50ω type-n calibration kits
Table of Contents

Advertisement

Standard Definitions
Standard Class Assignments
Standard Class Assignments
Class assignment organizes calibration standards into a format compatible with the error
models used in the measurement calibration. A class or group of classes corresponds to the
systematic errors to be removed from the measured network analyzer response.
lists the classes of the devices in this calibration kit.
Table A-1
Standard Class Assignments
Calibration Kit Label:
Class
S
A
11
S
B
11
S
C
11
S
A
22
S
B
22
S
C
22
Forward Transmission
Reverse Transmission
Forward Match
Reverse Match
Response
Response and Isolation
A-2
N 50
A
B
C
2
8
1
7
3
2
8
1
7
3
4
4
4
4
1
7
2
1
7
2
D
E
F
G
8
4
8
4
Table A-1
Standard Class Label
Opens
Shorts
Load
Opens
Shorts
Load
Thru
Thru
Thru
Thru
Response
Response & Isolation