POL-EKO SMART CLN 15 Instruction Manual

POL-EKO SMART CLN 15 Instruction Manual

Laboratory incubators, drying ovens, drying ovens with nitrogen blow, laboratory sterilizers, pass-through sterilizers
Table of Contents

Advertisement

Instruction manual SMART
models:
CLN 15, CLN 32, CLN 53, CLN 115, CLN 180, CLN 240
CLW 15, CLW 32, CLW 53, CLW 115, CLW 180, CLW 240, CLW 400,
CLW 750, CLW 1000
models:
SLN 15, SLN 32, SLN 53, SLN 75, SLN 115, SLN 180, SLN 240
SLW 15, SLW 32, SLW 53, SLW 75, SLW 115, SLW 180, SLW 240,
SLW 400, SLW 750, SLW 1000
Drying ovens with nitrogen blow
models:
SLWN1 15, SLWN1 32, SLWN1 53, SLWN1 115, SLWN1 240
SLWN2 15, SLWN2 32, SLWN2 53, SLWN2 115, SLWN2 240
models:
SRN 53, SRN 115, SRN 240
SRW 53, SRW 115, SRW 240, SRW 400, SRW 750, SRW 1000
Pass-through sterilizers
models:
SRWP 115, SRWP 240
Before using the equipment, please read carefully this instruction manual!
Laboratory incubators
Drying ovens
Laboratory sterilizers
Version 1.24
Issued 07.05.2021

Advertisement

Table of Contents
loading
Need help?

Need help?

Do you have a question about the SMART CLN 15 and is the answer not in the manual?

Questions and answers

Subscribe to Our Youtube Channel

Summary of Contents for POL-EKO SMART CLN 15

  • Page 1 Instruction manual SMART Laboratory incubators models: CLN 15, CLN 32, CLN 53, CLN 115, CLN 180, CLN 240 CLW 15, CLW 32, CLW 53, CLW 115, CLW 180, CLW 240, CLW 400, CLW 750, CLW 1000 Drying ovens models: SLN 15, SLN 32, SLN 53, SLN 75, SLN 115, SLN 180, SLN 240 SLW 15, SLW 32, SLW 53, SLW 75, SLW 115, SLW 180, SLW 240, SLW 400, SLW 750, SLW 1000 Drying ovens with nitrogen blow...
  • Page 2 Instruction manual CL, SL, SLWN, SR, SRWP SMART Manufacturer's address: POL-EKO-APARATURA sp. j. ul. Kokoszycka 172 C 44-300 Wodzisław Śląski Country of origin: Polska As a manufacturer, we inform you that we took the necessary measures to ensure that this device fully meets your expectations and is reliable for a long period of use.
  • Page 3: Table Of Contents

    Instruction manual CL, SL, SLWN, SR, SRWP SMART Contents INTENDED USE AND IMPORTANT INFORMATION FOR THE USER ............6 PACKAGE CONTENTS ..........................7 BEFORE THE FIRST USE ..........................8 3.1. Installation of shelves ........................... 10 3.2. Remarks on the placement of samples ....................10 3.3.
  • Page 4 Instruction manual CL, SL, SLWN, SR, SRWP SMART 6.7. Quick Change of parameters ........................ 39 6.7.1. Quick change of set temperature ..................... 39 6.7.2. Quick change of set time ........................40 6.7.3. Quick change of fan efficiency and degree of air-flap opening ............41 6.8.
  • Page 5 Instruction manual CL, SL, SLWN, SR, SRWP SMART 12.1. Exterior cleaning ........................... 66 12.2. Interior cleaning ............................ 66 12.3. Cleaning the touch screen ........................67 12.4. Consumables ............................68 13. ADVICE ON HOW TO SAFELY STORE THE DEVICE ................68 14.
  • Page 6: Intended Use And Important Information For The User

    Instruction manual CL, SL, SLWN, SR, SRWP SMART 1. INTENDED USE AND IMPORTANT INFORMATION FOR THE USER CL laboratory incubators, SL drying ovens, SR laboratory sterilizers are laboratory devices designed for incubation, drying and sterilization with hot air of samples at following temperature ranges: •...
  • Page 7: Package Contents

    Instruction manual CL, SL, SLWN, SR, SRWP SMART • overload the shelves (the maximum load is described in technical data). You should: • place samples in such a way to provide proper air circulation in the chamber, • open the door for the shortest period of time to reduce temperature fluctuations, •...
  • Page 8: Before The First Use

