Accuracy (cont'd)
AC Current
Range (40 Hz to 400 Hz)
400.0 µA
4000 µA
40.00 mA
400.0 mA
4.000 A
10.00 A*
* 2 A continuous, > 2 A to 10 A for 10 second max with 15 minutes cool down interval.
10.00 A range is specified from 20% to 100% of range.
DC Current
Range
400.0 µA
4000 µA
40.00 mA
400.0 mA
4.000 A
10.00 A*
* 2 A continuous, > 2 A to 10 A for 10 second max with 15 minutes cool down interval.
10.00 A range is specified from 20% to 100% of range.
Frequency
Range
9.999 Hz
99.99 Hz
999.9 Hz
9.999 kHz
99.99 kHz
999.9 kHz
9.999 MHz
Input Voltage: 1 V rms ~ 20 V rms
14
Accuracy
± (1.8% + 1.0 µA)
± (1.8% + 10 µA)
± (1.8% + 0.10 mA)
± (1.8% + 1.0 mA)
± (2.5% + 0.010 A)
± (2.5% + 0.10 A)
Accuracy
± (1.0% + 0.7 µA)
± (1.0% + 7 µA)
± (1.0% + 0.07 mA)
± (1.0% + 0.7 mA)
± (1.5% + 0.007 A)
± (1.5% + 0.07 A)
Accuracy
± (1.0% + 0.005 Hz)
± (1.0% + 0.05 Hz)
± (1.0% + 0.5 Hz)
± (1.0% + 0.005 kHz)
± (1.0% + 0.05 kHz)
± (1.0% + 0.5 kHz)
Unspecified
Burden Voltage (typical)
Burden Voltage (typical)
Duty Cycle
Range: 5% ~ 95%
Accuracy: ± (2% + 0.7)
Input Voltage: 4 V ~ 10 V peak to peak
Frequency Range: 4 Hz ~ 1 kHz
51 µV/µA
2.7 mV/mA
0.04 V/A
51 µV/µA
2.7 mV/mA
0.04 V/A