application – lighted plunger

How to Apply Your PROSE Device with a Lighted Plunger

Wet the fenestrated end of the application plunger with
1
preservative-free saline. Attach the plunger to the plunger light
by inserting the light into the hole at the end of the fenestrated
plunger.
While holding a cleaned and disinfected PROSE device
2
gently by the edges, apply to an application plunger by gently
squeezing the plunger at the line while placing the device on
the cup and releasing the 'squeeze' to create suction. Refer
to page 8 for complete instructions. Turn on the green light by
pressing the black button.
3
Holding the plunger vertically, overfill device with
preservative-free saline. Let the solution drip, not stream into the device, in
order to prevent bubbles from forming.
Maintain proper lid spread. Hold the eyelids at the edges, where the lashes
4
meet the lids. The contact points should be centered on both lids, so as to
form a perfectly round opening with plenty of clearance for the device.
Tuck your chin to your chest. Look straight down. Keep the
5
plunger centered in your eye by focusing on the green light in
the center of the plunger.
Continue approaching your eye smoothly and steadily, and
6
stare at the green light as if you are looking through it. Some
people find it useful to shut the eye they are not applying the
device to.
You will feel the fluid but you are not there yet. Press your
7
device onto the eye with firm pressure.
Capture the device with the eyelids by releasing the lids. Don't
8
squint. Maintain the upward pressure on the device while
squeezing the plunger in order to release the plunger from
the device.
The device has now been released onto your eye! Continue squeezing the
9
plunger while you pull the plunger away from the eye.
Check your device in the mirror. Shine a penlight, at an angle, onto the surface
10
of the device and look for the presence of any bubbles in or under the device.
If you detect bubbles the device MUST be removed and reapplied.
After application of both devices, clean your application plunger with warm
11
soapy water, rinse well, and allow to air dry.
10