Page 1
Instruction manual SMART PRO Climatic chambers with ultrasonic humidifier models: KK 115, KK 240, KK 350, KK 400, KK 500, KK 700, KK 750, KK 1200, KK 1450 Climatic chambers with steam humidifier models: KKS 115, KKS 240, KKS 400, KKS 750 Before using the equipment, please read carefully this instruction manual! Version 1.29 Issued 12.05.2022...
Page 2
Instruction manual KK, KKS SMART PRO Manufacturer's address: POL-EKO-APARATURA sp. j. ul. Kokoszycka 172 C 44-300 Wodzisław Śląski Country of origin: Polska As a manufacturer, we inform you that we took the necessary measures to ensure that this device fully meets your expectations and is reliable for a long period of use.
Instruction manual KK, KKS SMART PRO Contents: INTENDED USE AND IMPORTANT INFORMATION FOR THE USER ............6 PACKAGE CONTENTS ..........................7 BEFORE THE FIRST USE ..........................8 3.1. Reverse osmosis system (standard for KKS) ..................10 3.1.1. Working conditions and requirements ....................10 3.1.2.
Page 4
Instruction manual KK, KKS SMART PRO Quick Note – user’s message ....................45 6.5.4. 6.5.5. Alarm bar ............................45 6.6. Quick Program ............................46 6.7. Programs............................. 48 6.7.1. Creating / editing a program ......................49 6.7.2. Segments edition ..........................50 6.7.3.
Page 5
Instruction manual KK, KKS SMART PRO 6.20. E-mail reports ..........................89 6.21. Automatic defrosting function (standard for KK and KKS)............91 Temperature – additional temperature sensor Pt 100 (option) ............. 92 6.22. 6.23. Corrections ........................... 93 INTERFACE ..............................94 7.1. MODBUS TCP ............................
Instruction manual KK, KKS SMART PRO 1. INTENDED USE AND IMPORTANT INFORMATION FOR THE USER KK and KKS climatic chambers are devices that allow you to control temperature, humidity and optionally illumination. There are two series of chambers that differ in the temperature and humidity range and type of humidifier: •...
Instruction manual KK, KKS SMART PRO Failure to comply with the above recommendations may result in damage to the device or deterioration of technical parameters, as well as loss of warranty. 2. PACKAGE CONTENTS KK SMART PRO climatic chambers are delivered with: Device KK (FIT) Capacity...
Instruction manual KK, KKS SMART PRO 3. BEFORE THE FIRST USE The manufacturer sends the device protected by cardboard profiles and foil. The device should be transported in an upright position and the package should be secured against sliding during transport. After receiving the device, visually assess its condition and equipment in the presence of the person delivering the goods.
Page 9
Instruction manual KK, KKS SMART PRO Wheels The device has been equipped with leveling wheels. After placing the unit at its destination, secure the device against movement by locking the wheels. Each climatic chamber is equipped with wheels. They must be locked and leveled AFTER placing the chamber in the place of use.
Instruction manual KK, KKS SMART PRO Water and sewage installation for the KKS climatic chamber with steam humidifier The KKS climatic chamber is equipped with a pump for pumping out used water and with a reverse osmosis system, Section 3.1. Technical data of the pump: •...
Instruction manual KK, KKS SMART PRO 3.1.2. Delivery range Item Name Quantity Preview drawing reverse osmosis system 1 pcs. • WASTE • FEED 2m hose Ø6 4 pcs. • DEIONIZED • TO OSMOSIS • 2m hose Ø12 1 pcs. SEWAGE SYSTEM filters key 1 pcs.
Page 12
Instruction manual KK, KKS SMART PRO Figure 1 Connection scheme Starting the system (connect according to the above scheme) Connect the hose FEED Ø6 to the water connection and to the reverse osmosis system. Connect the hose WASTE Ø6 to the drainage (sewage system) and to reverse osmosis system (check valve). Connect the hose DEIONIZED Ø6 to the drainage (sewage system) and to reverse osmosis system.
Instruction manual KK, KKS SMART PRO Unscrew the water supply ball valve. To fill the water storage pressure tank, leave the system for about 2 hours. Due to the carbon refill, the outflowing water may be cloudy and may have a dark color. Close the water supply ball valve and wait until water stops flowing from the hose DEIONIZED.
