Set Up: Your PROSE Case

To start, you will need to have a clean work surface and the following supplies:
• One PROSE case
• One unopened package of Clear Care
• Please note that you can use original Clear Care or Clear Care Plus
Remove the lid from the PROSE case.
1
Remove the lid from the small Clear Care case.
2
Remove/break off the catalyst from the arbor on the bottom of the Clear Care
3
case (as illustrated below).
Discard the Clear Care case.
4
*NEVER attempt to disinfect your PROSE device in the case included
with the Clear Care Solution
If your PROSE case still has the arbor (stem) attached to the bottom of the
5
basket assembly, it will need to be removed by using pliers to grasp the arbor
and gently bending the arbor until it breaks. If not, continue to step 6.
Drop the platinum disc from the Clear Care case into the PROSE case so that
6
it sits on the bottom of the PROSE case.
We recommend replacing the platinum disc every two months. Under NO
circumstances should the platinum disc be used for more than 90 days.
Your PROSE case is now properly set up for nightly disinfection.
7
We recommend replacing your PROSE case every 3 months.