    Instruction manual CL, SL, SLWN, SR, SRWP SMART Pass-through sterilizers (SRWP) in SMART version are delivered with: SRWP Device Capacity Shelves [pcs.] Slides [pcs.] Power cord [pcs.] Rubber cap [pcs.] Key for door lock [pcs.] Stand Masking element Quality Control Certificate [pcs.] 3.
  • Page 9 Instruction manual CL, SL, SLWN, SR, SRWP SMART If you don’t comply with the above recommendations, the unit may get broken or it may worsen the technical parameters such as: • temperature fluctuation, • temperature variation, • power consumption and may result in loss of warranty. Wheels / leveling feet If the device has been equipped with wheels or leveling feet: •...
  • Page 10: Installation Of Shelves

    Instruction manual CL, SL, SLWN, SR, SRWP SMART 3.1. Installation of shelves To install the shelf or to change its position, follow these steps: Install the shelf slide at the selected height by inserting it into perforations on the wall of the device. Do the same with the slide on the opposite wall.
  • Page 11: Closing Chamber Door

    Instruction manual CL, SL, SLWN, SR, SRWP SMART Front view Shelf Hmax = 2/3H Top view u+v+w > 1/3 a x+y+z > 1/3 b u v w x y z Following the above rules will provide best optimal parameters of temperature fluctuation and variation. 3.3.
  • Page 12 Instruction manual CL, SL, SLWN, SR, SRWP SMART Front view rating plate access port ø30mm for external sensor internal glass door (for CL) adjustable feet shelf external solid door temperature sensors open door sensor touch control panel 10) USB port 11) handle with key lock...
  • Page 13 Instruction manual CL, SL, SLWN, SR, SRWP SMART Rear view main switch connector for connecting a hose with nitrogen (only for SLWN models) air-flap LAN port fuse main power socket C20...
  • Page 14: Design Of Srwp Devices

    Instruction manual CL, SL, SLWN, SR, SRWP SMART 4.2. Design of SRWP devices Below there’s a picture of pass-through sterilizer (exemplary photo) with a description of the important components of the device. Front view main power socket C20 rating plate access port ø30mm for external sensor external solid door with handle with a key lock touch control panel...
  • Page 15 Instruction manual CL, SL, SLWN, SR, SRWP SMART Rear view green diode – lights up when the device is turned on yellow diode - pulsates when the set temperature is being reached; lights up when the set temperature is reached red diode - it lights up in the event of an alarm external solid door handle with a key lock...
  • Page 16: Device Equipment (Standard And Optional)

    Instruction manual CL, SL, SLWN, SR, SRWP SMART 5. DEVICE EQUIPMENT (STANDARD AND OPTIONAL) 5.1. External door with viewing window (optionally for CL, SL, SR) The external door with a viewing window is an optional equipment for CL, SL, SR models (except for CL/SL 15,32). In SL drying ovens with door with viewing window, the maximum operating temperature is lowered to +250°C at the factory.
  • Page 17: Internal Socket (Optionally For Cl)

    Instruction manual CL, SL, SLWN, SR, SRWP SMART 5.4. Internal socket (optionally for CL) An internal socket with grounding and IP44 protection is an optional equipment for CL laboratory incubators. The internal socket (depending on the model: 230V 50Hz / 230V 60Hz / 115V 60Hz) is intended for EU plugs or plugs type B.
  • Page 18: Access Port For External Sensor (Standard For All Units)

    Instruction manual CL, SL, SLWN, SR, SRWP SMART 5.6. Access port for external sensor (standard for all units) A Ø30 mm access port can be used to insert an external temperature sensor for independent temperature control inside the device. The access port has been secured with a rubber plug. The plug should cover the access port while the unit is operating.
  • Page 19: Usb Port (Standard For All Units)

    Instruction manual CL, SL, SLWN, SR, SRWP SMART 5.9. USB port (standard for all units) The USB port on the front panel is used only to transfer data from the device's internal memory to the flash drive. To do this, insert the flash drive into the USB port on the front panel and then: •...
  • Page 20: Device Operation

    Instruction manual CL, SL, SLWN, SR, SRWP SMART Batteries should be replaced every 12 months. During the warranty period, the replacement should be performed by an authorized service. Otherwise, you will lose your warranty. 6. DEVICE OPERATION This symbol means that a given window can be moved in the direction shown in the picture.
  • Page 21: Main Screen

