Setting up the AlphaTackPlus
Place the instrument on a heavy and stable table that is usable for weights of more than 50
kg. Furthermore the table should not be sensitive for resonance that can be affected by
vibrations of the instrument.
For reproducible tack results it is essential that the AlphaTackPlus has to be surrounded
completely with air of same temperature. Don't exposure the unit the direct influence of air
conditions and/or heating.
Don't place any electronic devices on top of the instrument. Because the top is warmed up
by inner electrical devices the self-cooling of a device placed on top is not granted, so that
malfunctions of this device are possible.
4.2.1.
Instrument cables and connections
The AlphaTackPlus measuring system needs only a main cable and a USB cable. If any
damage is detected exchange the hose/cable before the instrument is taken in further
operation. In case of doubt
Fehler! Textmarke nicht definiert.).
4.2.1.1.
Please only use cable from the manufacturer. These are tested and delivered in mechanical
stable version. Your socket must have a secure, electrical ground connection.
4.2.1.2.
Please only use cable from the manufacturer. These are tested and delivered in mechanical
stable version.
4.2.1.3.
The AlphaTackPlus can be switched on, if all connections to the instrument are made. In the
instrument display serial number, software version and operation time are shown.
Start WinAlphaTack now. The connection to the AlphaTackPlus is made automatically. You
indicate a fine connection when several instrument instructions are output within the
communication window on lower frame edge.
After installation and before a first test is started several adjustments must be proven
because they can be changed during transport.
contact
your supplier or Novomatics (please refer to page
Mains cable
USB cable
Switching on the AlphaTackPlus, Start of WinAlphaTack
AlphaTackPlus