analog signal will be increased at this time, and the motor speed will run at the set maximum speed and will not increase with the analog...
5. External Control Interface The external control interface of 15 hole base is shown in the figure below: The external control interface is 15 hole base, the function of 15 hole pin is: Pin1 is the external control start/stop port; Pin2 is the external control reversing port;...
function of external control speed regulation. Control motor speed from 0rpm to maximum speed by analog signal. Ø Pin6 0-5V: 0V-5V voltage signal input terminal Ø Pin7 0-10V: 0V-10V voltage signal input terminal Ø Pin8 4-20mA: 4-20mA current signal input terminal Ø...
7. Main Function and Feature Ø 2.4 inch true color LCD display. Ø Ultra-quiet drive settings, accurate control, low vibration and noise. Ø Mechanical keypad and menu interface are convenient for users setting parameters. Ø Easy to operate speed regulation by digital knob. Ø...
Page 13
Electrical surge as a cause of failure is excluded. Ø Chemical attack is excluded. Ø Improper operation or man-made damage as a cause of failure is excluded Flow Controller Systems 2950 Buskirk Ave Suite 300 Walnut Creek CA 94597 Tel: 925-516-4459 Website: flowcontrollersystems.com Email: justin@fcs-us.net...