Page 1
Blood Glucose and Uric Acid Monitoring System Blood Glucose and Meter User Manual Uric Acid Safe AQ UG For in vitro diagnostic use only...
Page 2
Safe AQ UG Dear Owner: Sinocare thanks you for choosing the Safe AQ UG blood glucose and uric acid monitoring system. This user manual contains important information about the meter and how it works. Please read this user manual carefully before using your new meter, or the protection provided by the meter may be impaired.
Important Health-Related Information Specifications About Your Safe AQ UG Meter Maintenance Button Functions Changing the batteries Setting Up Your Safe AQ UG Blood Glucose and Uric Acid meter Troubleshooting Control Solution Testing Explanation of symbols Blood Glucose and Uric Acid Testing Warranty...
The Safe AQ UG Blood Glucose and Uric Acid Monitoring System includes: Safe AQ UG blood glucose and uric acid meter, Safe AQ UG blood glucose test strip, Safe AQ UG uric acid test strip, blood glucose control solution, uric acid control solution Unit type: Safe AQ UG Blood Glucose and Uric Acid meter displays blood glucose results in mg/dL or mmol/L and uric acid results in mg/dL or μmol/L.
Important Safety Information Patient Cautions ▶ For self-testing customer, your Safe AQ UG meter and lancing device are for single person ▶Not for neonatal (newborn or infant) use use only. Do not share them with anyone else, even family members.
Page 6
professional immediately. Expected Results for people without diabetes: 4. If uric acid test results low than 181 μmol/L (3.0mg/dL), the meter will display “LO”. Plasma Blood Glucose Result Retest immediately with a new test strip. If your reading is still “LO”, contact healthcare Fasting Plasma Glucose <100 mg/dL (<5.55 mmol/L) professional immediately.
About Your Safe AQ UG Meter Full Screen Display Symbol for inappropriate Uric Acid Test Strip port: Blood Glucose Test Strip port: Uric acid test result testing temperature Insert the uric acid test strips Insert the blood glucose test here...
Function Action Setup steps for date/time The Safe AQ UG blood glucose and uric acid meter is preset with the date and time. You may To turn the meter on and enter the blood Press and hold for 3 seconds.
Page 9
sound. STEP 4: Setting the Year If there is an error, the meter will display E-4. Please pull out the code chip and insert again. If The year appears flashing at the left side of the still failed, please call Customer Service at +86-731-89935581/+86-731-89935582. meter display.
Page 10
STEP 5: Setting the Month STEP 6: Setting the Day The month appears flashing at the left side of The day appears flashing at the left side of the meter display. the meter display. 1. Setting the Month: Press and release ▼or▲ 1.
Page 11
STEP 7: Setting the Hour STEP 8: Setting the Minutes The hour appears flashing at the lower left side The minutes appear flashing at the lower left of the meter display. side of meter display. 1. Press and release ▼ or ▲ to decrease or 1.
● When you want to check whether the test steps are correct. Important Information ● Use only Sinocare control solution with your Meter. ● Check the expiration date on the control solution bottle. Do not use if expired. ● Use the blood glucose test strips within a period of 6 months from the date that you first...
Page 13
If the result is out of range, test again. If still so, do not use test strips or meter for Insert a Safe AQ UG Test Strip with printing face up and the contact end into the test strip testing.
Blood Glucose and Uric Acid Testing Questionable Results: Probable Cause Action Prepare meter, strip, lancing device and lancet before test. Operation mistake Retest according to correct steps STEP 1: Wash your hands or use an alcohol swab to Control solution not being shaken well Shake control solution well and retest with a new test strip.
Page 15
Retain the lancet protector for safe lancing lancet disposal. The Sinocare lancing device is prepared and ready to lance your finger for a blood STEP 5: Screw the lancing device cover back on. sample. The lancet is for single use only. To prevent the risk of blood borne pathogen transmission, it shall not be shared with other people.