Page 14
Instruction manual KK, KKS SMART PRO Connecting the hose: Push the hose into the quick coupler (Fig.2). Disconnecting the hose: Press the quick coupling collar symmetrically and pull out the hose (Fig.3). 3.1.4.2. Connecting the system to the KKS chamber Figure 4 Connection diagram...
Page 15
Instruction manual KK, KKS SMART PRO Connection of reverse osmosis system to the KKS climatic chamber (connect it according to the diagram above) Connect the hose FEED Ø6 to the KKS and to the water connection. Connect the hose TO OSMOSIS Ø6 to the KKS and to reverse osmosis system. Connect the hose WASTE Ø6 to the KKS and to the reverse osmosis system (check valve).
Instruction manual KK, KKS SMART PRO 3.1.5. Consumables and service life of reverse osmosis system The user is obliged to perform maintenance throughout the life of the equipment. Service life Consumable Symbol Preview drawing (months) mechanical FCPS20 insert I carbon cartridge FCCBL mechanical FCPS5...
Instruction manual KK, KKS SMART PRO Turn off the equipment. Note: Unplug the KKS climatic chamber from the power socket. Close the water supply ball valve and tank valve. Using a key for filters, unscrew the housings of filter refills and remove used refills. Note: The housings are filled with water.
Instruction manual KK, KKS SMART PRO Note: The system may be under pressure. You should be particularly careful. Use a membrane wrench or manually unscrew the housing head. Remove the used membrane from the housing (Figure Figure 5 TFC-75 GPD Aquafilter housing head hose Lubricate with technical vaseline rubber seals (O-rings) of the new osmotic membrane and the gaskets in the...
Page 19
Instruction manual KK, KKS SMART PRO Turn off the equipment. Note: Unplug the KKS climatic chamber from the power socket. Close the water supply ball valve and tank valve. Odłączyć wężyki od wkładu liniowego. Note: Remove the blue safety clips before disconnecting the hoses. Note: The system may be under pressure.
Instruction manual KK, KKS SMART PRO Note: secure the connection with a blue securing clip. Open the water supply ball valve and the tank valve. Turn on the KKS climatic chamber. 3.1.9. Tank replacement Required tools and materials: teflon tape wrench 10mm cross screwdriver PH2 Turn off the equipment.
Instruction manual KK, KKS SMART PRO Remove the tank clamp. Tank valve Teflon tape Tank Air valve Replace the tank. In the absence of a valve in a new tank, screw the valve to seal the thread with teflon tape. Due to the teflon seal, do not pull back the connector when screwing in - it may cause unsealing and water leakage.
Instruction manual KK, KKS SMART PRO The water tank for demineralized water should be placed on a shelf mounted at the back of the climatic chamber. A PZ2 cross screwdriver is required for mounting the shelf. To mount the shelf you have to: •...
Page 23
Instruction manual KK, KKS SMART PRO Installation sequence: Connect the orange power cord to the drain pump (1). Slide the condensate tray into the guides under the equipment (2). condensate tray 3. Secure the condensate tray by tightening the knurled thumb screws (3, 4)
Page 24
Instruction manual KK, KKS SMART PRO 4. Connect the drain hose to eg a sewage system or tank (5) outflow Technical data of the pump: • maximum lifting height - 2m, • maximum hose length - 5m (slope min.1%). Positioning of the hose Ø6/9mm:...
Instruction manual KK, KKS SMART PRO 3.4. Installation of shelves In the KK 350, KK 500, KK 700, KK 1200, KK 1450 models To install the shelf or to change its position, follow these steps: Install the shelf slide at the selected height by inserting it into proper slots on the wall of the device. Do the same with the slide on the opposite wall.
Instruction manual KK, KKS SMART PRO Slide the shelf into the installed shelf slides. Now, the shelf is correctly installed! To remove a shelf, perform the above steps in reverse order. 3.5. Internal glass door (standard for KK and KKS) Internal glass door is a standard equipment in KK and KKS climatic chambers.
Instruction manual KK, KKS SMART PRO • frequency of door openings, • temperature of the samples. Too high relative humidity in the chamber of the device may cause icing of the cooling element, and thus reduce the cooling capacity and increase electricity consumption. Cartons, sponges and other hygroscopic materials should not be used to store samples, as they may increase the humidity in the chamber.
