Page 8
14.3 Guidelines and manufacturer's statement - electromagnetic immunity ......98 14.4 Recommended isolation distance ..................100 14.5 Basic EMC performance of T6 ventilator ................101 15 Product warranty ..........................102 16 Classification details of toxic and harmful substances ..............103...
Page 9
Product information Thank you for purchasing the T6 ventilator. To use this device correctly please read and understand the contents of this Device manual carefully before use. After reading, keep this manual at a proper place where it is easy to access.
Page 10
Intellectual property rights @2021 Ambulanc (Shenzhen) Tech. Co., Ltd., All rights reserved The intellectual property rights of this product and its operation manual belongs to Ambulanc (Shenzhen) Tech. Co., Ltd., including but not limited to the patent right, trademark right, and copyright, etc. Ambulanc (Shenzhen) Tech.
Page 11
Statement Ambulanc (Shenzhen) Tech. Co., Ltd. reserves the right to modify contents of this Manual without prior notice. Ambulanc (Shenzhen) Tech. Co., Ltd. reserves the right to change the technology without prior notice. Ambulanc (Shenzhen) Tech. Co., Ltd. reserves the right to modify the product specifications without prior notice.
Page 12
Maintenance service Free maintenance range: • All devices in line with warranty service regulations of Ambulanc (Shenzhen) Tech. Co., Ltd. can enjoy free service. Charged maintenance range: • Any equipment beyond the range of warranty services regulations of Ambulanc (Shenzhen) Tech. Co., Ltd. will be charged by Ambulanc for services; •...
Page 13
Return Return procedures Please follow the following steps if you do need to return the goods to Ambulanc (Shenzhen) Tech. Co., Ltd.: • Obtain the right of return: Contact the after-sales service department of Ambulanc (Shenzhen) Tech. Co., Ltd. to inform serial number of Ambulanc product, which has been marked on the outer packing box.
Page 14
not designated by Ambulanc. 6. This system is designed to provide doctors with the auxiliary tools needed for clinical treatment. 7. The doctors are responsible for the course of treatment.Ambulanc has no responsibility for the course of treatment. 8. Important data should always be backed up to external storage media, such as clinical records, notebooks, etc.
• Follow the instructions in "9 Cleaning and Disinfection" to prevent infection or bacterial infection. • T6 can only be used after getting the proper medical training and technical guidance on ventilator usage as improper use may lead to death of the patient.
• If the device is used in dirty environment, the filter should be replaced as described in "11.14 Replace of filter cotton ". • Do not immerse T6 in any liquid. If any liquid gets into the cover, it can cause damage to the device.
Ventilation/operation • The patient and ventilator must be continuously observed during ventilation. • Prolonged reliance on T6 for breathing may cause respiratory muscles of the patient to atrophy. • Ventilation for a long time can dry out the respiratory tract.
• Connect the external power supply of T6. Caution: [Maintaining of battery installation] In order to enable continuous operation of T6, it is recommended to use a fully charged battery for the whole time (even when an external power supply is connected to supply power).
Page 20
application part Equipotential Protection level Do not discard in an ordinary trash Power cord disconnection Refer to the manual provided with On/Off switch of the device together/Operation main unit Manual Battery power AC power supply supply Clearing alarms other than Nebulization Advanced Alarms Function menu...
It is used for ventilation assistance and respiratory support for adults, children and infants. The T6 may be operated only if it is safely installed and fixed or placed on a licensed carrier platform. Warning: The ventilator shall not be covered or placed in a position that affects operation and performance of the ventilator.
Ambulanc (Shenzhen) Tech. Co., Ltd. has designed all components required for the T6 ventilation system. Appearance description 2.6.1 Mainframe - front view Fig. 2-1 Mainframe (front view) Components Description 1 Alarm light In case of an alarm, the display turns into red, yellow...
