Patient Information - Meditech ABPM-05 User Manual

Ambulatory blood pressure monitors
Hide thumbs Also See for ABPM-05:
Table of Contents

Advertisement

如果是手动编程,则没有时间设置。如果时间设置不精确,则测量时间可能是错误的。如果要使用手动编程功能,请不要让设备长时间没有
电池。如果可能发生这种情况,请将电池重新插入设备,通过PC对设备进行编程来设置内部时钟,并将电池留在设备中(建议的周期为24
小时)。第一次测量具有控制目的,在编程后的第二分钟开始,然后在接下来的五分钟内不能开始其他测量。其余的测量以特定的
15/20/30/40分钟间隔进行,在6:00和22:00小时轮班进行测量。最后一次测量正好是第二次测量后24小时。
imprecise, the time of measurements may be false. If you want to use the manual
programming function, do not leave the device without batteries for a longer period.
If it could happened so, insert batteries again into the device, set the inner clock by
programming the device by a PC and leave the batteries in the device (the
suggested period is 24 hours).
The first measurement has a controlling purpose and it starts in the second minute
after programming, then other measurements cannot be started in the next five
minutes. The rest of the measurements are taken at specific 15/20/30/40 minute
intervals and there are measurements at the 6:00 and 22:00 hour shifts. The last
measurement is exactly 24 hours after the second measurement.
LCD displays

Patient Information

Wearing a thin shirt under the cuff is recommended.
It does not influence the accuracy of the blood pressure measurement, but it
prevents problems caused by long-time wear of the cuff (sweat, itching, soreness,
etc.)
The cuff should be properly placed on and connected.
The cuff tube should be pointed towards the patient's shoulder and the white
tissue-sign (textile cuff) or the 'artery' indicator (PU fabric leather cuff) on the cuff
should be placed above the brachial artery.
Hold your arm slightly away from your chest during a measurement.
Patients should avoid excess movement during blood pressure measurements.
They should hold their arm loose, slightly away from their chest.
Should the blood pressure measurements cause bloodshots, torpidity or pain in the
hand, the cuff should be removed from the arm immediately and disconnected from
the recorder.
Such occurrence should be reported to the physician latest after the monitoring
session.
BP5BB5LR_UMEN_V05-20170703
Normal status: time
is displayed.
10 second delay
state
1. measurement
plan: 15/30 minute
day/night intervals
2. measurement
plan: 20/40 minute
day/night intervals
3. measurement
plan: 30/30 minute
day/night intervals
Programming is in
progress: blinking
signal
13

Hide quick links:

Advertisement

Table of Contents
loading

This manual is also suitable for:

Bluebp-05

Table of Contents