INSTRUCTIONS FOR USE BODYSTAT 1500MDD This “Instructions for Use” contains information on how to operate, storage and disposal BODYSTAT 1500MDD DUAL FREQUENCY BODY COMPOSITION & WELLNESS MONITORING UNIT Version 9/24 Date of Issue: 08/02/2024...
SECTION B HOW TO USE BODYSTAT 1500MDD ..8 INTRODUCTION ............................9 WHAT DOES THE 1500MDD UNIT MEASURE?..................10 IMPORTANT GUIDELINES FOR EQUIPMENT CARE ................11 OPERATIONAL FEATURES ........................12 CONNECTING THE CABLE LEADS ......................13 INSTALLATION MODE (OPTIONS) ......................14 BODYSTAT MEASUREMENT ........................
YOUR HARDWARE UNIT The BODYSTAT 1500 MDD package consists of the following: Bioelectrical Impedance Dual Frequency Measuring Unit • & Black 1.2 metre Cable Leads (2) • Hardware User’s Guide • Disposable Electrodes 300) • Duracell/Procell AA Batteries 1.5 V (6) •...
Bio-Electrical Impedance Analysis Fat –Free Mass (Lean Weight) FFMI Fat-Free Mass BFMI Body Fat Mass Index TRANSLATIONS If this BODYSTAT 1500MDD User Guide is required in any of the following languages, please email info@bodystat.com: Bulgarian English Hungarian Russian Czech Finnish...
Page 6
Although all BODYSTAT hardware units and software programs are tested before release, no claim is made concerning the accuracy of the product. BODYSTAT LTD, as the manufacturers of the product and authors of the software, their distributors, and dealers, cannot assume liability or responsibility either implied or expressed for any loss or damage either real or consequential arising out of the use of this product.
As with any other common electronic device, the BODYSTAT 1500MDD unit of Bodystat Limited emits electromagnetic energy to perform its intended function. Nearby electronic equipment may be affected. The device is not suitable for oxygen-enriched environments. The device is not suitable for use with flammable gases, including environment with flamable anesthetics.
INTRODUCTION The BODYSTAT® 1500 MDD body composition monitoring unit is designed to provide a fast and effective method of measuring body composition using modern technology and is ideal for use in the medical/clinical environment. Using the scientifically validated principle of...
WHAT DOES THE 1500MDD UNIT MEASURE? The Bodystat® 1500MDD measures/estimates: Options Displayed on the Explanation 1500MDD unit & Normal Range Fat as a % of Total Body Weight Weight* & Normal Range Est. FAT Weight determines Health Risk Lean & Normal Range...
Therefore, it is recommended that only leads supplied by Bodystat Limited are to be used. • You should be aware that computers and other electrical equipment can cause interference which may adversely affect the impedance measurement.
The most recent 100 tests are always automatically stored in the memory of the ® BODYSTAT 1500MDD unit and subject data may be recalled at any time. Data stored in memory is not lost after the unit is switched off (refer...
IMPORTANT GUIDELINES FOR EQUIPMENT CARE. The BODYSTAT 1500MDD unit has two main cable leads. Each lead is to be inserted to the connectors at the rear of the unit. The red dot on the lead wire should be lined up with the red dot on the device.
INSTALLATION MODE (OPTIONS) Installation Mode comprises the following: • DISPLAY OPTIONS • IMPERIAL / METRIC OPTIONS • MID-RANGE OPTIONS • LANGUAGES OPTIONS This facility allows the operator to choose: Information required to be entered, e.g., Gender, Age, Height & Weight Information required for display,...
Page 15
Press the <> to change the appropriate unit selection and once complete, press the proceed button to confirm. After pressing the proceed button, the selection reverts to the 4 options for Units, Inputs, Outputs or Printer. Further selections can be made or alternatively, press the home button to return to the main screen.
Page 16
Activity Level options: The activity level has no effect on On-screen Description the body composition or V Low Very Low hydrational/nutritional results. It is Low/Medium used only in the estimation of the Medium Medium/High Activity Energy Requirement. V High Very High Range (Min/Max) Age 1 to 99 years Height 1' to 7'6"...
Page 17
PRINTER OPTIONS ® The Bodystat 1500MDD device has the option of printing to a thermal printer. When using the printer for the first time, it is important that the printer is first ® paired with the Bodystat 1500MDD device.
Page 18
The Bodystat device will then continue to search for the printer. If the device does not find the printer, ensure that the printer is switched on. Once the device has been paired with the printer, a message will be displayed confirming the pairing has been completed.
NOT recommended for subjects with pacemakers or any implantable electronic device. With the application of our single frequency BODYSTAT 1500MDD unit, certain pre-test protocols are recommended. However, in the medical environment it is unlikely that these requirements necessarily apply and will vary from circumstance to circumstance.
Page 20
RIGHT foot, in order to avoid the battery current (low voltage) passing through the side of the body where the heart is situated. The BODYSTAT® 1500MDD has two sets of main leads which are then divided into a and a BLACK lead. It does NOT matter which of the two sets of main leads is connected to the right hand or right foot.
Page 21
Attach the crocodile clips to the exposed metal tab strip of the electrode. For convenience, the metal tab on the electrode should point in the direction from which the leadwire tension will occur as illustrated. The BLACK (measuring) leads are then connected to the electrodes on the right wrist and right ankle.
