EasyABPM: Device/Read data
7. Customize, review & print data
7.1 Customize your standard report
CardioVisions: Tools/Options/Standard reports/ABPM report
EasyABPM: Tools/Settings/Report
7.2 Review the study and edit data if necessary.
7.3 Create and save or print the report

ABPM-05 Manual Programming

ABPM-05 if programmed manually can be used by CardioVisions software 1.13 or
later versions. Plans are stored in the inbuilt memory of the device and they cannot
be changed. The following three measurement plans can be selected during the
programming of the device:
PLAN A): measurements every 15 minutes at day and every 30 minutes at night.
PLAN B): periods with 20 minutes at day and 40 minutes at night.
PLAN C):30-minute periods independent of day or night time.
Other settings are the same in all the three plans: undecided cuff size, 300 mmHg
pressure limit, LCD display enabled, manual day/night shift disabled. Daytime
starts at 6:00, while nighttime starts at 22:00, special session is disabled. Patient
data can be selected or created later in CardioVisions database.
How to program manually
In case of manual programming there is no time setting. If the time setting is
BP5BB5LR_UMEN_V05-20170703
Press and hold the Start and the Day/night buttons
simultaneously. Measurement
measurement plans will be displayed after 10 seconds
for 3-3 seconds.
To choose the measurement plan, release the buttons
when the specific plan is displayed. You will hear two
beeps and the LCD will display four blinking „o"
letters, which indicates that programming the device
is in progress. After a successful programming you
can hear 5 beeps, and the selected plan can be seen
again. If programming fails for some reason, the E90
error code will be displayed on the LCD.