Download Print this page

Troubleshooting - Breathe Pure Temp-A-Sure User Manual

Noncontact infrared thermometer

Advertisement

Troubleshooting

Problem
LCD Display "LO" or "HI"
No response when pressing
measurement button
No display or display Error
No beeper
Shut off right after switching on
Symbols
Definitions
BF type applied part
Refer to operation manual
Cautions, please refer to attached documents
Level of protecting against liquid inlet
Manufacturer
Date of manufacture
European union representative
Serial number
Separate collection
Keep dry
Keep away from sunlight
Device used within the Magnetic Resonance (MR)
environment is prohibited.
Solution
Breeze, water, sweating, cosmetic on forehead may affect
measurement.
Check deviation value setting. Defaulted value is 0.8° C.
While if the teasting environmental temp changes so enormously or
if the trmometer is used directly from high-temp object to very low-
temp one, the measurement difference will happen. The thermometer
should be kept in a relative stable environment for 10 minutes to get
heat balance before stating a new measurement.
Ensure measurement distance is 3 cm to 5 cm
Take out and reassemble battery
Check if the thermometer is under menu setup. In procedure of menu
setting, thermometer is unable to measure and therfore no response.
Take out battery and install battery again.
Check if beeper is switched off.
Check battery level or take out and install the battery again.