EMC information

Medical electrical equipment should be used with precautions according to EMC,
and must be installed according to the EMC notices disclosed in this manual as
mobile RF transceivers could adversely affect it.
Electromagnetic emission
Meditech ambulatory blood pressure monitors are suitable for use in the specified
electromagnetic environment. The purchaser or user of the device should assure
that they are used in an electromagnetic environment as described below.
Emission test
Radiated and
conducted RF emission
CISPR 11
Radiated and
conducted RF emission
CISPR 11
Harmonic emission
IEC61000-3-2
Voltage fluctuations/
Flickers IEC61000-3-3
Electromagnetic immunity
Meditech ambulatory blood pressure monitors are suitable for use in the specified
electromagnetic environment. The purchaser or user of Meditech products should
assure that they are used in an electromagnetic environment as described below.
Monitor
Immunity test
ABPM-05
Electrostatic
discharge (ESD)
IEC 61000-4-2
BlueBP-05
ABPM-05 &
Electrical fast
BlueBP-05
transient/burst
IEC 61000-4-4
ABPM-05 &
Surge
BlueBP-05
IEC 61000-4-5
BP5BB5LR_UMEN_V05-20170703
Compliance
Group 1
ABPM-05/BlueBP-05 uses RF energy only for their
internal function. Therefore, the emission is very
low and it is not likely to cause any interference in
nearby electronic equipment. Except for Bluetooth
communication usage. (The Bluetooth
radiofrequency communication is carried out at the
2.4GHz frequency; class2, Bluetooth 2.0+EDR)
Class B
ABPM-05/BlueBP-05 is suitable for use in
domestic establishments and in establishments
directly connected to the low voltage power supply
network which supplies buildings used for
domestic use.
Not applicable
---
Not applicable
---
IEC60601 test
level
+ 6KV contact
+ 8 KV air
+ 2KV for power
supply
+ 1KV input/output
lines
+ 1KV differential
mode
+ 2KV common
mode
Electromagnetic environment
Compliance
Electromagnetic environment
level
+ 8 KV air
Floors are wood, concrete or
ceramic tile, or floors are
covered with synthetic
material and the relative
+ 6KV
humidity is at least 30
contact
percent.
+ 8 KV air
Not
Mains power quality is that of
applicable
a typical commercial and/or
hospital environment.
Not
Mains power quality is that of
applicable
a typical commercial and/or
hospital environment.