Hide thumbs Also See for ARA DOCK4:

Advertisement

Operation
1. Firmly place between 1 and 4 ION ARA single gas detectors face-down into the
bays.
2. To conduct a bump test, press the blue button labelled 'bump test'. The bay LED
status will illuminate amber during the test. Upon completion, the bay LED will
indicate green if passed and red if failed.
3. To conduct a calibration, press the green button labelled 'calibration'. The bay LED
status will illuminate amber during the test. Upon completion, the bay LED will
indicate green if passed and red if failed.
NOTE: The gas is only supplied to the bays where a detector is present.
Briefly applies gas to test the sensor response, downloads logged events
and tests the audible and visual alarms
Adjusts the sensor settings to match the gas applied, downloads logged
events and test the audible and visual alarms
ARA Status LEDs
Troubleshooting
• Check the built-in pressure gauge to ensure there is suffcient calibration gas
(34L & 57L are full at 500psi). Inspect the ARA inlet port. Clear any obstructions and
replace clogged or dirty filters.
• Clean ARA IR communication window with a non-alcohol-based cleaner.
• Bright sources of light may interfere with the ION ARA DOCK4 IR communication ports,
try relocating.
If you are still expeiencing problems after troubleshooting, please email
technical.support@ionscience.com or call 01763 208 503.
Pioneering Gas Sensing Technology.
ionscience.com
ARA DOCK4 Bump Test &
Calibration Station
Quick start guide
Test passed
Test in progress
Test failed
2.
Part No: 908225 V1.0