Instruction manual KK, KKS SMART PRO 4.1. Design of KK / KKS devices Below there’s a picture of climatic chamber (exemplary photo) with a description of the important components of the device. Front view rating plate internal glass door access port ø30mm for external sensor shelf condenser housing...
Page 29
Instruction manual KK, KKS SMART PRO Rear view of KK 500, KK 700, KK 1200, KK 1450 climatic chamber integrated power cable container for deionized water condensate tray with a pump (for its installation go to the Section 3.3.) main switch LAN port fuse...
Page 30
Instruction manual KK, KKS SMART PRO Rear view of KK 115, KK 240, KK 400, KK 750 climatic chamber container for deionized water integrated power cable connectors connection (hose from a tank with demineralized water or deionizer) service connection drain hose connection (to the sewa- ge system, sink, etc.) main switch LAN port...
Page 31
Instruction manual KK, KKS SMART PRO Rear view of KKS climatic chamber main switch integrated power cable connectors tap water connection connection to the osmosis system check valve demineralized water connection drain hose connection (to the se- wage system) service connection 10) LAN port 11) fuse...
Instruction manual KK, KKS SMART PRO 5. DEVICE EQUIPMENT (STANDARD AND OPTIONAL) 5.1. External door with viewing window (optionally for KK/KKS: 115, 240, 400, 750) The external door with a viewing window is an optional equipment for KK/KKS 115, KK/KKS 240, KK/KKS 400, KK/KKS 750 climatic chamber.
Instruction manual KK, KKS SMART PRO 5.4. Open door alarm (standard for all units) All units have been equipped with an open door sensor. If you open the door, the icon: will appear (the number above the icon presents open door counter. Press the icon to cancel the counter. The counter is also cancelled by turning of the device).
Instruction manual KK, KKS SMART PRO Data saved in the * .csv file can be opened in a spreadsheet. Data saved as * .plkx can be opened in the Lab Desk program (standard equipment). This program allows, among others, for data preview in the form of a table or a graph. It also allows you to prepare a statistic report for a selected range of data, see Section 6.12.
Instruction manual KK, KKS SMART PRO 5.8. Display battery backup (optionally for KK and KKS) Climatic chambers in the SMART PRO version can be optionally equipped with a battery backup of the display. The power loss and switching to the battery backup display mode is signaled by a pulsating red frame around the display and a sound signal (if it is turned on).
Instruction manual KK, KKS SMART PRO 6.2. First boot During the first boot, the screen (Figure 9) will display information about saving the "Download" folder (with instruction manual) on the USB flash drive in pdf format. In order to do it, insert the USB flash drive and wait a second to detect the hardware, then press If you press you quit downloading the folder.
Instruction manual KK, KKS SMART PRO Sometimes entered characters can be hidden (then they are replaced with „* ”). This happens when classified information is entered (e.g. when a user enters a password). 6.4. User logging in Setting device parameters is only possible by the logged in user. To log in, press in the main screen.
Instruction manual KK, KKS SMART PRO Figure 12 Main screen (program is switched off, no user is logged in) From this point, any action requires logging in. Figure 13 Main screen – running program in KK segment other temperature duration time parameters main program...
Page 39
Instruction manual KK, KKS SMART PRO Figure 14 Information panel The icon indicates information about which window is active. 6.5.1.1. Alarms panel On the first page of the information panel there’s alarms panel (Figure 15). Figure 15 Icon: Alarms panel alarms panel In the alarms panel there’s a list with active alarms or the alarms that have occurred but have not been confirmed.
Page 40
Instruction manual KK, KKS SMART PRO Figure 16 Alarm details With more alarms, a button appears on the right side of the list and allows you to enlarge the view to full screen. 6.5.1.2. Status panel The status of the device is displayed descriptively on the third page of the information panel (Figure 17).
Instruction manual KK, KKS SMART PRO Figure 18 Status – protection and alarms 6.5.2. The meaning of icons and symbols The icon allows you to go to the main screen. Automatic return to the home screen. Factory setting: disabled. The icon allows you to go to the main menu. Automatic logout.
Page 42
Instruction manual KK, KKS SMART PRO Icon is visible only when the chamber is heating up. Icon is visible only when the automatic defrosting function or defrosting program is running. Available when the program is running Clicking the icon allows you to quickly change the set temperature (Quick Change function).
