5-3 Back Panel - SHOKO SCIENCE Shodex RI-504 Operation Manual

Refractive index detector
Table of Contents

Advertisement

5-3 Back Panel
20
19
18
17
16
15
No.
Name of parts
Communication port
15
[COM]
Zero in terminals
16
[ZERO IN]
Purge terminals
17
[PURGE IN]
Error out terminals
18
[ERROR OUT]
Leak out terminals
19
[LEAK OUT]
Ready out terminals
20
[READY OUT]
Recorder terminals
21
[REC. OUT]
Ground terminal
for the signal cable
22
[FG]
Integrator terminals
23
[INTEG. OUT]
21
22
23
Fig. 5.3 Back panel of Shodex RI-504
LAN communication port
When these two terminals are short-circuited just as when the zero
Button ( 7 ) is pressed, the automatic zeroing is actuated.
When these two terminals are short-circuited, the flow line of
theeluent is changed just as when the purge button ( 4 ) is
pressed,and both sample and reference cells are filled with the
same eluent.
A contact signal is sent out through these terminals when
an error occurs. (ROM, RAM, Parameter, Home Position,
Overheating, Optical Balance, Intensity)
A contact signal is sent out through these terminals when
an eluent leak is detected.
A contact signal is sent out through these terminals when
10 minutes have passed since the temperature of the optical
system reached the designated value.
Signals to the recorder are sent out through these terminals.
The sensitivity of the output signal is 10 mV/FS.
The shield terminal of the signal cable should be connected to
this terminal.
Signals to the data processing unit are sent out through these
terminals. The sensitivity of the output signal is 2 mV/μ RIU
or 8 mV/μ RIU with the integrator range of 512 μ RIU/V or
128 μ RIU/V, respectively.
- 10 -
Function
24
25