Device Description; Purpose; Versions; Structure - Lauda LRZ 914 Operation Manual

Interface module
Table of Contents

Advertisement

4

Device description

4.1

Purpose

4.2

Versions

4.3

Structure

Fig. 1: Contact module LRZ 914
10 / 29
The contact module is used to set or query binary states of the constant
temperature equipment by opening and closing circuits:
Output state (read): The module activates the circuit of the defined
interface function, see Ä Chapter 7.2.1 "Read commands" on page 20.
Set state (write): The user activates the circuit of the defined interface
function from outside the module, see Ä Chapter 7.2.2 "Write com-
mands" on page 21.
Choosing the most suitable version depends entirely on the scope of control
required:
Contact module with single input and output (NAMUR): One switching
function and one query can be configured for one input and one output,
which is enough to switch the constant temperature equipment between
operating and standby mode as well as monitor the temperature, for
example.
Contact module with 3 inputs and outputs: There are three inputs and
three outputs to which you can assign various switching functions and
queries.
Only one of the two versions can be operated for each item of
constant temperature equipment. This is particularly important to
note when integrating additional modules via the module box.
Contact output ( OUT ) and contact input ( IN ) on interface module
LRZ 914 are implemented using two 3‑pin connections.
1
Cover with holes for M3x10 fastening screws
2
Contact output, connector, 3-pin
3
Contact input, socket, 3-pin
Refer to Ä Chapter 6.1.1 "Contact module with single input and output
(NAMUR)" on page 15 for more information on contact assignment.
Interface module LRZ 914 / 915