Instruction manual KK, KKS SMART PRO When the program is running, click the icon to quickly change the fan speed (Quick Change function). User message. Clicking on the icon allows you to enter a message. The icon appears in the event log and symbolizes entered user message. Active STM function (Smart Temperature Monitor) informs the user about the problem of reaching or maintaining the set temperature.
Page 44
Instruction manual KK, KKS SMART PRO Figure 19 Upper expandable menu when the program is running Figure 20 Upper expandable menu when the program is stopped Positions available on the upper bar can be personalized. Just drag the selected icon to a new location (Figure 21).
Instruction manual KK, KKS SMART PRO Quick Note – user’s message 6.5.4. During equipment’s operation, the user can save messages in unit’s memory with information about: the date of inserting a new sample, observed changes in the samples, the place of sampling, etc. To enter a message you must in the main screen in upper menu.
Instruction manual KK, KKS SMART PRO 6.6. Quick Program Quick Program allows you to quickly start the program from the main screen position without having to enter to the menu Quick Program has several features that guarantee its uninterrupted operation: •...
Page 47
Instruction manual KK, KKS SMART PRO Figure 24 Starting the Quick Program Press Set temperature Set fan if necessary Set humidity Set humidity Start Stopping the Quick Program has been made difficult on purpose (this prevents the program from being stopped acci- dentally) –...
Instruction manual KK, KKS SMART PRO Figure 25 Quick program on programs list In Quick Program editing mode, you can change: • settings of the data recording interval, • settings of the protection class. Temperature protection The highest protection class available for the device is set. The protection values depend on the set temperature: •...
Instruction manual KK, KKS SMART PRO Figure 26 List of programs Information on the number of created programs / the maximum number of programs that the user can create is at the top of the screen (programs: 1/10). 6.7.1. Creating / editing a program Press the button and a panel with program parameters will appear (Figure...
Instruction manual KK, KKS SMART PRO Cancels adding or editing of the program. Going to the edition of program segments. With more parameters, the window can be scrolled up and down. 6.7.2. Segments edition For each program, you can set maximum 100-segment time-temperature profiles that allow you to gradually increase or decrease the incubation temperature of the samples.
Page 51
Instruction manual KK, KKS SMART PRO The active value is highlighted in blue. The item highlighted in red means that the value is out of range and you should enter another one, e.g. the temperature is above / below the operating range of the device or the protection temperature.
Instruction manual KK, KKS SMART PRO Figure 29 6.7.3. Phytotron FIT (optionally for KK) The phytotron (FIT) function enables the simulation of day and night: duration and smooth regulation of light intensity. Phytotron version with illumination panels Before you start configuration of light parameters, turn on the illumination panel that will be used and configured (applies to devices with the FIT version in the form of lighting panels).
Page 53
Instruction manual KK, KKS SMART PRO Figure 31 Turning on/off the illumination panel Confirms changes Cancels the entered changes 6.7.3.1. Equipment with one illumination zone Equipment with one illumination zone can have fluorescent lamps mounted in: side walls/ doors/ walls and doors/ ligh- ting panel(s).
Instruction manual KK, KKS SMART PRO 2. To turn on the light in the segment, select the check box (Figure 33) and then set the intensity value for each turned on panel (Figure 34). Figure 33 Turning on the light Figure 34 Setting the light intensity in active panels Confirms and saves the changes Cancels the entered changes in the segments and goes to program parameters...
Instruction manual KK, KKS SMART PRO Figure 35 The summary of the segment Confirms and saves the changes Cancels the entered changes in the segments and goes to program parameters With more parameters, the window can be scrolled up and down. 6.7.5.
Instruction manual KK, KKS SMART PRO 6.7.7. Priority Can be set in terms of: Parameters: In the program without ramp – the device starts the countdown of the segment time when the set temperature is reached. In the program with ramp – first, the device counts down the time of the ramp and then proceeds to the segment countdown when the set temperature is reached.
Page 57
Instruction manual KK, KKS SMART PRO Figure 37 Main menu Figure 38 Program management menu If the program is running, the symbol appears next to the program name on the list. The symbol means that the program has been edited, but the changes have not been confirmed (Figure 39).
Instruction manual KK, KKS SMART PRO 6.8.2. The second way • In the main screen press the icon in the upper right corner (Figure 40) • In the upper left corner press „PROGRAM” • Select the program you want to start (Figure 41).
Instruction manual KK, KKS SMART PRO If the program is running, the symbol appears next to the program name on the list 6.9. Quick change of parameters You cannot make a quick change (of time / temperature) in a running program that belongs to another user.
