This section explains how to install the G1978B multimode source on a 6100
Series Single Quad LC/MS instrument. The G1978A multimode source is not
supported on these instruments.
The 6100 Series Single Quad LC/MS with multimode source is supported on
ChemStation B.03.01 or higher, or the ChemStation B.01.03 with a patch.

Step 1. Prepare to install

The Multimode Enablement Kit, G1978-60451, is shipped with the multimode
source. This kit needs to be installed before the multimode source is used.
Note that the multimode source and its accessories are to be installed by an
Agilent Customer Engineer.
1 Check that the Multimode Enablement Kit contains the following parts:
Figure 1
From left to right: G1960-60858, G1960-61015 and G1960-60873
2 Install the APCI Enablement Kit, G1947-60451, which is shipped with the
8
Multimode Bd HV cable, p/n G1960-60858
Multimode HV PCA, p/n G1960-61015
Multimode Bd Power/Data Cable, p/n G1960-60873
multimode source.
The APCI Enablement kit contains the following parts:
Fast APCI HV Supply, p/n G1946-80058
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide