Blood Glucose Monitoring System Blood Glucose Meter User Manual Safe AQ Smart、Safe AQ Voice、Safe AQ Air No. 265, Guyuan Road, Hi-Tech Zone, Changsha, Hunan Province, 410205,People’s Republic of China TEL:+86-731-89935581/89935582 FAX:+86-731-89825189 Email:info@sinocare.com Website:www.sinocare.com Shanghai International Holding Corp. GmbH (Europe) Eiffestraße 80, 20537 Hamburg, Germany.
Page 2
Dear Owner: Sinocare thanks you for choosing the Safe AQ Smart 、 Safe AQ Air 、 Safe AQ Voice blood glucose monitoring system. This user manual contains important information about the meter and how it works. Please read this user manual carefully before using your new meter, or the protection provided by the meter may be impaired.
XXI.Troubleshooting………………………………………………………………33 III.Patient Cautions …………………………………………………………………5 XXII.Explanation of symbols ………………………………………………………40 IV.Important Health-Related Information …………………………………………6 XXIII.Warranty………………………………………………………………………41 V.About Your Safe AQ Smart、Safe AQ Air、Safe AQ Voice Meter……………8 XXIV.References……………………………………………………………………41 VI.Full Screen Display ………………………………………………………………9 XXV.Index…………………………………………………………………………42 VII.Voice Function (For Safe AQ Voice only) …………………………………10 VIII.Pairing (Connecting) Meter with Smartphone (For Safe AQ Air only)……11 IX.Control Solution Testing…………………………………………………………12...
The Safe AQ Smart Blood Glucose Monitoring System includes: Safe AQ Smart blood glucose meter, Safe AQ blood glucose test strip, blood glucose control solution The Safe AQ Air Blood Glucose Monitoring System includes:...
Safe AQ test strip and lancet are for single use only. DO NOT REUSE. ▶ glucose result is displayed by your meter as mg/dL or mmol/L. Do not use other test strips and control solutions with the Safe AQ Smart、Safe AQ Air、Safe AQ Voice meter. ▶...
<140mg/dl (1-2 hours after starting a meal) FOR Safe AQ Smart, Safe AQ Air meter: 1. If test results greater than 600 mg/dL or 33.3mmol/L, the meter will display “HI”. Retest immediately with a new test strip. If your reading is still “HI”, contact healthcare professional immediately.
FOR Safe AQ Voice meter: V.About Your Safe AQ Smart、Safe AQ Air、Safe AQ Voice Meter 1. If test results greater than 600 mg/dL(33.3mmol/L), the meter will display Safe AQ Smart Meter: “HI” and read“The result is beyond the range”with red light on. Retest your blood Test strip port glucose immediately with a new test strip.
Safe AQ Voice Meter: VII.Voice Function (For Safe AQ Voice only) The Safe AQ Voice is preset as voice on. Slide the battery cover out of the meter. Remove the batteries. Voice switch is inside the battery chamber. Slide the voice switch to the “on” side, voice turns on and screen will display “...
Important Information phone and insert test strip again. • Use only Sinocare blood glucose control solution with your Meter. Note: Different App may be little different while pairing meter with smartphone. • Check the expiration date on the control solution bottle. Do not use if expired.
Page 10
• Discard any bottle that appears to be cracked or leaking. 2. Shake the control solution bottle well, then remove cap. Squeeze the bottle and • For in vitro diagnostic use only. discard the first drop. Squeeze the bottle again to get a second drop and bring the tip of the Safe AQ test strip to touch the drop of solution until the blood glucose meter beeps.
Questionable Results: X.Blood Glucose Testing Prepare meter, strip, lancing device and lancet before test. Probable Cause Action Operation mistake Retest according to correct steps STEP 1: Wash your hands or use an alcohol swab to clean the fingertip. MAKE SURE Control solution not being shaken well Shake control solution well and retest with YOUR HAND IS DRY BEFORE TESTING.