Instruction manual KK, KKS SMART PRO Figure 43 Quick change of set temperature 6.9.2. Quick change of set humidity In order to quickly change the set humidity of a running program, press the icon in the main screen (Figure 44). Choose the humidity value by scrolling the list up or down (Figure 45).
Instruction manual KK, KKS SMART PRO Figure 45 Quick change of set humidity 6.9.3. Quick change of set time In order to quickly change the duration time of a running program, press the icon in the main screen (Figure 46). Select the number of days, hours and minutes by scrolling the list up or down (Figure 47).
Instruction manual KK, KKS SMART PRO Figure 47 Quick change of set time 6.9.4. Quick change of fan efficiency In order to quickly change the fan efficiency, press the icon in the main screen (Figure 48). The value should be selected by scrolling the list up or down (Figure 49).
Instruction manual KK, KKS SMART PRO Figure 49 Quick change of fan efficiency 6.10. Schedules The option allows creating a list of programs to be implemented in a given time. You can create several independent schedules. The schedules window contains a list of all created schedules of the logged-in user (Figure 50).
Instruction manual KK, KKS SMART PRO Add the schedule Edit the schedule Delete the schedule 6.10.1. Creating / editing a schedule To create / edit a schedule, press the button . The panel with schedule parameters will appear on the screen (Figure 51).
Page 65
Instruction manual KK, KKS SMART PRO Figure 52 Adding a program to the schedule Select a program from the drop-down list and press on the field next to the inscription "program" (Figure 53). Informa- tion about the selected program will be displayed (Figure 54): number of segments, number of cycles, priority, tempera- ture protection, upper protection, lower protection.
Instruction manual KK, KKS SMART PRO Press the 'start date' field and then set the date and time of program start. Press the 'stop date' field and then set the date and time of the program end. You can assign up to 10 programs to one schedule. In total you can create ten schedules. When creating a schedule, consider the following restrictions: •...
Page 67
Instruction manual KK, KKS SMART PRO Figure 55 Main menu Figure 56 List of schedules Start the schedule 6.10.2.2. The second way • In the main screen press the icon (Figure 57), then press the SCHEDULE inscription. The schedule se- lection window will be displayed (Figure 58).
Instruction manual KK, KKS SMART PRO Figure 57 Main screen Figure 58 Selection of the schedule 6.11. Statistics Go to the main menu and press the icon . This panel (Figure 59) displays statistics of the currently running program or program that has ended Statistics are calculated separately for each segment. Data logging for calculation starts after 30 seconds from reaching the set temperature in the segment.
Instruction manual KK, KKS SMART PRO finished – the segment has been completed, interrupted – the segment was interrupted by the User before the set time has elapsed, • segment 1/2 – the number of the currently overviewing segment / number of the currently performed or completed segment.
Instruction manual KK, KKS SMART PRO Figure 60 Data record Press to continue downloading data Recording data onto the USB flash drive. .csv files are available - separated by semicolon when opening e.g. with a spreadsheet, .plkx - opening with the Lab Desk application Before removing the USB flash drive from the USB port, it must be unmounted, see Section 5.5.
Page 71
Instruction manual KK, KKS SMART PRO Figure 62 Temperature graph Returns to displaying the entire chart (undo all magnifications) / returns to the list of charts. Returns to data register Figure 63 Humidity graph Returns to displaying the entire chart (undo all magnifications) / returns to the list of charts. Returns to data register...
Instruction manual KK, KKS SMART PRO Figure 64 Temperature and humidity graph Returns to displaying the entire chart (undo all magnifications) / returns to the list of charts. Returns to data register The opening time of the chart depends on the number of saved data samples. The greater the number of saved samples, the longer this window will open.
Instruction manual KK, KKS SMART PRO The date / time, temperature of the sensor installed in the chamber and indications of additional sensors in the unit (humidity, additional temperature sensor) are saved to the file. The frequency of saving to the file is equal to the frequency of saving the data in the register set in the program, see section 6.7.1.
Page 74
Instruction manual KK, KKS SMART PRO Recording data onto the USB flash drive. .csv files are available - separated by semicolon when opening e.g. with a spreadsheet, .plkx - opening with the LabDesk application Before removing the USB flash drive from the USB port, it must be unmounted, see Section 5.5.
