Gaging Connectors; Connector Gage Accuracy - Agilent Technologies 85032B/E 50 User's And Service Manual

50ω type-n calibration kits
Table of Contents

Advertisement

Use, Maintenance, and Care of the Devices

Gaging Connectors

Gaging Connectors
The gages available from Agilent Technologies are intended for preventive maintenance
and troubleshooting purposes only. (See
information.) They are effective in detecting excessive center conductor protrusion or
recession, and conductor damage on DUTs, test accessories, and the calibration kit devices.
Do not use the gages for precise pin depth measurements.

Connector Gage Accuracy

The connector gages are only capable of performing coarse measurements. They do not
provide the degree of accuracy necessary to precisely measure the pin depth of the kit
devices. This is partially due to the repeatability uncertainties that are associated with the
measurement. Only the factory—through special gaging processes and electrical testing—
can accurately verify the mechanical characteristics of the devices.
With proper technique, however, the gages are useful in detecting gross pin depth errors on
device connectors. To achieve maximum accuracy, random errors must be reduced by
taking the average of at least three measurements having different gage orientations on
the connector. Even the resultant average can be in error by as much as
to systematic (biasing) errors usually resulting from worn gages and gage masters. The
information in
Table 2-2 on page 2-4
these systematic errors were not included in the uncertainty analysis. As the gages
undergo more use, the systematic errors can become more significant in the accuracy of the
measurement.
The measurement uncertainties (see
assembly materials and design, and the unique interaction each device type has with the
gage. Therefore, these uncertainties can vary among the different devices. For example,
note the difference between the uncertainties of the opens and shorts in
The observed pin depth limits in
typical factory pin depth values to provide practical limits that can be referenced when
using the gages. See
information on the design of the calibration devices in this kit.
NOTE
When measuring pin depth, the measured value (resultant average of three
or more measurements) is not the true value. Always compare the measured
value with the observed pin depth limits in
the condition of device connectors.
3-6
Table 6-3 on page 6-5
assumes new gages and gage masters. Therefore,
Table 2-2 on page
Table 2-2 on page 2-4
"Pin Depth" on page
2-4) are primarily a function of the
add these uncertainties to the
2-3. Refer to
"Kit Contents" on page 1-2
Table 2-2 on page 2-4
for part number
0.0001 inch due
Table
2-2.
for more
to evaluate
85032B/E