6.. FAQ

Anomalies
The device does not turn on
Display shows an incorrect
value
Display value is not reset to 0

7. Charging

Uni-Link is charged using a standard high quality USB
charger. Always on stock at Vetec or available in any store.
Charge the Uni-Link in your car, connected to your computer or using a wall charger. Charge tem-
perature 0°C to 45°C. A completely discharged battery takes up to 6 hours to recharge.
As the charger is not waterproof charging the unit in a safe dry place is recommended. Place the
Uni-Link on a stable surface and make sure it is turned off. Plug in and charge up using the USB
port on the bottom of the Uni-Link.
8. Service, repairs and maintenance
To prevent damage to the Uni-Link, make sure that repairs and recalibration are carried out by
qualified personnel.
It is recommended that recalibration and repairs are done at Vetec's service department
located in Denmark or by authorized partners. Contact Vetec's service department for further
information. When you return a Uni-Link for repair, make sure to always include the invoice
number or the delivery note along with a description of the problem.

8.1. Battery replacement

The battery indicator blinks when it needs to be replaced. It is recommended to purchase the
battery replacement kit that includes a Li-on battery and a new foil.
- Remove the foil.
- Use a suitable screwdriver to remove the battery cover.
- Replace the battery with a new battery.
- Put the battery cover back on and tighten the screws so that the cover is flush with the
edge of the battery case. Ensuring that it is is in correct position.
- Replace the foil with a new foil. This will ensure that the device is waterproof.
Possible causes
Discharged batteries
Electronics defective
More than 1 year since last
calibration
Malfunction of electronics
Has been overloaded
Has not been reset
Has been overloaded
- 8 -
Corrective actions
Charge the battery
Contact Vetec's service depart-
ment
Re-calibrate at Vetec service
department
Reset: Turn off and on and retest
Contact Vetec's service department
Press →0←
Contact Vetec's service depart-
ment.