Page 75
Instruction manual KK, KKS SMART PRO DeviceOff the device is switched off (on the main switch) Door opened the door is opened Open door alarm start open door alarm has been activated Door closed the door is closed Open door alarm stop open door alarm has been deactivated Program Restarted program has been reasumed after power failure...
Instruction manual KK, KKS SMART PRO Power Fail Stop power reasumed, returned to maintain program parameters User login date and time of login User logout date and time of logout Quick note To view message details, click . In this window (Figure 68) you can see the content of the message, the name of the user who entered it and the name of the program during which the message was written.
Page 77
Instruction manual KK, KKS SMART PRO Figure 69 Info window (example) to save the “Download” folder (with instruction manual) on the USB flash drive. After inserting the flash Press icon drive into USB port wait few seconds until the information “Flashdrive connected” will appear on the display - for more to write the service data on the USB flash drive –...
Instruction manual KK, KKS SMART PRO 6.15. Users Go to the main menu and press the icon . In this panel (Figure 71) you can add a new user, edit an existing one or delete it. Adding a new user Editing selected user Deleting selected user - his programs and data register will be deleted with the user.
Instruction manual KK, KKS SMART PRO Figure 72 Editing a user Confirms and saves the user Cancels introduced changes and returns to the users list The device can have maximum 5 users. There are 40 programs available which can be freely distributed among users.
Page 80
Instruction manual KK, KKS SMART PRO Temporarily silencing the alarms ✔ ✔ ✔ ✘ ✘ Enabling / disabling the sound ✔ Saving a Quick Note ✔ ✔ ✔ Creating users accounts ✔ ✘ ✘ Changing user's settings ✔ ✘ ✘ Changing equipment's name ✔...
Page 81
Instruction manual KK, KKS SMART PRO Admin account Has access to programs menu and has rights and limitations in accordance with table Figure 74 Menu available for Admin User account: • has access to programs menu , where User can start programs previously assigned to him, check their statistics (statistics, data register), check events history of the equipment (event log) and the information about the system (info), •...
Instruction manual KK, KKS SMART PRO 6.16. User settings panel Go to the main menu and press the icon . In this panel (Figure 76) you can: Change the name of the equipment – by default, the device serial number is entered. Change the language in the equipment´s menu.
Instruction manual KK, KKS SMART PRO 6.16.1. Unlocking the touch screen When the automatic touch screen lock is enabled (Section 6.16), slide the blue circle into the white circle to unlock the screen. Figure 77 Unlocking the touch screen 6.17. Time Go to the main menu and press the icon...
Page 84
Instruction manual KK, KKS SMART PRO If the unit is permanently connected to the internet, the time will be synchronized with the time server. Time synchronization is performed: • manually using the button • while enabling the automatic synchronization option and then every 12 hours ( •...
Instruction manual KK, KKS SMART PRO 6.18. Alarms Go to the main menu and press the icon . You can set parameters related to alarms. • lower temperature alarm – an alarm will be activated if the temperature drops below the value given in this field, •...
Instruction manual KK, KKS SMART PRO 6.18.1. STM function The STM function (Smart Temperature Monitor) informs the user about the problem of reaching or maintaining the set temperature. The user can enable / disable the function (Figure 81). If the STM function is on, the STM symbol will appear on the screen next to the temperature of the main sensor.
Instruction manual KK, KKS SMART PRO Figure 83 Option enabled, warning about problems with reaching / maintaining the temperature Possible causes: damaged heater, the sample inserted into the chamber absorbs / dissipates too much energy. If the color was red before opening the door, the color changes to blue after opening the door. In case of activation of the function (detection of problems with reaching / maintaining the temperature): •...
Page 88
Instruction manual KK, KKS SMART PRO • Gate – Server’s IP address or router’s that manages the Ethernet network • DNS – IP address of the domain name system • MAC – the address of the network card, read-only • DHCP –...
Instruction manual KK, KKS SMART PRO Figure 85 WiFi settings Confirms changes Cancels the entered changes 6.20. E-mail reports Go to the main menu and press the icon . In this window you can set the parameters needed to activate e- mail notifications.
Page 90
Instruction manual KK, KKS SMART PRO Confirms changes Cancels the entered changes If the "activate" option at the top of the panel is not enabled, emails will not be sent! Sender and recipients (Figure 87) • Sender – sender’s e-mail address •...
Instruction manual KK, KKS SMART PRO Figure 88 E-mail: email account configuration Connection test Confirms changes Cancels introduced changes Before testing the connection, make sure that the device is connected to the network and has a properly configured network connection, see Section 6.19.
Instruction manual KK, KKS SMART PRO Figure 89 Defrosting program Confirms changes Cancels the entered changes Temperature – additional temperature sensor Pt 100 (option) 6.22. Go to the main menu and press the icon . In this panel (Figure 90). You can set parameters related to the temperature measurement in the equipment using an additional temperature sensor.
Instruction manual KK, KKS SMART PRO Figure 90 Temperature measurement settings with an additional sensor Confirms changes Cancels the entered changes 6.23. Corrections Go to the main menu and press the icon . In this window (Figure 91) you can correct: •...
Instruction manual KK, KKS SMART PRO 91 User’s correction Figure Confirms changes Cancels the entered changes 7. INTERFACE 7.1. MODBUS TCP The device allows status monitoring using the MODBUS TCP communication interface. Connection parameters: • IP address: same as device’s (set in the panel Section 6.19) •...
Instruction manual KK, KKS SMART PRO 8. TEMPERATURE PROTECTION The device is factory fitted with sample protection - temperature protection. If any of the elements responsible for maintaining the set temperature is damaged or the user sets the temperature unconsciously, the set protection will work.
Instruction manual KK, KKS SMART PRO The device is equipped with a protection mechanism against damaging the cooling system. The mechanism makes it impossible to turn on cooling when the temperature exceeds 45ºC. As a result if the device has been programmed to go down to a lower temperature (e.g.
Instruction manual KK, KKS SMART PRO wipe the container with a dry cloth until it’s completely dry, install the disc and the protective strainer in the condensate level sensor container Pay attention to the correct assembly of the protective disc (see the drawing below) slide back the condensate tray under the device.
Instruction manual KK, KKS SMART PRO for: KK 500, KK 700, KK 1200, KK 1450 view from the top (the aggregate is located in the upper part of the device) for: KK 350, KK/KKS 115, KK/KKS 240, KK/KKS 400, KK/KKS 750 Failure to clean regularly may result in damage to the compressor and loss of the rights for repair under warranty.
Instruction manual KK, KKS SMART PRO Figure 92 Locking the screen The screen is ready to be cleaned. To unlock the touch screen, slide the blue circle into the white circle Figure 93 Unlocking the screen 12. ADVICE ON HOW TO SAFELY STORE THE DEVICE Remove all objects from the chamber.
• get full contact details of technical service • access to POL-EKO-APARATURA online catalogue, and information about accessories and related products • receive additional product information and special offers To receive information or technical assistance, contact the Service Department or visit the website: www.pol-...
POL-EKO-APARATURA warrants that this product will be free from defects in material and workmanship for a period of two (2) years from date of the invoice. If a defect is present, POL-EKO-APARATURA will, at its option and cost, repair, replace, or refund the purchase price of this product to the customer, provided it is returned during the warranty period.
Instruction manual KK, KKS SMART PRO 15. RATING PLATE The rating plate is located on the left side wall in the upper left corner. Below there is an example of a rating plate: Manufacturer’s data Type of device Serial number (the two marked digits indicate the year of manufacture of the device) Temperature protection class according to DIN 12880 Degree of protection against electric shock (class I: protection against indirect contact) and IP enclosure protection rating...
3/11 3/14 3/11 3/11 5/16 2 x 3/11 2 x 3/11 Warranty 24 months Manufacturer POL-EKO-APARATURA the above parameters apply to standard devices (without optional equipment) depth does not include 50mm of power cable on uniformly loaded surface reinforced shelf...
Page 104
Instruction manual KK, KKS SMART PRO KK 500, 700, 1200, 1450 KK 115, 240, 400, 750...
Instruction manual KK, KKS SMART PRO 17. DECLARATIONS OF CONFORMITY...
Page 107
❑ warranty and post-warranty service ❑ consultancy in the selection, maintenance and operation of laboratory equipment POL-EKO LAB is Accredited by the Polish Centre for Accreditation (a member of ILAC) and provides accredited calibration of: ❑ thermostatic and climatic chambers (incubators, drying ovens, thermostatic cabinets, climatic chambers, freezers) ❑...
Need help?
Do you have a question about the SMART PRO KKS 115 and is the answer not in the manual?