    Instruction manual CL, SL, SLWN, SR, SRWP SMART Figure 2 Downloading files 6.3. Main screen After switching on the device, the main screen (Figure 3) appears. It contains the information about the device status. After starting the program, additional information appears on the screen (Figure Figure 3 Main screen...
  • Page 22: Information Panel

    Instruction manual CL, SL, SLWN, SR, SRWP SMART Figure 4 Main screen – running program main segment quick change of start/stop menu temperature duration time fan efficiency program current temperature in the chamber alarm bar side-scrolling information panel current program active date/hour status...
  • Page 23 Instruction manual CL, SL, SLWN, SR, SRWP SMART Figure 6 Icon: Alarms panel there’s a list with active alarms or the alarms that have occurred but have not been con- In the alarms panel (Figure 7) firmed. When the alarm is active, the alarm bar is red and the alarm event is displayed in the list with the status "active". When the alarm event stops, the state changes to "inactive".
  • Page 24: The Meaning Of Icons And Symbols

    Instruction manual CL, SL, SLWN, SR, SRWP SMART Figure 9 Icon: Statusu panel The status of the device is indicated also by description. Figure 10 Status – description program name the name of running program current segment currently running segment / total number of segments in the program priority of time or parameters time set...
  • Page 25 Instruction manual CL, SL, SLWN, SR, SRWP SMART Automatic screen lock. Factory setting: disabled. Unmounting the USB flash drive before removing it from the USB socket. Internal light is switched on. The icon appears in the devices with the internal light (OWW).
  • Page 26: Upper Expandable And Configurable Menu

    Instruction manual CL, SL, SLWN, SR, SRWP SMART Removing selected program from the list. Going to alarms panel. Going to the status panel with information about the program parameters. Going to the menu to create, edit, delete and start programs. Canceling adding or editing of the program.
  • Page 27 Instruction manual CL, SL, SLWN, SR, SRWP SMART Figure 11 Upper menu when the program is running Figure 12 Upper menu when the program is stopped Positions available on the upper bar can be personalized. Just drag the selected icon to a new location (Figure 13) Figure 13 Changing icon´s position...
  • Page 28: Alarm Bar

    Instruction manual CL, SL, SLWN, SR, SRWP SMART 6.3.4. Alarm bar The Alrm Bar is a quick visual information about the device status. The colour of the bar indicates the status of the device: – blue - the device is working properly –...
  • Page 29 Instruction manual CL, SL, SLWN, SR, SRWP SMART In order to go to Quick program click the icon in the main screen. By clicking the appropriate icon you can set: − temperature (Figure 15) − fan efficiency − degree of opening of the air-flap Clicking the icon starts the program in continuous mode (time set to infinity).
  • Page 30: Programs

    Instruction manual CL, SL, SLWN, SR, SRWP SMART Figure 16 Quick program on programs list In Quick Program editing mode, you can change: • settings of the data recording interval, • settings of the protection class. When the program is running you can change the parameters (temperature, air-flap and fan if available) by pressing the icon .
  • Page 31: Creating / Editing A Program

    Instruction manual CL, SL, SLWN, SR, SRWP SMART Figure 17 List of programs Information on the number of created programs / the maximum number of programs that can be created is at the top of the screen (programs: 5/5). 6.5.1. Creating / editing a program Press the button , and a panel with program parameters will appear (Figure...
  • Page 32: Segments Edition

    Instruction manual CL, SL, SLWN, SR, SRWP SMART Going to the edition of program segments. With more parameters, the window can be scrolled up and down. 6.5.2. Segments edition For each of the 5 programs, you can set maximum 6-segment time-temperature profiles that allow you to gradually increase or decrease the incubation temperature of the samples.
  • Page 33 Instruction manual CL, SL, SLWN, SR, SRWP SMART With more parameters, the window can be scrolled up and down. Figure 19 Program segment edition The navigation between: segments, program parameters and summary is done by swiping your finger to the side or by touching the icon If, when editing a program, you automatically return to the home screen, the edited program will not be lost, but saved as a draft (see below).
  • Page 34 Instruction manual CL, SL, SLWN, SR, SRWP SMART Figure 20 6.5.2.1. Summary of segments In the segments summary (Figure 21) all segments can be seen along with introduced parameters: • number of segments, • temperature, duration time, target time of reaching temperature of a given segment, •...
  • Page 35: Protection Class