Page 23
4 Battery sign Battery indicator. 5 Battery The indicator will flash during charge, and will stay on indicator untill the battery is fully charged or in use. Tip: To indicate the ventilator is powered by an internal battery or an external power supply during operation. 6 On/off key If it is in the shutdown state, click this button to turn on,if it is in standby state, click this button to turn off, and if it is in...
sensor interface 16 Nebulizer Used to connect the nebulizer. interface 17 Leak test Used to calibrate compliance plug 18 Inspiratory Used to connect the inspiratory line. branch 2.6.2 Mainframe - Rear view Fig. 2-1 Mainframe (rear view) Components Description 1 DC power input port Used to connect to the vehicle power supply.
upgrade. 6 Hook Hook width: 45mm. 7 Gas supply Connect to high pressure gas supply port(high pressure) 8 Gas supply port(low Connect to low pressure gas supply. pressure) 用于连接外部校准设备,活可以连接医疗级外部 设备。 Used to remove and replace the rechargeable built in 9 Battery latch 支持与...
2.6.5 T6 component diagram Fig. 2-2 T6 component diagram Components Description 1 Support arm Used to support and suspend respiratory line of patients. Include the oxygen line, electronic system, mechanical 2 Mainframe structure, display, and carbon dioxide module. 3 Trolley Used to support the mainframe, support arm, oxygen cylinder and humidifier, etc.
⚫ After installation, you must do functional inspection (please refer to the "11 Maintenance and inspection") to ensure that the device works properly. Packing items T6 ventilator is packed in a single box. Please refer to "12 T6 accessories" for packing items. Installation of battery The battery used in T6 is a rechargeable lithium battery.After insertion, push it by hand,...
Connection of oxygen source Fig. 3-2 Oxygen source interface The ventilator can provide two gas source interfaces: high-pressure oxygen and low-pressure oxygen. When the ventilator is connected with the high-pressure oxygen gas source, the normal working gas source pressure is 300-600kPa. If pressure of gas source is lower than 300kPa, performance of the ventilator will be affected, or even the ventilation may be disabled.
interface shall be pressed, and then the gas source hose shall be pulled out. Power supply connection The ventilator can be connected to DC power supply and AC power supply. 3.4.1 Connect AC power supply Fig. 3-3 AC power interface 1.
Insert the DC power cord directly into the DC power input interface, and then rotate clockwise until a "Pop" sound is heard which shows that it has been installed in place. Install the mainframe Fig. 3-5 Mainframe installation diagram ①Mainframe ②Positioning columns ③...
2. Tighten knob of the fixing block. Patient respiratory line assembly and its connection Respiratory line assembly of T6 is divided into repetitive respiratory line assembly and disposable respiratory line assembly. And connection mode of the line is: double-line connection mode. Following steps shall be followed in the connection mode (as shown...
Page 33
Components Components 1 Expiratory branch filter 7 Humidifier 2 Expiratory line 8 Humidifier gas inlet 3 Flow sensor 9 Humidifier gas outlet 4 Simulated lung (patient) 10 Inspiratory line 5 Seeper trap in expiratory 11 Inspiratory branch filter branch 6 Seeper trap in inspiratory branch Caution: 1.
Install humidifier Fig. 3-8 Humidifier installation diagram 1. Align humidifier pulley with the humidifier support mounting seat and slide into it, and then tighten the screws. 2. Install the filter to the inspiratory and expiratory interface. 3. Connect the filter in inspiratory branch to the humidifier inlet via the line. 4.
3.10 Infant noninvasive pipeline connection Fig. 3-10 Schematic diagram of noninvasive pipeline connection for infants parts parts ①Infant nasal oxygen tube ② pressure pipeline tube ③Expiratory line ④Humidifier Warning: Infant patients: •You can only select CPAP, PCV mode or switch from CPAP, PCV mode to other modes during standby.
3.11 Install nebulization Nebulization refers to the atomization of medicines into aerosols, which are then inhaled into the patient's body for the purpose of treatment. Connect the nebulizer to the machine correctly, as shown in the figure below: Fig. 3-11 Nebulizer installation diagram Parts Parts 1 Nebulization port...