Page 12
Retain the lancet protector for safe lancet lancing disposal. The Sinocare lancing device is prepared and ready to lance your finger for a blood sample. STEP 5: Screw the lancing device cover back on. The lancet is for single use only. To prevent the risk of blood borne pathogen transmission, it shall not be shared with other people.
Page 13
For Safe AQ Smart, Safe AQ Air, if the meter display “E-2”; for Safe AQ Voice, if the meter display “E-2” and play “ The temperature is beyond the range”, it means the environment temperature is beyond the range.
Page 14
“The result is xx point xxx millimol per liter”. For Safe AQ Smart, Safe AQ Air, if the meter displays “E-3”; for Safe AQ Voice, if the meter displays “E-3” and plays “Test method error”, please read the blood glucose testing section carefully and use a new test strip test again.
XII.Memory After the blood glucose meter counts down from 5, The Safe AQ Smart、Safe AQ Air、Safe AQ Voice meter stores 10 blood glucose your blood glucose results appears along with the test results. You can review the individual results by entering the memory mode.
6. Wash hands thoroughly with soap and water after handling the meter, lancing device or test strips. • Only trained healthcare professional be allowed to use Safe AQ Smart、Safe AQ 7. Cleaning and disinfecting meter and lancing device before use on next patient.
XVII.Performance Characteristics meter can’t work properly, please contact customer service at +86-731-89935581/ +86-731-89935582 or contact local dealers. Accuracy: 95% test result of Safe AQ Smart, Safe AQ Air, Safe AQ Voice meet the requirement below: Concentration Range Bias% To disinfect the meter: <5.55 mmol/L (100 mg/dL)
Precision: test result of Safe AQ Smart, Safe AQ Air, Safe AQ Voice meet the requirement below: Memory 10 blood glucose test results Concentration Range Requirement Testing condition 10℃~35℃ ≤80% RH Construction <5.55 mmol/L (100 mg/dL) SD< 0.34 mmol/L (6.0 mg/dL)
XIX.Maintenance XX.Changing the batteries • Use of this instrument in a dry environment, especially if synthetic materials are Make sure your meter is off when you change the batteries. present (synthetic clothing, carpets etc.) may cause damaging electrostatic discharges that may cause erroneous results. Caution: Keep batteries away from children.
XXI.Troubleshooting Display What it means What to do Display What it means What to do Your result is below the Use a new test strip to test again, Test method error: insert strip into Insert test strip into meter before measurement limit, which is if meter still display “LO”, call meter after apply sample to strip.
Page 21
Call Sinocare customer service at The test strip package is damaged, Use test strip from a new vial. Test strip port malfunction +86-731-89935581/ and desiccant tablet becomes soft;...
Page 22
USE-BY date. (Only for Safe AQ Voice) result Meter or test strip malfunction Call Sinocare customer service at Test result is >70mg/dL +86-731-89935581/+86-731-89935582 (>3.9 mmol/L) and < 126 or contact your local dealers. mg/dL(< 7.0 mmol/L), but...
XXII.Explanation of symbols Consult instructions for use Keep away from sunlight Caution Serial number Temperature limit (Store at) Authorized representative in the European Community Manufacturer CE Marking and Notified Body Number In vitro diagnostic medical device Biological risks Symbol for the marking of electrical and electronics devices Fragile, handle with care according to Directive 2012 /19 / EC .
AQ Smart、Safe AQ Air、Safe AQ Voice blood glucose monitoring system: Control solution, test…..12 1. Changsha Sinocare Inc. warrants this meter to be free of defects in materials and Prepare, lancing device…..16 workmanship from the date of purchase. If the meter does not work properly, Sinocare will Test result, reviewing…..23...
Page 25
Thank you for choosing Safe AQ Smart、Safe AQ Air、 Safe AQ Voice blood glucose meter,a free replacement under normal operation is provided.Please do complete the warranty card and send it back to our agent or our factory.we will inform you with information of the meter.
Need help?
Do you have a question about the Safe AQ Smart and is the answer not in the manual?
Questions and answers
E 3 error has come up twice
The E 3 error on the Sinocare Safe AQ Smart means "Insufficient samples." To fix this, use a new test strip and test again.