    Instruction manual CL, SL, SLWN, SR, SRWP SMART 6.5.3. Protection class The device is equipped with sample protection – temperature protection which is carried out on the basis of the temperature value measured by an independent temperature sensor, the so-called security sensor. The main aim of sample protection is to protect against uncontrolled rise or fall of temperature.
  • Page 36: Loop

    Instruction manual CL, SL, SLWN, SR, SRWP SMART 6.5.6. Loop The option is available if the number of segments is equal to 2 or more. When the program finishes the last segment, the device starts the program again from the first segment. You can define if the program should be carried out once (loop: 1) or multiple times (loop: 2 to 255).
  • Page 37 Instruction manual CL, SL, SLWN, SR, SRWP SMART Figure 23 Main menu Figure 24 Program management menu If the program is running, the symbol appears next to the program name on the list. The symbol means that the program has been edited, but the changes have not been confirmed (Figure 25).
  • Page 38: The Second Way

    Instruction manual CL, SL, SLWN, SR, SRWP SMART 6.6.2. The second way • In the main screen (Figure 26) press the icon in the upper right corner • Select the program you want to start (Figure 27). You have two additional options to start the program: Immediate start of the program.
  • Page 39: Quick Change Of Parameters

    Instruction manual CL, SL, SLWN, SR, SRWP SMART 6.7. Quick Change of parameters Although the ramp time has been included in the program, the Quick Change of parameters will take place immediately while the temperature is being reached. 6.7.1. Quick change of set temperature In order to quickly change the value of set temperature of a running program, press the icon in the main screen (Figure...
  • Page 40: Quick Change Of Set Time

    Instruction manual CL, SL, SLWN, SR, SRWP SMART 6.7.2. Quick change of set time In order to quickly change the duration time of a running program, press the icon in the main screen (Figure 30). Select the number of days, hours and minutes by scrolling the list up or down (Figure 31).
  • Page 41: Quick Change Of Fan Efficiency And Degree Of Air-Flap Opening

    Instruction manual CL, SL, SLWN, SR, SRWP SMART 6.7.3. Quick change of fan efficiency and degree of air-flap opening In order to quickly change the fan efficiency and/or degree of air-flap opening, press the icon in the main screen (Figure 32).
  • Page 42: Data Record

    Instruction manual CL, SL, SLWN, SR, SRWP SMART • average temperature [°C] – average temperature • segment – status of the segment: in progress – currently running segment (data is being constantly updated), finished – the segment has been completed, interrupted –...
  • Page 43: Event Log

    Instruction manual CL, SL, SLWN, SR, SRWP SMART Figure 35 Data record Press to display the rest of the data. Recording data onto the USB flash drive. .csv files are available - separated by semicolon when opening e.g. with a spreadsheet, .plkx - opening with the Lab Desk application (option) Before removing the USB flash drive from the USB port, it must be unmounted, see Section 5.9.
  • Page 44 Instruction manual CL, SL, SLWN, SR, SRWP SMART Figure 38 Event log Recording data onto the USB flash drive. .csv files are available - separated by semicolon when opening e.g. with a spreadsheet, .plkx - opening with the LabDesk application (option) Before removing the USB flash drive from the USB port, it must be unmounted, see Section 5.9.
  • Page 45 Instruction manual CL, SL, SLWN, SR, SRWP SMART Information event Alarm event Error Warning Possible events: Program Start starting the program Program Stop stopping the program Program Edit changing the program parameters Program End program is completed DeviceOn the device is switched on (on the main switch) DeviceOff the device is switched off (on the main switch) Door opened...
  • Page 46: Info

    Instruction manual CL, SL, SLWN, SR, SRWP SMART 6.11. Info Go to the main menu and press the icon . The panel contains the following information: • name of device, • device’s temperature range • serial number of the device •...
  • Page 47: Unlocking The Touch Screen

    Instruction manual CL, SL, SLWN, SR, SRWP SMART Turn on/off the sound . Critical alarms will continue emitting a sound. Set the time after which the user will return to the home screen. Available settings: off, 1 min, 3 min, 5 min, 10 min.
  • Page 48: Time

    Instruction manual CL, SL, SLWN, SR, SRWP SMART 6.13. Time Go to the main menu and press the icon . In this panel you can: • change the date / system time If the date / system time is changed to the later date / time comparing with the data and events which are stored in the memory, they will remain in the register.
  • Page 49: Alarms