3.13 Patient breathe valve Patient inspiratory valve 3.13.1 Fig. 3-13 Patient inspiratory valve Components Description Can be used to replace diaphragm of the safety valve. 1 Safety valve cover Used to connect to patient inspiratory line, and is 2 Gas outlet provided with a Φ15mm/22mm coaxial interface.
4 Interface description Main interface components Fig. 4-1 Main interface Components Description 1 System date and time Display current time and date. 2 Power supply service state Display battery power and charging status. 3 Adapter state Show whether the adapter is in place. Indicate that current respiration is triggered by 4 Patient trigger the patient...
Page 40
8 Screen lock Display locking status of screen 9 USB Display that the USB is connected now 10 Freezing mode Displays the waveform freeze status. 11 Alarm mute prompt icon Display current alarm mute state. 12 Patient type icon Display current patient type (adult/Pediatric). 13 Ventilation mode Display current mode.
Waveform interface Select the [Waveforms] button on the main interface screen to open the interface as shown in the figure below Fig. 4-2 Waveform interface 4.2.1 Monitor waveform switching In the main interface, click the waveform to be switched, double click to pop up the switching options, and then select to switch Fig.
Loops interface Select the [Loops] button on the main interface to open the interface as shown in the figure below. This interface can display the combination of 2 loops and 1 waveform on the same screen Fig. 4-4 Loops interface Monitoring value interface Select the [values] on the main interface to open the interface as shown in the figure below...
Trend Chart interface In the main interface, select the [Trends] option to pop up the trend setting interface. Enter the trend interface. You can set up the trend Chart or view trend data and trend waveforms in this interface. And view the 30min, 60min, 2h ,6h,12h or 24h trend charts.The trend chart can be saved for a maximum of 168 hours.
storage of up to 5,000 messages. When the maximum number of log messages is reached, the newer event will overwrite the older ones. The following points shall be noted when viewing the logs: Events structure: time + type + message. •...
Page 45
Fig. 4-7 System settings Keys Description 1 Time setting Year, month, day, hour, minute 2 Language Chinese (default). Adjustable range: Chinese, English 3 Sound Type Beep(default) 4 Volume Level 1 by default. Adjustable level 1-3 5 Luminance Level 5 (default). Adjustable range: level 1-8 6 style Night (default).
4.9.2 General Fig. 4-8 General setting Keys Description 1 Nebu time 1-90 min (adjustable) 2 Nebu Position Inhalation pipe/ patient end optional 3 Hold Time 1-40s (adjustable) 4 Gas ATPD, STPD, BTPS (adjustable) 5 Humidifying Artificial nose and humidifier (optional) 6 LPO Open and close optional 7 Oxygen sensor...
5 Special functions Lung recruitment maneuver(RM) Lung recruitment maneuver is a pulmonary protective ventilation strategy. The applying of a pressure higher than the normal mean airway pressure and maintaining it for a certain period of time during mechanical ventilation can enable reexpansion of more collapsed alveoli on one hand, and prevent secondary atelectasis caused by small tidal ventilation on the other hand.
Step 2: Start CPR by following the voice prompts and the "Dong Dong Dong..."beat Caution: ⚫ Continuous mode It's not available in infant types. PEEPi PEEPi (intrinsic PEEP) refers to the positive end-expiratory alveolar pressure in the absence of extrinsic PEEP 1.
5.5 NIF NIF means the maximum negative pressure generated by the patient breathing spontaneously over a period of time. 1. Select [Tools] button → [Diagnostics] → [NIF]. 2. After pressing the [Exp Hold] button, the system will start the measurement. 3.
Suction Sputum suction is the process during which ventilator is used to assist the user to suck sputum of the patient. The ventilator automatically detects the action of the user disconnecting and connecting the patient's tube, applies oxygenated ventilation before and after sputum suction, and block relevant alarms during the process of sputum suction.