Page 16
STEP 10:Take Blood Sample STEP 9: Insert strip into the test strip port, The Safe Make sure the lancing device is pressed firmly to AQ UG meter will beep and turn on. your finger. Press the button to lance. The port marked with “G” is for blood glucose strip and Attention: If you didn’t get enough blood sample, the port marked with “U”...
Page 17
STEP 14: Remove the lancing device cover. With Caution: the lancet protector on the table, stick the needle DO NOT smear or scrape the blood onto the Test Strip of the lancet into the protector. Eject the lancet DO NOT apply blood to the Test Strip when the Test Strip is not in the test strip port. and screw the lancet cover back on.
Viewing test result If blood glucose test results greater than 33.3mmol/L (600 mg/dL), the meter will display “HI”. Retest After the meter counts down from 5, your blood glucose results appear along with the unit of immediately with a new test strip. If your reading is measure.
Review uric acid test result: ● While the Safe AQ UG meter is turned off, press the The Safe AQ UG meter stores the 200 most recent results (100 blood glucose test results button for 3 seconds to turn on the meter, the screen μmol/L...
3 seconds to Press and hold ▼ for 3 seconds advance to the screen showing the last ● Only trained healthcare professional be allowed to use Safe AQ UG blood glucose and test result, and then press and ▼ the Show the latest blood glucose uric acid monitoring system on multiple patients.
When comparing results between Safe AQ UG blood glucose and uric acid monitoring system 3. Let meter air dry thoroughly before using to test. and a laboratory system, Safe AQ UG blood glucose and uric acid monitoring system blood 4. Wash hands thoroughly again after handing meter.
Power Saving Test result of uric acid meet the requirements below: If there is no blood applied to the test strip within 3 minutes, the meter will automatically Concentration Range Requirement turn off. ≤297 μmol/L(5 mg/dL) SD < 22.2 μmol/L(0.37mg/dL) After test, the test results will be displayed on the screen with test strip in strip port, If there >297 μmol/L(5 mg/dL) CV <...
Dimension 108*66*22(mm) Additional features Weight About 100g Auto electrode inserting detection Power source 3V DC, 10mA, 2 AAA alkaline batteries Auto sample loading detection Battery life perform up to 1,000 tests Auto reaction time count-down Display 500 blood glucose test results with date and time Temperature Warning 100 blood glucose control solution test results Battery power Warning...
Troubleshooting frequency interference at the frequency range and test levels specified by IEC 61326-2-6: 2012. Display What it means What to do ● Do not use this instrument in close proximity to sources of strong electromagnetic radiation, as these may interfere with the proper operation. Your blood glucose test Use a new test strip to test again, if ●...
Page 25
Display What it means What to do Display What it means What to do Your test result is higher than Use a new test strip to test again, if For blood glucose testing 20.0 mg/dL or 1188μmol/L. meter still display “HI”, call healthcare Test method error: insert strip Insert test strip into meter before professional immediately.
Page 26
Display What it means What to do Display What it means What to do Battery incorrectly installed Check that the battery is correctly Expired test strip or damped Use a new test strip to test again. installed. Pay attention to the “+”sign test strip Battery power is low Replace battery...
Page 27
Display What it means What to do Display What it means What to do Insufficient sample Repeat the test with a new test strip Expired test strip Use new test strip and enough blood sample The meter does not Meter displays Defective meter or test strip Contact the manufacturer or the dealer begin to test after...
Keep dry Consult instructions for use Safe AQ UG blood glucose and uric acid monitoring system: 1. Changsha Sinocare Inc. warrants this meter to be free of defects in materials and Caution Keep away from sunlight workmanship from the date of purchase. If the meter does not work properly, Sinocare will replace it with a new meter or equivalent product free of charge.
Index Thank you for choosing Safe AQ UG blood glucose meter, Hyperuricemia (HUA)……..1 a free replacement under normal operation is provided.Please do Reference value…..5 Control solution, test…..18 complete the warranty card and send it back to our agent or our Prepare, lancing device…..22...
Need help?
Do you have a question about the Safe AQ UG and is the answer not in the manual?
Questions and answers