    Instruction manual CL, SL, SLWN, SR, SRWP SMART 6.14. Alarms Go to the main menu and press the icon . You can set parameters related to alarms. • lower alarm – an alarm will be activated if the temperature drops below the value given in this field, •...
  • Page 50: Network

    Instruction manual CL, SL, SLWN, SR, SRWP SMART Figure 46 Setting the time of mute function 6.15. Network Go to the main menu and press the icon . In this panel (Figure 47) you can change LAN settings: • IP – the device’s IP address •...
  • Page 51: Corrections

    Instruction manual CL, SL, SLWN, SR, SRWP SMART Figure 47 – LAN settings Confirms changes Cancels the entered changes 6.16. Corrections Go to the main menu and press the icon . In this window (Figure 48) you can correct temperature value indicated on the display by adding the correction value.
  • Page 52: Archives

    Instruction manual CL, SL, SLWN, SR, SRWP SMART Confirms changes Cancels the entered changes 6.17. Archives Go to the main menu and press the icon . The archive folder contains data and events transferred there while changing the date / time to an earlier one than the date / time with which they have been stored in the device memory.
  • Page 53: Temperature Protection

    Instruction manual CL, SL, SLWN, SR, SRWP SMART register INPUT REGISTERS function READ_INPUT_REGISTERS (0x30001) Address Type Multiplier Description temperature from the main sensor bool Open door b0 – door alarm b1 – upper temperature alarm b2 – lower temperature alarm b3 –...
  • Page 54: Slwn Drying Ovens With Nitrogen Blow

    Instruction manual CL, SL, SLWN, SR, SRWP SMART 9. SLWN DRYING OVENS WITH NITROGEN BLOW According to the requirements of the PN-ISO 589:2006 norm which refers to the determination of total moisture in hard coal requires that coal, the samples susceptible to oxidation should be dried at a temperature of + 105°C in a drying oven with a nitrogen flow.
  • Page 55 Instruction manual CL, SL, SLWN, SR, SRWP SMART inlet pressure gauge (high pressure) connector to connect to the cylinder pressure adjustment screw outlet pressure gauge (low working pressure) shut-off valve end for the hose connection Reducer scale of the pressure gauge flow control valve knob end for the hose connection Laboratory rotameter...
  • Page 56: Connection

    Instruction manual CL, SL, SLWN, SR, SRWP SMART scale of the pressure gauge flow control valve knob end for the hose connection Technical rotameter 9.2. Connection Before using: • check carefully the condition of the reducer, especially the inlet fitting, connecting nut, pressure gauge, rota- meter and safety valves, •...
  • Page 57: Connection Of The Kit B

    Instruction manual CL, SL, SLWN, SR, SRWP SMART 9.2.2. Connection of the Kit B In order to connect the Kit B you have to: The supplied hose cut into two parts, the length should be chosen so it will be possible to connect the cylinder with the rotameter and the rotameter with the equipment.
  • Page 58: Connection Of The Kit C

    Instruction manual CL, SL, SLWN, SR, SRWP SMART 9.2.3. Connection of the Kit C In order to connect the Kit C you have to: The supplied hose cut into two parts, the length should be chosen so it will be possible to connect the cylinder with the rotameter and the rotameter with the equipment.
  • Page 59: Shut-Off Solenoid Valve For Gas Supply (Optionally)

    Instruction manual CL, SL, SLWN, SR, SRWP SMART 9.3. Shut-off solenoid valve for gas supply (optionally) The solenoid valve opens the gas supply to the equipment at the beginning of the program and closes it when the program is finished. If the program has been started with a delay, while waiting for the start of the first segment of the program, the gas supply is closed.
  • Page 60: Methodology For Calculating The Nitrogen Flow Rate In The Drying Oven

    Instruction manual CL, SL, SLWN, SR, SRWP SMART 9.5. Methodology for calculating the nitrogen flow rate in the drying oven Calculation of nitrogen flow rate in drying ovens intended for drying oxidizing coals. Depending on the capacity of the drying oven, calculate the flow rate (set on the rotameter). An example of how to calculate the flow rate for the SLWN 115 STD at the temperature of 105ºC is given below.
  • Page 61 Instruction manual CL, SL, SLWN, SR, SRWP SMART where: - gas temperature on the rotameter: T = 273,15 + t = 273,15 + 25 = 298,15 [K] =105ºC: T - gas temperature in the chamber, in this case t = 273,15 + t = 273,15 + 105 = 378,15 [K] = 105 [ºC] : Calculation for the temperature in the chamber t...
  • Page 62: Srwp Pass-Through Sterilizers