Caution: ⚫ Not available in HFNC, CPR and CPAP mode. ⚫ You can not activate Insp Hold function in CPAP/PSV ventilation mode. If apnea ventilation occurs, the Inspiration Hold function is supported. Exp.holding Expiratory holding is the practice of artificially prolonging the patient's expiratory stage and preventing the patient from inspiration for a certain period of time.
After clicking the → [General] → [Nebu time], and selecting the [Save] button after setting, and the ventilator will work according to the set nebulization time. After the nebulization function is enabled, remaining time of this function will be displayed in the system prompt message area.
Page 54
⚫ It is not available in CPAP/PSV, non-invasive and apnea ventilation modes. ⚫ It is not available during the process of sputum suction and within 1 minute after the process, and not available within 1 minute after the last P-V test.
6 Alarm Alarm message The alarm messages with the highest priority at present will be displayed in the alarm prompt bar on the main interface. If there are multiple alarms, you can click the alarm prompt message bar in the main interface, and the alarm message interface will be brought up to view other alarm messages (as shown below).
Type Voice alarm displayed!!! time. |0.1s|0.1s|0.5s|0.1s|, the duration of pulse is 0.2s, and the interval of pulse group is 7s. Medium-priority alarms The alarm area on The yellow 1. Three continuous the main interface light flashes “beep--beep--beep” will turn red and Flash sounds will be heard, corresponding...
Page 57
Alarm code Alarm name Alarm description Priority 1011 Failure 1011! Motor temperature read failure High Oxygen supply pressure 1012 Failure 1012! High sensor failure Ambient pressure sensor 1 1013 Failure 1013! High failure Ambient pressure sensor 2 1014 Failure 1014! High failure Incorrect ambient pressure...
Page 58
Alarm code Alarm name Alarm description Priority Patient side pressure Patient side pressure sensor 1031 High sensor failure!!! failure 1032 No pressure sensor!!! No pressure sensor High Input gas temperature greater 1033 High temp of input gas!!! High than 50°C Output gas temperature is 1034 High temp of output gas!!!
Alarm code Alarm name Alarm description Priority Oxygen concentration is 2005 Low FiO2!!! High below lower limit 2006 High EtCO EtCO is above upper limit High 2007 Low EtCO EtCO is below lower limit High 2008 High PEEP!!! PEEP is above upper limit High 2009 Low PEEP!
High temperature during Battery temp is too high, battery discharge, and the 1051 High may turn off!!! system is about to shut down. (>75 °C) Battery temp high, High temperature during 1052 connect external Medium battery discharge (≥ 65 °C) power!! External power External power supply...
Condition Horn alarm messages are displayed in the alarm interface. Multi-ple Only the alarm The yellow light If it is a beep low-priority messages with the is always on alarm, it will be alarms are given highest priority will be given with low synchronously displayed in the...
Alarm cancellation: ⚫ When the alarm limit is set to [OFF], the system will turn off the physiological alarm of the corresponding parameters. Namely, the text message, visual alarm, audible alarm and parameter flashing of the physiological alarm are all canceled. Setting of alarm limits Alarm limits can be set according to physiological characteristics of the patient's breathing.
Page 63
Alarm parameter Alarm range Lower limit of positive end Closed, 1cmH2O-40 cmH2O expiratory pressure Upper limit of airway pressure 12cmH O~100 cmH Lower limit of airway pressure Closed, 1cmH O-90 cmH Apnea 5s~60s Gas source pressure is Pressure of oxygen source is insufficient below 250 kPa Pressure of oxygen source is...
Power-on Plug the power cord into power outlet. Make sure the external power indicator is on. 1. Press the [ON/OFF] button to turn on T6. 2. At this time, a progress bar indicating that the ventilator is subjected to self-inspection will appear on the screen.
System calibration includes: Self check and zeroing ⚫ Hose resistance ⚫ Hose compliance ⚫ Flow sensor ⚫ Caution: The automatic self-inspection function is not intended as a substitute for function ⚫ inspection. When using the machine, function inspection of the machine shall be carried out according to the contents described in "10 Maintenance and inspection".