Page 22
Only the best quality BODYSTAT electrodes as those supplied by • BODYSTAT LTD, or their accredited distributors, should be used and used only once. This is to ensure accuracy of measurement and good reproducibility of tests.
ENTER SUBJECT DATA Switch the BODYSTAT® 1500MDD unit ON. No message will be displayed on the touch screen LCD screen if the battery is flat, in which case replace the batteries with new Duracell/Procell Batteries. Select the New Test button on the screen to perform a Bodystat measurement.
Page 24
defaulted option for those results (unless the Waist Hip Ratio has been omitted in the options menu). If the cursor is flashing on the weight option and the next button has been, the next screen showing waist/hip and activity level is displayed. The first option is waist, followed by hip.
Page 25
Once all options have been entered, press the proceed button to move to the measurement screen. Alternatively, press the red abort < > button to exit the screen without measuring. The next screen displayed is for connecting the electrodes. The unit has already been connected to the four surface electrodes as fully explained in the PERFORMING A MEASUREMENT section of this manual and the required...
Page 26
Wait a few seconds for the results to be displayed on the LCD screen. The data has now automatically been saved and stored enabling both input and output subject data to be recalled later. The electrodes and leads may at this stage be removed. The next screen that is automatically displayed is dependent on how the operator has set up the user selectable DISPLAY OPTIONS...
Page 27
DISPLAY RESULTS After successful measurement and processing of the subject data, the results are displayed as selected by the operator. The first screen displayed is the test summary screen which shows the Wellness Marker, Impedance, Resistance, Reactance and Phase Angle at 50 kHz. Once the summary screen has been displayed, a new test can now be performed by selecting the new test button.
Page 28
Each result displays the Estimated Measured Result. Underneath the result displays the Recommended Normal Level (where appropriate) for the subject measured. Select the Next button to move to the next results screen or alternatively press the summary button to move back to the summary page. The display results are automatically stored in memory and the operator can safely scroll forwards through the data by using the <>...
Page 29
However, it would be more efficient to download the data into our Software Program supplied with the BODYSTAT® 1500MDD unit and then in turn to easily save the subject data in Excel, XML or CSV files.
Page 30
NOTE: Tests are automatically stored in memory after subject data has been entered and the BODYSTAT 1500MDD unit has successfully performed a measurement. Data is not lost after the unit is switched off or during battery replacement. To access this recall option, select the recall data button from the home screen.
Page 31
Press the home button to return to the main screen. The first screen displayed is the input data that was entered for that test. By pressing the next button, the result output data is then displayed. Once all the results have been reviewed, the user has the option of selecting the back arrow or alternatively going back to summary screen.
INFORMATION ABOUT THE DEVICE Access to the BODYSTAT® 1500MDD unit serial number may be necessary from time to time. This will be required for warranty purposes, etc. The serial number may be accessed as follows: While switching the unit and then select the information button from the home screen.
There are two regression equations or formulas used in the program of the BODYSTAT 1500 MDD unit for the determination of body composition. One is suited for the age groups 18 to 70 years and is found in our popular ...
Page 34
Water is a good conductor of electricity and the less fluid in the body the higher therefore the resistance factor in the body. This results in a higher Impedance value being registered by the BODYSTAT 1500 MDD unit and hence higher fat % figures and lower body fluid volumes.
24-hour period should be made when entering the data in the BODYSTAT ® 1500MDD unit. This can only be a subjective assessment, but the following table may help: ACTIVITY LEVEL Level of Activity...
4. If the unit remains outside the calibration range, replace ALL batteries with ONLY Duracell® Procell® batteries RECHARGABLES). 5. If the unit still remains out of range, contact BODYSTAT LTD or your local BODYSTAT dealer. Bodystat Calibrator Red & Black Crocodile Clips Impedance to read between 496 and 503...
If there is no activity on the device for 3 minutes, then system will turn itself off to conserve battery power. It is highly recommended that the batteries should be removed from the Bodystat unit should the unit not be used for some time.
Refer to the BODYSTAT web page for the principles of Bioelectrical Impedance Analysis (BIA), validation of the accuracy of the BODYSTAT technology, the most recent updated BODYSTAT research papers and its use in a variety of applications. CONTACTS: For Sales Enquiries –...
RoHS compliance (please refer to our website for details) Cardboard packaging and another packing/storage material should be recycled where possible. At the end of the life cycle of your Bodystat device, it is important that all batteries are removed & recycled according to the battery manufacturer’s instructions.
Table 1 - Guidance and manufacturers declaration – electromagnetic immunity The Bodystat 1500 is intended for use in the electromagnetic environment specified below. The customer or the user of the Bodystat 1500MDD should assure that it is used in such an environment. Emissions test...
Page 43
Table 2 - Guidance and manufacturers declaration – electromagnetic immunity The Bodystat 1500MDD is intended for use in the electromagnetic environment specified below. The customer or the user of the Bodystat 1500MDD should assure that it is used in such an environment.
Page 44
Table 3 - Guidance and manufacturers declaration – electromagnetic immunity The Bodystat 1500MDD is intended for use in the electromagnetic environment specified below. The customer or the user of the Bodystat 1500MDD should assure that it is used in such an environment Immunity Test...
Page 45
PB840 Ventilator is used exceeds the applicable RF compliance level above, the Bodystat 1500MDD should be observed to verify normal operation. If abnormal performance is observed, additional measures may be necessary, such as re-orientating or relocating the Bodystat 1500MDD.