    Instruction manual CL, SL, SLWN, SR, SRWP SMART 10. SRWP PASS-THROUGH STERILIZERS 10.1. Precautions When installing the device, follow the health and safety rules. Two people are needed to assemble the device. 10.2. Package contents • pass-through sterilizer • stand •...
  • Page 63: Installation Of Srwp Sterilizer

    Instruction manual CL, SL, SLWN, SR, SRWP SMART Wall Floor 10.4. Installation of SRWP sterilizer Fastening bolts Screw in the 4 fixing bolts - tilt the device in such a way that you will be able to screw the fixing bolts into the special holes in the four corners of the bottom of the device.
  • Page 64 Instruction manual CL, SL, SLWN, SR, SRWP SMART Placing the device in the aperture Insert the device with the mounted stand into the aperture in the wall to such a depth that it will be possible to screw the frame on the opposite side.
  • Page 65: Connecting The Device To A Computer

    Instruction manual CL, SL, SLWN, SR, SRWP SMART Frame and stand installation To install the frame and the stand you need the following equipment (not supplied by the manufacturer): • For masking element installation: - screws M4x8 (SRWP 115 -20 pcs.,SRWP 240 - 30 pcs.) •...
  • Page 66: Exterior Cleaning

    Instruction manual CL, SL, SLWN, SR, SRWP SMART 12.1. Exterior cleaning The housing of the device should be cleaned at least once a week, depending on the working conditions. The housing and door should be cleaned with caution using a soft cloth dampened with water. Only mild cleaning products should be used to clean the device.
  • Page 67: Cleaning The Touch Screen

    Instruction manual CL, SL, SLWN, SR, SRWP SMART 12.3. Cleaning the touch screen The touch screen is exposed to dirt, so it must be cleaned regularly. To clean the touch screen, use a clean and dry microfiber cloth. It is a very delicate material and collects dirt well. Before using the cloth, make sure that on the surface there are no crumbs or particles.
  • Page 68: Consumables

    • get full contact details of technical service • access to POL-EKO-APARATURA online catalogue, and information about accessories and related products • receive additional product information and special offers To receive information or technical assistance, contact the Service Department or visit the website: www.pol-...
  • Page 69: Warranty Conditions

    POL-EKO-APARATURA warrants that this product will be free from defects in material and workmanship for a period of two (2) years from date of the invoice. If a defect is present, POL-EKO-APARATURA will, at its option and cost, repair, replace, or refund the purchase price of this product to the customer, provided it is returned during the warranty period.
  • Page 70: Rating Plate

    Instruction manual CL, SL, SLWN, SR, SRWP SMART 16. RATING PLATE The rating plate is located on the left side wall in the upper left corner. Below there is an example of a rating plate: Manufacturer’s data Type of device Serial number (the two marked digits indicate the year of manufacture of the device) Temperature protection class according to DIN 12880 Degree of protection against electric shock (class I: protection against indirect contact) and IP enclosure protection...
  • Page 71: Technical Data

    Power supply 230 [V] ±10% / 50 [Hz] Shelves fitted/max 3/10 3/14 5/16 6/22 Warranty 24 months Manufacturer POL-EKO-APARATURA working chamber is always smaller depth does not include 50mm of power cable reinforced shelf reinforced version on uniformly loaded surface...
  • Page 72 Instruction manual CL, SL, SLWN, SR, SRWP SMART CL, SL, SLWN, SR, SRWP...
  • Page 73: Declarations Of Conformity

    Instruction manual CL, SL, SLWN, SR, SRWP SMART 18. DECLARATIONS OF CONFORMITY...
  • Page 74 Instruction manual CL, SL, SLWN, SR, SRWP SMART...
  • Page 75 Instruction manual CL, SL, SLWN, SR, SRWP SMART...
  • Page 76 Instruction manual CL, SL, SLWN, SR, SRWP SMART...
  • Page 77 Instruction manual CL, SL, SLWN, SR, SRWP SMART...
  • Page 78 ❑ warranty and post-warranty service ❑ consultancy in the selection, maintenance and operation of laboratory equipment POL-EKO LAB is Accredited by the Polish Centre for Accreditation (a member of ILAC) and provides accredited calibration of: ❑ thermostatic and climatic chambers (incubators, drying ovens, thermostatic cabinets, climatic chambers, freezers) ❑...

Table of Contents