Fig. 7-3 Patient setting Caution: ⚫ Each time you switch patient types, a system calibration must be performed to initiate ventilation. ⚫ If you select the previous patient’s ventilation, the machine will use the previous patient ventilation settings and alarm settings by default. Ventilation type This ventilator has two ventilation types: invasive ventilation and non-invasive ventilation.
intubation. 7.4.2 Non-invasive ventilation(NIV) Non-invasive ventilation refers to the fact that the ventilation of patient is enabled by a nasal mask or respiratory mask without endotracheal intubation or tracheotomy intubation. In non-invasive ventilation mode, the following is available for adult and pediatric types: P-A/C, P-SIMV, APRV, BiPPV, CPAP/PSV, and in infant types: CPAP, PCV.
save, or [×] to cancel the operation. 4. Follow the same method to set other parameters that needs to be set. Fig. 7-4 Ventilation parameter setting 7.5.2 Ventilation parameters in each mode Ventilation mode Set parameters Tidal volume/inspiratory time/breath rate/PEEP/oxygen IPPV concentration/upper pressure limit/inspiratory pause/sigh Tidal volume/inspiratory time/breath rate/PEEP/oxygen...
Oxygen therapy flow/oxygen concentration/upper pressure HFNC limit 7.5.3 IPPV and PCV • Intermittent positive pressure ventilation: IPPV During mechanical ventilation of IPPV (Intermittent Positive Pressure Ventilation), the ventilator always provides intermittent positive pressure ventilation. The pressure rise during inhalation/inspiration is manifested as positive pressure, while during exhalation/expiration the pressure returns to baseline pressure .
ratio imbalance caused by ARDS or COPD. Even when the respiratory line leaks, it can also ensure the supply of tidal volume. PCV should be used when the trachea is leaking. Typical pressure waveform of PCV is as follows: Fig. 7-6 PCV typical pressure waveform 7.5.4 V-A/C and P-A/C •...
Fig. 7-8 P-A/C typical pressure waveform The [Sigh] is a deep inspiration greater than current tidal volume/pressure at every other set ventilation times on the basis of the specified ventilation frequency, and is applicable for the patients who need mechanical ventilation for a long time. 7.5.5 V-SIMV and P-SIMV Synchronized intermittent mandatory ventilation: SIMV SIMV (Synchronized intermittent Mandatory Ventilation) is a ventilation technology...
P-SIMV (pressure-limited synchronized intermittent mandatory ventilation) refers that the machine provides support pressure if the patient is able to trigger the ventilator outside the trigger window during mechanical ventilation. In a specific trigger window, the ventilator detects the patient's inspiratory effort according to setting of trigger sensitivity and immediately gives forced ventilation once according to the preset pressure, so that supply of the mandatory ventilation is synchronized with the patient's inspiratory force.
The typical pressure waveform of CPAP/PSV is shown below: Pressure Apnea ventilation Pressure support ventilation Apnea pressure Support pressure Inspiratory trigger Rise time Time Fig. 7-11 CPAP/PSV typical pressure waveform 7.5.7 BiPPV bilevel positive airway pressure BiPPV means that during mechanical ventilation or autonomous breathing, the ventilator alternately provide two different levels of positive airway pressure.
7.5.8 APRV APRV mode is known as the airway pressure release ventilation mode, which can be viewed as giving periodic and transient airway pressure release in CPAP mode. The typical pressure waveform of APRV: Fig. 7-13 APRV typical pressure waveform 7.5.9 PRVC In PRVC mode, volume control is carried out in the manner of pressure-controlled ventilation.
Time Time Fig. 7-14 PRVC mode controlled typical pressure and velocity waveforms PRVC-SIMV 7.5.10 PRVC-SIMV mode known pressure-regulated volume control-synchronous intermittent mandatory ventilation mode, and is a ventilation mode that ensures the lowest preset ventilation frequency. The mechanical ventilation mode provided is the volume mode (PRVC mode).
