11.2.1.2 Avoid exposure to water, impact and excessive heat or direct exposure
to sunlight.
11.2.2
Care of the leads:
11.2.2.1Clean leads with a damp cloth if required.
11.2.2.2 Cables should not be wrapped around the unit, tangled or tightly rolled
when not in use.
12. Troubleshooting
Problem
Results Out Of Range
Faulty Electrode
Check Connection
LCD screen display is
blank
ECW% Fat Readings are
always incorrect
Strange characters on
LCD screen
PC-POD-CP-012-v01
Possible Cause
Device is not correctly set-up
Cable leads have not been
properly attached to the
electrodes
Battery power is dead
Malfunction of the electronic
memory micro processor
Malfunction of the liquid
crystal display
Printed copies are not controlled.
PC-POD-CP-012-v01
Solution
1. Recheck input data
2. Medications that result in
unnaturally high fluid levels.
1. Ensure participant is in
correct position
2. Recheck the electrodes.
3. Ensure leads are inserted
properly
4. Ensure leads are not tangled
5. Ensure alligator clips are
attached securely
6. Perform calibration
1. Check to see if batteries are
inserted properly
2. Replace with new Duracell
batteries
1. Perform calibration
2. Change batteries
3. Check cables
4. Return unit to dealer
1. LCD screen should be
replaced
Page 11 of 11