To convert from multimode to ESI, APCI or APPI

10 Place the source shipping cover on the source.
To convert from multimode to ESI, APCI or APPI
Never touch the source surfaces, especially when you analyze toxic substances or
WA R N I N G
when you use toxic solvents. The source has several sharp pieces which can pierce
your skin including the APCI corona needle, vaporizer sensor and counter current
electrode.
1 If the source to be installed is an APPI source, disconnect the multimode
2 Unscrew and remove the multimode spray shield with the field shaping
3 Install the new source and the standard spray shield, making sure that the
4 For APCI and APPI ion source, connect the vaporizer heater cable and the
5 For all sources, reconnect the nebulizer gas line tubing and the LC/MS
14
high voltage PCA RS-232 serial cable from the serial port B connector of
Smart Card.
electrodes.
hole in the spray shield is in the 12 o'clock position.
APCI high voltage cable. For the APPI source, connect the RS-232 cable to
the serial port B connector of Smart Card.
sample tubing.
Multimode Source for 6100 Series Single Quad LC/MS Set-Up Guide