The typical pressure and flow velocity waveforms controlled by CPAP mode are as shown below: Positive airway Pressure pressure Time Flow rate Time Fig. 7-16 CPAP mode controlled typical pressure and flow velocity waveform 7.5.12 HFNC HFNC is the high flow oxygen therapy function. This mode is indicated for the patient who has autonomous respiration.
Caution: ⚫ HFNC can only be used for the patients with autonomous respiration. ⚫ Patients must be given oxygen therapy under the supervision of medical care professional. If something goes wrong or the patient does not have enough autonomous respiration, the medical care professional can help immediately. ⚫...
automatically adjust the gas supply pressure so that the pressure at intubation end is as consistent as possible with set value of the ventilator pressure. Once you select the ventilation mode, select [Compensate] → [ATC] in the parameter setting screen. When it is activated, you can perform corresponding parameter setting. After clicking [Save] button, the system will activate the intubation Compensation function;...
there is residual gas pressure in the cylinder when you return it to refill, which will prevent moist/humid air in the surrounding environment from intruding and causing corrosion/rust. 1. Please check the barometer on the pressure reducing valve that shows the pressure of the cylinder to check the gas storage condition of the cylinder.
7.9 Data Export This machine provides the patient data export function, you can use it to export the current patient’s log information and trend information, etc. 1. Insert the USB drive into the USB port of the ventilator. Click → [Maintenance] → [User Maintenance] → [Data Export] in Standby screen 3.
8 CO monitoring Overview The CO module of this ventilator uses infrared absorption technology to measure the concentration in patient’s respiratory line.The principle is based on the fact that molecules can absorb infrared light energy of a specific wavelength, and the amount of energy absorbed is directly related to the concentration of CO .
monitoring setting 1. Connect the sensor to CO module. 2. The CO module is the measurement mode by default. After connecting the CO module, the machine will automatically start CO monitoring. 3. The zeroing calibration operation shall be carried out as described in Section 8.4 (EtCO zero calibration).
9 Cleaning and disinfection The T6 mainframe and its accessories must be cleaned and disinfected after each use to keep it in a good standby state, to avoid cross-infection. Functional inspection should be carried out after each cleaning and disinfection (see the "11 Maintenance and inspection"...
4. Check the mask and replace it immediately if there are any damaged parts. 5. Re-useable respiratory hoses, patient respiratory valves, silicon parts and respiratory masks can also be sterilized/disinfected with high temperature. Valve accessories Warning: Risk of explosion! It is strictly forbidden to put valve accessories (pressure reducing valves, oxygen cylinders, etc.) in disinfectants or other similar liquids.
Warning: Use of the ventilator without a filter shall be prohibited, in order to avoid affecting performance of the machine, or even cause damage to the machine. Handling method Parts Cleaning Disinfection Rinsing in a Sterilization washer Use a piece Wipe to disinfect Not allowed Not allowed...
Page 87
Medical alcohol(75%) Disinfectant Isopropyl alcohol Disinfectant glutaraldehyde Disinfectant Phthalaldehyde disinfectant Disinfectant Soap water(pH from 7.0 to 10.5) Detergent clean water Detergent...
10.1 Technical faults Faults Causes Remedies Deliver to the manufacturer or T6 malfunction/fault seller for repair T6 cannot be started The battery is used Recharge the battery The gas supply line Identify and correct gas Obvious oxygen loss leakage leakage points T6 cannot be shut Press and hold the "ON/OFF"...
limit is reasonable or not apnea apnea The apnea time Check condition of the exceeds the set patient time value Check that the set time value is reasonable or The set upper limit Check condition of the is exceeded patient Check condition of the Airway obstruction patient...
10.4 Abnormal power failure alarms T6 has the function of shutdown alarm caused by abnormal power failure of the system. The alarm will be triggered when the mainframe is shut down due to the power failure caused by any abnormality, and the alarm duration is not less than 15 seconds;...
11 Maintenance and inspection 11.1 Routine inspections Before each use: • Functional inspection shall be carried out once. After each use: • The reusable respiratory hoses and patient respiratory valves must be cleaned and disinfected once according to the instructions for use in Chapter 6. After each use or removal: •...
Relieve pressure from the system: for this, it is necessary to close the oxygen cylinder. Turn on T6 for a while until reading of pressure gauge of the oxygen cylinder is "0". Then turn T6 off again.
Page 93
When T6 is connected for self-inspection, the low battery power alarm will be inspected automatically. If T6 works normally and no alarm is given after the T6 is powered on when the oxygen cylinder is opened, it shows that the voltage is normal.
7. Trigger pressure function test. The specific steps are as follows: a) The ventilation mode is set to CPAP/PSV ventilation mode, and CPAP pressure is set to 0. b) The trigger pressure is set to -3mbar. c) Inhale through the mask. The ventilator shall supply gas when the inspiratory negative pressure reaches to -3mbar, and will stop supplying gas when the support pressure reaches to the target pressure.
11.7 Flow sensor calibration Calibration of flow sensor shall be carried out when error of flow monitoring value is large or the flow sensor is replaced. Calibration of flow sensor can be performed according to the following steps: 1. Connect respiratory line and flow sensor. 2.
The T6 is equipped with a rechargeable lithium-ion battery, which is powered by a built-in battery through the T6 mainframe. The charging time shall not be less than 8 hours, and the working time under standard conditions shall not be less than 6 hours after full charging.
40%-60%. The battery shall be fully charged before use. Battery replacement 11.12.3 Make sure the T6 mainframe is in shutdown state; Open the battery lock in the direction of the arrow (see Fig. 57); Remove the dead battery from battery case (see Fig. 58);...
Fig. 11-2 Placement with a new battery Battery status description 11.12.4 The user can view whether the battery is connected, whether the battery is charging, the battery power and other information in the interface. Status of the battery is described as follows: Components Description The battery is not connected...
11.14 Storage If T6 is not used for a long period of time, following measures are recommended: Clean and disinfect (see Section 6 "Cleaning and disinfection").
12 T6 accessories Name/model Function Manufacturer Remarks Mainstream carbon Support CO Witleaf Purchased dioxide module M401B monitoring part Notices: ⚫ The accessories listed in this section are applicable to the ventilator. The hospital shall be responsible for ensuring the compatibility between ventilator and accessories.
13 Product specifications 13.1 Safety specifications Medical device management category Category Category III medical devices Electric shock protection Category I device, including internal power supply type Electric shock protection Defibrillation-proof BF type class Operating mode Continuous operation It shall not be used in the presence of flammable Degree of safety for anesthetic gas mixed with air or with oxygen or flammable anesthetic gas...
Function With touch screen Interface Network interface Support to connect to PC for software upgrade function Software of the ventilator can be upgraded via USB port, USB interface configuration information and historical data (e.g., trend data, logs, etc.) can also be exported via USB port, and configuration can be transferred between the machines with the same model via USB flash disk drive.
Battery type Lithium-ion battery Battery capacity 9600mAh Rated battery voltage DC 14.8V Minimum power supply 6h (a new fully charged battery operated in standard time operating conditions) 13.5 Gas supply specifications Gas supply specifications Gas supply Medical oxygen High-pressure gas 3.0-6.0 bar source pressure High-pressure pipe input...
L/min; No more than 6 cmH2O (adult) at a flow rate of 60 L/min; Expiratory pressure No more than 6 cmH2O (pediatric) at a flow rate of 30 L/min; No more than 6 cmH2O (infant ) at a flow rate of 5 L/min; Trigger mode Trigger mode Pressure trigger, flow trigger...
Page 105
2:1~1:4:±10% of set value; 4:1~1:10 Others:±15% of set value Upper pressure 10-100 cmH ± (2cmH O+ 5% of set value) limit ± (0.4 cmH2O + 10% of the setting Pressure trigger -20~-0.5 cmH value) Positive end 0-40cmH ± (0.9cmH2O + 5% of the setting expiratory value) pressure...
Page 106
Inspiratory pause 0%-60% Monitored parameters ±2bpm or ±5% of actual reading, Respiratory rate 0~250bpm whichever is greater Inspiratory tidal 0-3,000ml ± (2mL+ 15% of actual reading) (infant volume mode); ± (3mL+ 15% of actual reading) (pediatric mode); ±15% actual reading (adult mode) Expiratory tidal ±...
Notices: ⚫ When the ventilator is operated beyond the range specified by the manufacture, it may malfunction. Please ensure that the ventilator works under the specified working conditions, so as to maintain stable operation. ⚫ The total system response time of CO concentration is 1 sencond.
13.8 Gas line diagram F ig. 13-1 T6 Product Structure Diagram 13.9 Parts list Symbol NAME Symbol NAME Air filter cartridge Air inlet flow sensor Temperature sensor Primary acoustic box Turbine Secondary acoustic box Temperature sensor Check valve Fresh gas flow sensor...
13.10 Principle Description There are two types of oxygen supplies, including high pressure oxygen supply and low pressure oxygen supply: high pressure oxygen is connected via high pressure oxygen inlet 13; and low pressure oxygen is connected via low pressure oxygen inlet 14. Select one oxygen supply type: High pressure oxygen supply or low pressure oxygen supply.
⚫ The guidelines and manufacturer's statements are detailed in the annexes. Warnings: ⚫ The T6 ventilator shall not be used near or stacked with other devices. If it must be used near or stacked with any other devices then it should be observed to verify that it can operate normally under the configuration used.
14.2 Battery immunity declaration - requirements for all devices and systems Electromagnetic immunity declaration - requirements for all devices and systems The T6 ventilator can be used in following specific electromagnetic environments, and users shall ensure to use the device in following specified electromagnetic environments. Electromagnetic...
Page 112
Common mode: Common mode: ±2kV ±2kV Power frequency 3A/m 3A/m The power magnetic field frequency 50/60Hz magnetic field (50/60Hz) shall have the (GB/T 17626.8) level characteristics of power frequency magnetic field at a typical location in a typical commercial or medical environment.
14.3 Guidelines and manufacturer's statement - electromagnetic immunity Guidelines and manufacturer's statement - electromagnetic immunity The T6 ventilator is intended to be used in the electromagnetic environment specified below and the purchaser or user shall ensure that it is used in this electromagnetic environment.
Page 114
If the measured field intensity at the location of the T6 ventilator is higher than the applicable RF compliance level mentioned above, the T6 ventilator shall be observed to verify its normal operation.
Recommended isolation distance between portable and mobile radio frequency communication device and T6 ventilator The T6 ventilator is intended to be used in an electromagnetic environment in which the radio frequency radiation disturbance is controlled. The purchaser or user, depending on...
14.5 Basic EMC performance of T6 ventilator Basic EMC performance of T6 ventilator The T6 ventilator can work normally according to the parameter settings. For details, please see Chapter 11 of the Manual. The alarms can be given according to real-time monitoring on status of the T6 ventilator.
15 Product warranty Ambulanc (Shenzhen) Tech. Co., Ltd. can carry out free of charge maintenance for the product quality problems occurred during normal use, based on product instructions within two years from the date of purchase. If the warranty period indicated on the product is less than two years, the warranty will expire at the end of the expiration date indicated on the package or in the Operation Manual.
16 Classification details of toxic and harmful substances Name and content of toxic and harmful substances or elements Hexaval Polybro Polybrominat Mercu minated Lead ed diphenyl Part name mium chromiu bipheny (Pb) ethers (Cd) (Hg) (PBDE) Cr(VI) (PBB) Display screen ×...
17 Storage and transportation The pack products are allowed to be transported by road, air or rail. Shock and violent vibration shall be prevented during transportation. See the description in the table below: Graphic symbols Description Graphic symbols Description This side up Handle with care Stacking layer limit: Keep dry...
Need help?
Do you have a question about the T6 and is the answer not in the manual?
Questions and answers